(ATOMIC ABSORPTION)
* Alternate air volumes may be necessary to achieve good analytical sensitivity.
Division of Physical Measurements and Inorganic Analyses OSHA Salt Lake Technical Center Salt Lake City, Utah 1. Introduction This method can determine the amount of specific metal and metalloid particulates in the workplace atmosphere. The airborne particulates are collected on filters using calibrated sampling pumps. These samples are then analyzed using flame atomic absorption or emission spectrometry. This method can also determine specific metals and metalloids contained in wipe and bulk samples. The identification and quantification of the particulate is directly determined as the element. The elements are:
For some analytes, there are alternate methods or procedures which may be more sensitive, accurate, or specific. When a separate OSHA method or procedure exists, that method shall take precedence over this method unless special circumstances render it inapplicable. Elements or compounds having alternate methods or stopgap procedures are:
Depending on advances in technology or changes in exposure limits, substances may be added or deleted from the above lists.
Air and wipe samples containing metal and metalloid particulate have always been analyzed at the OSHA Salt Lake City Analytical Laboratory using atomic absorption or emission spectrometry (8.1.). Constituents in bulk samples have been determined semi-quantitatively using this technique. 1.2. Principle Air samples of the workplace are taken using calibrated sampling pumps with cassettes containing either mixed cellulose ester (MCE) or polyvinyl chloride (PVC) filters. These samples are prepared in the laboratory using concentrated (conc.) acids or extracted with deionized water if a soluble fraction is required. The sample solution is diluted to a known volume after any necessary matrix modifiers are added. The sample is then aspirated into the flame of an atomic absorption or emission spectrophotometer (AAS or AES) and the molecules in the sample solution are subjected to the following processes:
2) desolvation 3) liquefaction 4) vaporization 5) atomization 6) excitation (atoms converted from "ground" to excited state) 7) ionization The absorption or emission of light occurring during processes 5 and 6 is then measured at the characteristic wavelength for the element of interest. For absorption, a hollow cathode lamp or an electrodeless discharge lamp (EDL) is used as the light source. A double beam spectrophotometer is normally used where the lamp radiation alternately passes through and around a flame into which the sample is being aspirated. The sample is atomized and the metal or metalloid atoms absorb light from the source at their characteristic wavelengths. This absorption is proportional to the concentration of the element present in the sample solution. A monochromator isolates the characteristic radiation of the element being analyzed. A photosensitive device then measures the intensity of the transmitted radiation from the two light paths to determine the amount of absorbance occurring in the flame. For emission, a light source is not used. The sample is introduced into the flame, atomized and excited, and then the light emission from excitation is isolated and measured. The intensity of the light emitted is proportional to the concentration of the element present. The following flames are used in this method for absorption or emission:
b) Nitrous oxide/Acetylene mixture (N2O/C2H2) c) Air/Hydrogen mixture (Air/H2) The use of a specific flame is dependent on the respective element's analytical stability, sensitivity, and interferences. 1.3. Advantages and Disadvantages
1.3.1. This analytical method is specific for the element to be determined and does not
distinguish different compounds. When an analysis for a compound is requested, an
elemental analysis is performed on the sample. A gravimetric factor is then applied
to calculate the compound value (Note: For some compounds, additional analytical
procedures (i.e. ion chromatography or 1.3.2. The analysis will also not differentiate between different particle size ranges, such as dusts and fumes. 1.3.3. Metallic analytes having Permissible Exposure Limits (PELs) designated as the soluble form (i.e. iron soluble salts, nickel, etc.) can be analyzed using this method. Samples for soluble analytes are extracted with deionized water and an elemental analysis is performed on the extract. 1.3.4. Some compounds may not dissolve using the digestion procedures described herein. In these cases, an alternative digestion method should be used. 1.3.5. Several elements can be determined from the same filter sample using this method; however, digestion procedures may solubilize only certain metals. If a combination of metals is requested on the same filter, all of the metals must be soluble in the digestion procedure used. 1.3.6. The equipment used is inexpensive and does not require specialized training. 1.4. Use of Metal and Metalloid Compounds in Industry Metals, their alloys, and compounds are used in a wide variety of industries. In certain operations (e.g., welding, smelting, grinding, etc.), particulate matter containing metals and their compounds may be released into the workplace atmosphere. These substances pose a potential health hazard to workers exposed to them (8.2.-8.4.). Further documentation regarding industrial use, toxicity, and physical properties may be found in NIOSH criteria documents for the particular substance. 2. Analytical Range and Sensitivity This method uses detection limit, linearity, and sensitivity terms which are characteristic of atomic absorption. These terms are further defined in Appendix A. Any detection limits, linear ranges, and sensitivities mentioned in this method are for analyses using the primary analytical wavelength, a flow spoiler, an Air/C2H2 flame, and a hollow cathode lamp unless otherwise noted.
