SULPROFOS
Carcinogen And Pesticide Branch 1. General Discussion
This evaluation was undertaken because OSHA recently adopted the sulprofos TLV as a PEL. The OVS-2 sampling tube was tested as an effective sampling device for sulprofos. This method follows the procedure developed for other organophosphorus pesticides. (Ref. 5.1.) 1.1.2. Toxic effects (This section is for information only and should not be taken as the basis of OSHA policy.)
The toxicity of sulprofos is dependent on the species tested. The oral LD50 in rats is 1.1.3. Potential workplace exposure Sulprofos is used as a selective insecticide. There was no information available on the number of workers exposed to sulprofos. 1.1.4. Physical properties (Ref. 5.2. to 5.5.)
1.2. Limit defining parameters The detection limit of the analytical procedure, including a 28:1 split ratio, is 0.3 ng per injection. This is the amount of analyte which will give a peak whose height is approximately five times the baseline noise. (Figure 1.) 2. Sampling Procedure
2.1.2. OVS-2 tubes, which are specially made 13-mm o.d. glass tubes that are tapered to 6-mm o.d., packed with two sections of cleaned XAD-2 adsorbent and a 13-mm diameter glass fiber filter. The sampling section and backup section contain 270 and 140 mg respectively. The backup section is retained by two foam plugs and the sampling section is between a foam plug and the glass fiber filter. The glass fiber filter is held next to the sampling section by a polytetrafluoroethylene (PTFE) retainer. (Figure 2.) 2.2. Reagents No sampling reagents are required. 2.3. Sampling technique
2.3.2. Attach the small end of the tube to the sampling pump with flexible tubing. 2.3.3. Attach the tube vertically in the employee's breathing zone in such a manner that it does not impede work performance. 2.3.4. After sampling for the appropriate time, remove the tube and seal it with plastic caps. 2.3.5. Wrap each sample end-to-end with an OSHA seal (Form 21). 2.3.6. Record the air volume for each sample, and list any possible interferences. 2.3.7. Submit at least one blank for each set of samples. Handle the blank in the same manner as the samples, except no air is drawn through it. 2.3.8. Submit bulk samples for analysis in a separate container. Do not ship with air samples. 2.4. Desorption efficiency (glass fiber filter and XAD-2 adsorbent) Six vials each containing a 13-mm glass fiber filter and 270-mg of XAD-2 adsorbent were each liquid spiked on the glass fiber filter with 23 µL of a 10.648 mg/mL sulprofos standard and allowed to dry overnight in a drawer at ambient temperature. These samples were each desorbed with 2.0 mL of toluene, shaken for 30 min and analyzed as in Section 3. The results are listed in Table 2.4.
Desorption Efficiency
2.5. Retention efficiency Eighteen OVS-2 tubes were each liquid spiked with 23 µL of a 10.648 mg/mL sulprofos standard on the glass fiber filter. These were allowed to dry overnight and then 240 L of humid air (~81% relative humidity) were drawn through each tube at 1 L/min. Six of the tubes were each desorbed with 2.0 mL of toluene, shaken for 30 min and then analyzed as in Section 3. The results are listed in Table 2.5. The remaining samples were stored, 6 in a drawer at ambient temperature and 6 in a freezer.
Retention Efficiency
2.6. Sample storage After 7 days of storage, 6 tubes, 3 from the ambient storage group and 3 from the freezer storage group, were each desorbed with 2.0 mL of toluene, shaken for 30 min and then analyzed as in Section 3. The remaining tubes were desorbed and analyzed after 9 days of storage. The results are given in Tables 2.6.1. and 2.6.2.
Ambient Storage
Freezer Storage
2.7. Recommended air volume and sampling rate
2.7.2. The recommended flow rate is 1.0 L/min. 2.8. Interferences (sampling) It is not known if any compounds will interfere with the collection of sulprofos. Any suspected interferences should be reported to the laboratory. 2.9. Safety precautions (sampling)
2.9.2. Follow all safety practices that apply to the work area being sampled. 3. Analytical Procedure
3.1.2. A mechanical shaker. 3.1.3. A GC equipped with an FPD using a sulfur filter. A Hewlett-Packard (HP) 5890 equipped with an autosampler was used in this evaluation. 3.1.4. A GC column capable of separating sulprofos from any interferences. A 30 m × 0.32 mm i.d. (1.0 µm film) DB-5 capillary column was used in this evaluation. 3.1.5. An electronic integrator, or some other suitable means for measuring detector response. The Waters 860 Laboratory Data System was used in this evaluation. 3.1.6. Volumetric flasks and pipets. 3.1.7. Vials, 2-mL. 3.2. Reagents
3.2.2. Sulprofos, reagent grade. A standard obtained from EPA (EPA # 6350, 95.5% purity) was used in this evaluation. 3.3. Standard preparation Prepare sulprofos stock standards by weighing 10 to 15 mg of sulprofos. Transfer the sulprofos to separate 10-mL volumetric flasks, and add toluene to the mark. Make working range standards of 4.2 to 266 µg/mL by diluting of the stock standards with toluene. Store stock and diluted standards in a freezer. 3.4. Sample preparation
3.4.2. Add 2.0 mL of toluene to each vial and seal with a Teflon-lined cap. 3.4.3. Shake the vials for 30 minutes on a mechanical shaker. 3.4.4. If necessary, transfer the samples to 2-mL vials for use on an HP autosampler. 3.5. Analysis
3.5.2. Chromatogram (Figure 3.) 3.6. Interferences (analytical)
3.6.2. GC conditions may generally be varied to circumvent interferences. 3.6.3. Retention time on a single column is not proof of chemical identity. Analysis by an alternate GC column or detector, high performance liquid chromatography (HPLC) and confirmation by mass spectrometry are additional means of identification. 3.7. Calculations
3.7.2. Determine the µg/mL of sulprofos in both sections of each sample and blank from the calibration curve. 3.7.3. Blank correct each section by subtracting the µg/mL found in each blank section from the µg/mL found in each corresponding sample section and then add the values together. 3.7.4. Determine the air concentration by using the following formula.
3.8. Safety precautions (analytical)
3.8.2. Avoid skin contact with all solvents. 3.8.3. Wear safety glasses at all times. 4. Recommendation for Further Study
4.2. A GC having an FPD with a phosphorus filter may yield better sensitivity.
![]()
Detection Limit Chromatogram of Sulprofos with Another Compound ![]()
OVS-2 Sampling Tube ![]()
Chromatogram of Sulprofos with Another Compound ![]()
Calibration Curve 5. References
5.2. Registry of Toxic Effects of Chemical Substances 1985-86 Edition; DHHS(NIOSH) Publication No. 87-114, U.S. Department of Health and Human Services: Cincinnati, OH, 1987; p 3415. 5.3. Farm Chemicals Handbook; Berg, Gordon L. Ed.; Meister: Willoughby, Ohio, 1989; p C45. 5.4. Merck Index, 10th ed.; Windholz, Martha ED.; Merck: Rahway, N.J., 1983; p 1292. 5.5. Documentation of Threshold Limit Values and Biological Exposures Indices; American Conference of Governmental Industrial Hygienists Inc., Fifth Edition 1986, p 547. |