CATECHOL
Organic Service Branch I 1. General Discussion
The OSHA Technical Center has received many requests for a sampling and analytical procedure for
catechol. OSHA promulgated an exposure standard for catechol of 5 ppm (20 mg/m3)
TWA. OSHA method 32 recommends collection of phenol and cresol on 1.1.2. Potential workplace exposure (Ref. 5.2. and 5.3.) Catechol is used as a topical antiseptic, reagent, antifungal preservative on seed potato pieces, photographic developer, and developer in fur dyes. Catechol is used as an antioxidant in many industries including rubber, chemical, dye, photographic, pharmaceutical, fat, cosmetics, and oil. 1.1.3. Toxic Effects (This section is for information purposes and should not be taken as the basis for OSHA policy.)(Ref. 5.3.) The lethal human dose of catechol is 50 to 500 grams/kilogram, or 1 teaspoon to 1 ounce for a 70 kilogram (150 pound) person, with death resulting from pulmonary failure. Catechol is a skin, eye, mucous membrane, and pulmonary irritant. It is readily absorbed from the gastrointestinal tract, and through the skin. Catechol can cause elevated blood pressure through vasoconstriction, degeneration of the renal tubes in the kidneys, and diminished liver function. 1.1.4. Physical properties (Ref. 5.2.):
1.2. Limit defining parameters
1.2.2. The overall detection limit is 0.014 ppm catechol. (All ppm amounts in this study are based on a 100 liter air volume and 3 mL desorption.) 1.3. Advantages
1.3.2. The analytical method is reproducible and sensitive. 1.3.3. Reanalysis of samples is possible. 1.3.4. It may be possible to analyze other compounds at the same time. 1.3.5. Interferences may be avoided by proper selection of column and GC parameters. 1.4. Disadvantages none known 2. Sampling procedure
2.1.2. The sampling media consists of OVS-7 tubes. The OVS-7 tubes are specially made
13 mm O.D.. glass tubes that are tapered to 6 mm O.D.. These tubes are packed with a 13 mm
diameter glass fiber filter then a 270 mg sampling section followed by a 140 mg backup
section of purified 2.2. Sampling technique
2.2.2. Connect the OVS-7 to the sampling pump with flexible tubing. 2.2.3. Place the tubes in a vertical position to minimize section towards the channeling, with the smaller pump. 2.2.4. Air being sampled should not pass through any hose or tubing before entering the 2.2.5. Seal the OVS-7 with plastic caps immediately after sampling. Seal each sample lengthwise with OSHA 2.2.6. With each batch of samples, submit at least one blank tube from the same lot used for samples. This tube should be subjected to exactly the same handling as the samples (break ends, seal, & transport) except that no air is drawn through it. 2.2.7. Transport the samples (and corresponding paperwork) to the lab for analysis. 2.2.8. Bulks submitted for analysis must be shipped in a separate mailing container from other samples. 2.3. Desorption efficiency
Table 1
2.3.2. Six filters were placed into separate 4 mL vials and spiked at each loading of 0.193 mg
(0.429 ppm), 0.96 mg (2.13 ppm), 1.93 mg (4.29 ppm), and 4.00 mg (8.88 ppm) catechol and allowed
to equilibrate overnight at room temperature. They were extracted with 3 mL of the desorbing
solution for 30 minutes with occasional shaking, and were analyzed by
Table 2
2.4. Retention efficiency
The filters of six OVS-7 tubes were spiked with 2.0 mg (4.44 ppm) catechol, allowed to
equilibrate overnight and then had 100 liters humid air (91% RH) pulled through them. The
glass fiber filter was placed before the Teflon spacer to insure that no catechol spiked
onto the filter was in contact with the
Table 3
2.5. Storage
Glass fiber filters(GFF) from the OVS-7 tubes were removed and spiked with 2.23 mg (4.95 ppm)
catechol and placed in a 4 mL vial containing the front section of