2.2. The upper linear range for each element is also given in Table 2. These ranges were taken from reference 8.6. Instrument response is linear to greater concentrations if an alternate wavelength is used; however, the detection limit may also increase. Samples can be diluted to bring the concentration of the element(s) within the linear range. The upper linear range for most elements is usually found near 0.25 to 0.30 absorbance units (ABS). 2.3. The sensitivity for each element is also listed in Table 2. These values are for a nebulizer which has been optimized to give an ABS of 0.25 for an aqueous solution containing 5 µg/mL Cu (8.6.). The actual sensitivity obtained will depend on the particular instrument and flame used, the sample matrix, and instrument operating parameters. 3. Method Performance - Precision and Recoveries Listed in Table 3 are data compiled from quality control (QC) samples which were spiked with aqueous solutions of various analytes and then analyzed in single blind tests. Each analyte was spiked onto an individual MCE filter, allowed to dry, and then prepared and analyzed along with survey samples previously taken by industrial hygienists. These samples were analyzed from 1986 to 1989. Due to the limited number of survey samples received for a few substances, QC samples were not prepared and analyzed for all analytes included in this method. 4. Interferences Interferences occur at the analytical level and can be characterized as chemical, matrix, ionization, spectral, or as background absorption.
4.2. Matrix interferences occur when the physical characteristics (viscosity, surface tension, etc.) of the sample and standard solutions differ considerably. This may occur when samples contain large amounts of dissolved salts or acid, when different solvents are used for samples and standards, or when the temperatures of samples and standards are appreciably different. To control this, samples and standards must be matrix matched, or the sample must be diluted until any matrix effect becomes insignificant. 4.3. Ionization interferences occur when the flame temperature is sufficiently high to ionize the atoms of interest. This changes the absorption spectrum of the analyte and effectively removes atoms from the flame, causing a loss of sensitivity. Ionization interferences are controlled by adding large amounts (usually >0.1%) of an easily ionized metal such as Na, K, Cs, or rubidium (Rb). The excess electrons released in the flame greatly reduces the degree of ionization of the metal being determined. 4.4. Spectral interferences occur when an element other than the one analyzed absorbs at the same wavelength. This causes a positive bias in the results obtained when the interfering element is present in the samples. In this case, an alternate line should be used. Spectral interferences also occur when a multielement hollow cathode lamp is used which contains elements with absorbing wavelengths close to one another and the analytical slit width used is wide enough to allow the wavelengths of more than one element to pass. If the sample contains two or more of these elements, a positive bias will occur. To resolve this, a single element lamp, an alternate wavelength, or in certain cases, a narrower slit width can be used. 4.5. Background absorption interferences include flame absorption, molecular absorption, and light scattering:
b) Molecular absorption is controlled by using hotter flames to break down molecular species or by DABC. c) Light scattering occurs at shorter wavelengths when samples have a large salt content; this is controlled using DABC. 4.6. Large amounts of silicates or other particulates may interfere and may also cause aspiration problems (8.7.). If present, they should be removed by filtration. The particulate should then be re-digested and analyzed to ensure the analyte(s) of interest have been completely extracted.
4.7. This analytical method is normally not 4.8. Potential interferences for several of the elements determined by this method are listed in Appendix B. 5. Sampling 5.1. Equipment - Air Filter Samples
5.1.1. Mixed cellulose ester (MCE) filters (0.8 µm pore size), cellulose backup pads, and
cassettes, 5.1.2. Gel bands (Omega Specialty Instrument Co., Chelmsford, MA) for sealing cassettes. 5.1.3. Sampling pumps capable of sampling at 2 liters per minute (L/min). 5.1.4. Assorted flexible tubing. 5.1.5. Stopwatch and bubble tube or meter for pump calibration. 5.2. Equipment - Wipe Samples
5.2.1. Smear tabs (part no. 5.2.2. Deionized water.
5.2.3. Scintillation vials, 5.3. Equipment - Bulk Samples 5.3.1. High-volume sampling pump with appropriate sized MCE collection filters.
