LASSO
Carcinogen/Pesticide Branch 1. General Discussion
Recently the OSHA Analytical Laboratory received a set of field samples requesting
analysis for Lasso. The air samples had been collected on glass fiber filters. This report
describes the analytical procedure developed and the preliminary validations of the
sampling method. Lasso (Alachlor) is a 1.1.2. Toxic effects (This section is for information only and should not be taken as the basis of OSHA policy) Possible genetic damage by the herbicide, Lasso was indicated by a paper. (Ref. 5.2.) 1.1.3. Potential workplace exposure
Lasso (Alachlor) is used in the 1.1.4. Physical properties
1.2. Limit defining parameters 1.2.1. Detection limit of the analytical procedure The detection limit of the analytical procedure is 1.5 ng Lasso per injection. This is the amount of analyte which will give a peak whose height is approximately five times the amplitude of the baseline noise. See figure 1. 1.2.2. Detection limit of the overall procedure The detection limit of the overall procedure is estimated to be 0.5 ug per sample or 0.005 mg/m3 based on the recommended air volume, assuming 100% recovery from the sampling device. The recovery test at this level has not been performed. 1.2.3. Sensitivity The sensitivity of the analytical procedure over a concentration range of 0.98 to 12.2 ug/mL is 47,890 area units per ug/mL of Lasso. The sensitivity is determined by the slope of the calibration curve. See figure 2. 1.3. Advantages The analytical procedure is rapid, sensitive, and reproducible. 1.4. Disdvantages
1.4.2. Both the glass fiber filter and the backup pad must be analyzed for Lasso separately. 2. Sampling Procedure
2.1.2. Glass fiber filter,
2.1.3. Backup pad,
2.1.4. Filter holder for 2.2. Sampling technique
2.2.1. Assemble the filter in the 2.2.2. Attach the outlet of the filter cassette to the personal sampling pump inlet with flexible tubing. 2.2.3. Air being sampled should not pass through any hose or tubing before entering the filter cassette. 2.2.4. A sample size of 100 liters is recommended. Sample at a flow rate of 1.0 liters/minute. The flow rate should be known with an accuracy of ±5%. 2.2.5. With each batch of samples, submit a blank filter from the same lot of filters used for the sample collection. This filter must be subjected to exactly the same handling as the samples except that no air is drawn through it. Label this filter as a blank. 2.2.6. The cassette should be shipped in a suitable container designed to prevent damage in transit. The samples should be shipped to the laboratory as soon as possible. 2.2.7. A sample of the bulk material should be submitted to the laboratory in a glass container with a Polyseal cap. Never transport, mail, or ship the bulk sample in the same container as the sample or blank filter. 2.3. Retention efficiency Three glass fiber filters were spiked with 49 ug of Lasso. Humid air (85% relative humidity) 100 liters was drawn through the filters at Lpm. The recovery of the filter plus the backup pad was 93%.
2.4. Extraction efficiency glass fiber filter The average extraction efficiency from the glass fiber fibers spiked with 49 ug of Lasso was 91%.
2.5. Extraction efficiency backup pad The average extraction efficiency from the backup pad spiked with 49 ug of Lasso was 102%.
2.6. Storage Storage test was not done. 2.7. Recommended Air Volume and Sampling Rate 2.7.1. The recommended air volume is 100 liters. 2.7.2. The recommended sampling rate is 1 Lpm. 2.8. Interferences There are no known interferences associated with the sampling procedure. 2.9. Safety Precautions 2.9.1. Attach the sampling equipment to the worker in such a manner that it will not interfere with work performance or safety. 2.9.2. Follow all safety practices that apply to the work area being sampled. 3. Analytical Method
3.1.2. HPLC reverse phase 3.1.3. An electronic integrator or other suitable method to measure detector response. 3.1.4. Microliter syringe or automatic sampling device for making sample injections. 3.1.5. Volumetric flasks of convenient sizes for preparing standards. 3.1.6. Shaking device for extraction of samples. 3.2. Reagents 3.2.1. Lasso (EPA standard 8452) 3.2.2. Acetonitrile, HPLC grade 3.2.3. Water, HPLC grade 3.3. Sample Preparation 3.3.1. Remove the filter from the cassette with clean tweezers and place it in a 20-mL scintillation vial. 3.3.2. Add 5 mL acetonitrile to the vial and cap it. 3.3.3. Shake the vials vigorously on a shaker for 30 minutes. 3.4. Standard Preparation 3.4.1. Standard of Ramrod is prepared by dissolving 9 to 12 mg (accurately weighed) of Lasso in acetonitrile in a 10 mL volumetric flask and making it to volume. 3.4.2. Dilute to the working range of 0.9 to 12 ug/mL with acetonitrile. 3.4.3. Store standards in dark bottles under refrigeration. 3.5. Analysis 3.5.1. HPLC Conditions
3.5.2. Chromatogram See Figure 1. 3.5.3. Peak magnitude is measured by electronic integrator or other means. 3.5.4. An external standard procedure is used to prepare a calibration curve from the analysis of at least three different concentrations from two separate weighings. 3.5.5. Bracket the sample with analytical standards. 3.6. Interferences (analytical) 3.6.1. Any collected compound that has the same LC retention time as analyte and absorbs at 214 nm and 254 nm is an interference. 3.6.2. HPLC parameters may be varied to circumvent most interferences. 3.6.3. Retention time alone is not proof of a chemical identity. Confirmation by other means should be sought when possible. 3.7. Calculations 3.7.1. The integrator value in area units for each standard is plotted against its concentration in ug/mL and a calibration curve using the best fit straight line through the points is obtained. 3.7.2. Sample concentration is calculated from the calibration curve. 3.7.3. The air concentration of Lasso for a sample is calculated by the following equation:
3.8. Safety Precautions 3.8.1. Confine the use of solvents to a fume hood. 3.8.2. Wear safety glasses in all laboratory areas. 4. Recommendations for Further Study
FIGURE 1. DETECTION LIMIT
FIGURE 2. CALIBRATION CURVE
5. References
5.2. Njagi, G.DE; Gopalan, H.NB.; Cytologia 46(1-2) pp. 5.3. Ovellette, R.P. Chemical Week Pesticides Register. 1977, pp. 265. |