Table 4
2.6. Precision The precision was calculated using the area counts from six injections of each standard at concentrations of 0.0644 mg/mL (0.429 ppm), 0.320 mg/mL (2.13 ppm), 0.644 mg/mL (4.29 ppm), and 1.33 mg/mL (8.86 ppm) catechol in the desorbing solvent. The pooled coefficient of variation was 0.0105.(Table 5)
Table 5
where:
CVl, CV2, CV3, CV4 = Coefficients at each level 2.7. Air volume and sampling rate studied
2.7.2. The sampling rate studied is 1 liter per minute. 2.8. Interferences Suspected interferences should be listed on sample data sheets. 2.9. Safety precautions
2.9.2. Safety glasses should be worn at all times in designated areas. 2.9.3. Follow all safety practices that apply to the workplace being sampled. 3. Analytical method
3.1.2. GC column capable of separating the analyte and an internal standard from any interferences. The column used in this study was a 15 meter DB-WAX capillary column 0.25µ d.f., 0.32mm I.D. 3.1.3. An electronic integrator or some other suitable method of measuring peak areas. 3.1.4. Two and four milliliter vials with Teflon-lined caps. 3.1.5. A 10 µL syringe or other convenient size for sample injection. 3.1.6. Pipettes for dispensing the desorbing solution. The Glenco 1 mL dispenser was used in this method. 3.1.7. Volumetric flasks - 5 mL and other convenient sizes for preparing standards. 3.1.8. An analytical balance capable of weighing to the nearest 0.01 mg. 3.2 Reagents
3.2.2. Catechol, Reagent grade 3.2.3. Methanol, HPLC grade 3.2.4. Dimethyl formamide, Reagent grade 3.2.5. The desorbing solution is 0.25 µL/mL dimethyl formamide in methanol. 3.3. Sample preparation
3.3.2. Each section is desorbed with 3 mL of the desorbing solution. 3.3.3. The vials are sealed immediately and allowed to desorb for 30 minutes on a shaker, a roto-rack, or a sample rocker. 3.3.4. Samples were transfered to two milliliter vials for anaylsis, as this was the size needed to fit in the autosampler. 3.4. Standard preparation
3.4.2. At least two separate stock standards should be made. Dilutions of the stock standards are prepared covering the concentrations in the samples. The analytical standards used in this study ranged from 0.001 to 1.33 mg/mL of catechol the desorbing solution. 3.5. Analysis
3.5.2. Peak areas are measured by an integrator or other suitable means. 3.6. Interferences (analytical)
3.6.2. Retention time data on a single column is not considered proof of chemical identity. Samples over the target concentration should be confirmed by GC/Mass Spec or other suitable means. 3.7. Calculations
3.7.2. The area counts for the samples are plotted with the calibration curve to obtain the concentration of catechol in solution. 3.7.3. To calculate the concentration of analyte in the air sample the following formulas are used:
* All units must cancel. 3.7.4. The above equations can be consolidated to form the following formula. To calculate the ppm of analyte in the sample based on a 100 liter air sample:
3.7.5. This calculation is done for each section of the sampling tube and the results added together. 3.8. Safety precautions 3.8.1. All handling of solvents should be done in a hood. 3.8.2. Avoid skin contact with all chemicals. 3.8.3. Wear safety glasses, gloves and a lab coat at all times. 4. Recommendations for further study Collection studies need to be performed.
Figure 1. A diagram of an OVS-7 tube.
with 0.25 µL/mL dimethyl formamide internal standard. 5. References
5.2. Windholz, M., "The Merck Index", Eleventh Edition, Merck & Co., Rahway N.J., 1989, p.1272. 5.3. "Documentation of the Threshold Limit Values and Biological Exposure Indices", Fifth Edition, American Conference of Governmental Industrial Hygienists Inc., Cincinnati, OH, 1986, p. 112.
|