5.3.2. Scintillation vials, 5.4. Sampling Procedure - Air Filter Samples
5.4.1. Place a MCE filter and a cellulose backup pad in each
5.4.2. Calibrate each personal sampling pump with a prepared cassette 5.4.3. Attach prepared cassettes to calibrated sampling pumps (the backup pad should face the pump) and place in appropriate positions on the employee or workplace area. 5.4.4. Collect the samples at approximately 2 L/min for the recommended sampling times (unless otherwise noted):
The analytical sensitivity of a specific analyte may dictate the use of a different sampling time.
* When determining compliance with the Ceiling PEL for sodium hydroxide, take
5.4.5. Place plastic end caps on each cassette after sampling. Attach an 5.5. Sampling Procedure - Wipe Samples Certain analytes may have a skin designation (See Table 1). 5.5.1. Wear clean, impervious, disposable gloves when taking each wipe sample. 5.5.2. Moisten the wipe filters with deionized water prior to use. 5.5.3. If possible, wipe a surface area covering 100 cm². 5.5.4. Fold the wipe sample with the exposed side in.
5.5.5. Transfer the wipe sample into a 5.6. Sampling Procedure - Bulk Samples 5.6.1. In order of laboratory preference, bulk samples may be one of the following:
2) a representative settled dust (rafter) sample, 3) a sample of the bulk material in the workplace.
5.6.2. If possible, transfer the bulk material or filter into a 5.7. Shipment 5.7.1. Submit at least one blank sample with each set of air or wipe samples. Blank filter samples should be handled in the same manner as other samples, except that an air or wipe sample is not taken.
5.7.2. The type of bulk sample should be stated on the OSHA 91A and 5.7.3. Send all samples to the laboratory with the OSHA 91A paperwork requesting the specific analyte(s) of interest. If analysis of a mixture of different elements or compounds is necessary, contact the lab to ascertain which analytes can be analyzed together. 6. Analysis 6.1. Safety Precautions 6.1.1. Care should be exercised when handling any acidic solutions. Acid solution contact with work surfaces should be avoided. If any acid contacts the eyes, skin, or clothes, flush the area immediately with copious amounts of water. Medical treatment may be necessary. 6.1.2. All work with conc. acids is potentially hazardous. Always wear safety glasses and protective clothing. Prepare all mixtures, samples, or dilutions in an exhaust hood. To avoid exposure to acid vapors, do not remove any beakers from the hoods until they have returned to room temperature. 6.1.3. Extra care should be used when handling perchloric acid (HClO4). Perchloric acid should only be used in a hood that has been approved for HClO4 use. In this hood:
b) A water washdown system for the ducts and work surface must be installed and periodically used. c) Precautions should be taken to ensure that explosions or spontaneous ignition of sample material from HClO4 is prevented. Working with HClO4 is very hazardous. Be sure to wear safety glasses, a labcoat, and gloves. Always add nitric acid (HNO3) with HClO4. When digesting backup pads or other samples with HClO4, watch them carefully since there is a chance they could ignite. Always keep HNO3 nearby when using HClO4. In the event of sample media ignition, quickly douse the sample with a small portion of HNO3. 6.1.4. Care should be exercised when using laboratory glassware. Chipped pipettes, volumetric flasks, beakers, or any glassware with sharp edges exposed should not be used. 6.1.5. Pipetting is always performed using an automatic pipet or pipette bulb, never by mouth. 6.1.6. Before using any instrument, the operator should consult the Standard Operating Procedure (SOP) (8.8.) and any instrument manuals. 6.1.7. Since metallic elements and other toxic substances are vaporized during flame operation, it is imperative that an exhaust hood is installed and used directly above the burner chamber of the spectrometer. Always ensure the exhaust system is operating before proceeding with the analysis. 6.2. Equipment 6.2.1. Atomic absorption spectrophotometer consisting of a(an): Nebulizer and burner head. Pressure-regulating devices capable of maintaining constant oxidant and fuel pressures. Optical system capable of isolating the desired wavelength of radiation. Adjustable slit. Light measuring and amplifying device. Display, strip chart, or computer interface for indicating the amount of absorbed or emitted radiation. Deuterium Arc Background Corrector. This is usually required for determinations at short (<250 nm) wavelengths. Light source for absorption:
b) Electrodeless Discharge Lamp (EDL) for the specific element. This type of lamp may provide better sensitivity and detection limits for some elements, especially Se, Sn, and Sb. If used, a separate EDL power supply is usually necessary. 6.2.2. Oxidant: Compressed, filtered air free from water, oils and other contaminants. 6.2.3. Nitrous oxide (N2O). 6.2.4. Fuel (Use flash arrestors when using flammable gases. Consult with the manufacturer for appropriate use.):
CAUTION: Do not use grades of acetylene that contain solvents other than acetone. These solvents may damage PVC tubing in some instruments. Do not use acetylene when the tank pressure drops below 520 kPa (75 psi). b) Hydrogen is used as the fuel in the determination of certain elements. 6.2.5. Pressure regulators, Two-stage. 6.2.6. Flash arrestors (model 6103, Matheson Gas Products, East Rutherford, NJ). 6.2.7. Glassware
b) Volumetric flasks, Class A: c) Pipettes, Class A: Assorted sizes 6.2.8. Forceps. 6.2.9. Exhaust hood and hotplate, or microwave digestion system (model no. MDS-81, CEM Corp., Matthews, NC).
6.2.10. Filtering apparatus consisting of MCE filters, 0.45-µm pore size, 6.2.11. Analytical balance (0.01 mg). 6.3. Reagents (All chemicals should be reagent grade or better. Many of the chemicals listed below are only used in specific instances. Specific reagents are listed within the additional procedures in Table 4 and also in Table 5.) 6.3.1. Deionized water (DI H2O) with a specific conductance of less than 10 µS. 6.3.2. Ammonium fluoride (NH4F) solutions (used for specific insoluble compounds, see AP 6, Table 4).
b) Ammonium fluoride, 0.1 M in 4% HNO3: Carefully add 40 mL conc. HNO3 and 100 mL of the 1 M NH4F solution to 500 mL DI H2O and dilute to 1 L in a polyethylene volumetric flask. Store in a polyethylene bottle since acidic solutions of NH4F may form small amounts of HF and etch glass containers. 6.3.3. Hydrogen peroxide (H2O2), 30% (used for digestions of Cr, see AP 5, Table 4). 6.3.4. Mineral acids (used for digestions) CAUTION: Refer to Section 6.1.2. before using acids.
b) Hydrofluoric acid (HF), conc. (49%). c) Nitric acid (HNO3), conc. (69 to 71%). d) Perchloric acid (HClO4), conc. (69 to 72%). Please see Section 6.1.3. before using HClO4. e) Sulfuric acid (H2SO4), conc. (95 to 98%). f) Acid mixture for platinum digestions: Prepare a mixture of HCl/HNO3 by slowly and carefully adding 82 mL conc. HCl to 18 mL conc. HNO3 (CAUTION: Do not store this solution; dispose of properly after use). 6.3.5. Mineral acids (used for dilutions or cleaning glassware) CAUTION: Refer to Section 6.1.2. before using acids.
b) Nitric acid, 4% v/v: Carefully add 40 mL conc. HNO3 to 500 mL DI H2O and dilute to 1 L. c) Nitric acid 10% v/v: Carefully add 100 mL of conc. HNO3 to 500 mL of DI H2O and then dilute to 1 L. d) Nitric and hydrochloric acid v/v mixture (4% HNO3 / X% HCl, where X% is listed below): Carefully add the appropriate amount of conc. HCl to 500 mL of DI H2O:
Then carefully add 40 mL conc. HNO3 and dilute to 1 L with DI H2O. 6.3.6. Chemical or ionization interference suppressants
b) Potassium ion, 5,000 µg/mL: Dissolve 9.54 g potassium chloride (KCl) in DI H2O and dilute to 1 L. c) Sodium ion, 5,000 µg/mL: Dissolve 12.71 g sodium chloride (NaCl) in DI H2O and dilute to 1 L. 6.3.7. Stock standard solutions Commercially available aqueous standards are used. Expiration dates for standards should be followed. If there is no expiration date, dispose of after 1 year. As an alternative, standards can be prepared using the procedures described in the SOP (8.8.) or instrument manufacturer manuals (i.e. 8.6., 8.9., 8.10.). 6.4. Glassware Preparation
6.4.1. Place the Phillips beakers in an exhaust hood and add approximately 10 mL of a 1:1
HNO3/DI H2O mixture in each 6.4.2. Rinse all volumetric flasks with 10% v/v HNO3 and then rinse thoroughly with DI H2O. 6.5. Working Standards 6.5.1. Dilute stock standard solutions to the appropriate ranges using a diluent that will match the sample matrix. Use information in Tables 1 and 2 as guides for the ranges; use Table 5 for matrices. The standard concentrations should bracket the expected sample concentrations and the standard/sample matrices should match. 6.5.2. Store standards in appropriate containers. Protect Ag standards from light by storing them in actinic or brown plastic bottles. Store standards containing NH4F in polyethylene containers. 6.6. Sample Preparation
Note: Always prepare blank samples with every sample set. Prepare an additional blank media sample any time an extra procedure is used (i.e. wiping out the particulate contained inside a cassette with an MCE filter or preparing a contaminated backup pad). This blank media should be from the same manufactured lot as the prepared filter or backup pad.
1. Preparation of air and wipe samples
Open the filter cassette or scintillation vial, carefully remove the sample filter with forceps,
and place in a labeled Phillips beaker. Use If the backup pad appears to be discolored, it may be due to leakage of air around the filter during sampling. In these cases, the pad should also be prepared and analyzed. Place the backup pad in a separate beaker. 2. Preparation of bulk samples Review any available material safety data sheets to determine safe bulk handling. The safety data may also offer a clue as to the aliquot amount needed for adequate detection of the element(s) of interest. Measure by volume or weight an appropriate aliquot of any liquid bulk sample. Weigh the appropriate amount of any solid bulk sample.
Note: Aliquot amounts of bulks are dependent on the analytical sensitivity, detection limit,
and solubility of the material used. If uncertain, a
After measuring, transfer the aliquot to a 6.6.1. Extraction or digestion - all samples Consult Tables 4 and 5 to determine the reagents used during extraction or digestion for each element to be analyzed. Some elements (Ba, Sn, etc.) or compounds are not digested with conc. HNO3, but are prepared using alternate procedures (APs) listed in Table 4. These elements or compounds and their AP numbers are:
For the element or compounds listed above, follow the APs recommended and then proceed with Section 6.6.2. For other elements or compounds, follow the procedures a, b, or c listed below:
Place the beakers in an exhaust hood and add 3 to 5 mL conc. HNO3 to cover the filter. Place the beakers on a hot plate and heat the samples until about 1 mL remains. Add a second portion of approximately 1 to 2 mL of conc. HNO3. Apply heat until the appropriate amount of HNO3 remains in the beaker (1 mL of HNO3 will give a 4% HNO3 matrix when diluted to 25 mL final volume). b) Large wipe, PVC filters, or backup pads Place the beakers in an exhaust hood and add the following amount of conc. HNO3 to the beakers: Large wipes and backup pads 10 to 15 mL PVC filters 3 to 5 mL Place the beakers on a hot plate and heat the samples until about 1 mL remains. Add 2 mL of conc. HClO4 along with a second portion of 2 mL HNO3, heat the sample, and then remove when about 1 mL remains. (Note: Please see Section 6.1.3. before using HClO4.) As an alternative, an extraction of the backup pad or wipe sample using only HNO3 may be used. Add HNO3 to the media, digest on a hotplate, and continue to add HNO3 until the solution becomes clear. Remove the beaker from the hotplate when the appropriate amount of HNO3 remains. c) Bulk samples Add 10 to 30 mL HNO3, place the beaker on a hot plate, and digest the bulk sample until the material dissolves and the appropriate amount of solution remains (about 1 mL if diluting samples to 25 mL, 2 mL if 50 mL final volume, etc. After dilution this will give a final volume of 4% HNO3). If necessary, use other acids, or use a microwave digestion system to facilitate digestion [For further information regarding microwave digestion, see the Standard Operating Procedure (8.11.)]. 6.6.2. Filtration - all samples
Samples extracted with DI H2O should normally be filtered. If particulate is present, filter the extract through a 0.45-µm MCE filter. Save the extract as the soluble portion. If necessary, digest the particulate on both filters using procedure (a) above or the applicable AP to prepare the remaining insoluble material for additional analyses. To control for potential contamination, prepare blank samples in the same fashion as the filtered samples. 2) Samples Previously Digested: If particulate matter is present after digesting, cool the sample, add approximately 10 mL DI H2O, then filter the solution through a 0.45-µm MCE filter. Save the filtrate. Repeat digestion procedure (a) above for the filter containing the particulate. 6.6.3. Dilution - all samples Allow all digested samples to cool to room temperature in an exhaust hood before proceeding. Additional sample or filtrate treatment may be required for certain elements. Perform any special sample treatments recommended in Table 5, and then quantitatively transfer each sample and each filtrate solution to individual volumetric flasks. Add any reagents necessary to achieve the final solution concentrations listed in Table 5 for specific analytes. Dilute to volume with DI H2O and then mix well. Solution volumes are dependent on the following factors:
b) The detection limit of the analytical method. c) The PEL of the analyte. d) The number of analytes requested. Air samples are normally diluted to 25 mL unless one or more of the above factors suggests an alternate volume should be used. For routine analysis, at least 1/10 of the OSHA PEL should be detectable. Final solution volumes can be estimated using the following equation:
Where:
Quantitative detection limits are listed in Table 2.
The FV factor assists in determining the final volume. Sample solution volumes
normally used are: FV Factor = 11.1
and a final volume should be 6.7. Instrument Setup and Analysis
6.7.1. Set up the AAS or AES according to the SOP (8.8.) or the manufacturer's
instructions. Use the flame and wavelength recommended in Table 7. If alternate
conditions are necessary, consult the instrument manufacturer's manual for other
settings and operating procedures. Install an EDL or hollow cathode lamp for the
element of interest and allow it to warm up for 6.7.2. Aspirate and measure the ABS of a standard solution for the element of interest. The standard concentration should be within the linear range for the element. Compare the ABS to an expected sensitivity value (Note: Some values are listed in Table 7; these were adapted from reference 8.6. or obtained at the OSHA laboratory). Then aspirate the smallest standard to be used and assure the ABS reading is above the background level of the instrument. 6.7.3. Make any adjustments necessary for the particular analysis, such as: scale expansion, burner head rotation, background correction, or alternate wavelength. 6.7.4. Aspirate and measure the ABS of a prepared standard solution, then determine the baseline by aspirating DI H2O and measuring the ABS. 6.7.5. Analyze standards, samples, and blanks. Repeat the baseline determination after each solution is analyzed. The baseline readings will assist in correcting any instrument drift. If more than one solution has been prepared for a sample (i.e. filtrate and sample, or soluble and insoluble portions), analyze each for all requested elements. Standards must bracket the sample concentrations. Analyze a standard after every four or five samples. Standard readings should be within 10 to 15% of the readings obtained at the beginning of the analysis. 6.7.6. If any samples exceed the linear range, they should be diluted. When diluting a sample, be sure that the diluted sample has the same matrix as the original sample and standards. If a number of samples must be diluted, it may be more advantageous to use a less sensitive wavelength. 6.8. Analytical Recommendations 6.8.1. When a fresh standard is prepared, analyze the old and new standards and compare results to verify the new standard is correct. If two or more stock solutions are available for working standard preparations, rotate the preparation from one stock solution to the next to verify the quality. 6.8.2. Keep a permanent record of all standard preparation and comparison data. Assign and follow expiration dates for all standards. 6.8.3. Always analyze blank samples along with the other samples. Treat blanks in the same fashion as samples, including any filtration steps.
6.8.4. When analyzing for Ag,
6.8.5. In this method, many different matrices are used to digest and keep analytes in
solution. Occasionally, during multiple element analysis of the same sample, matrix
effects can occur if standards are not 7. Calculations 7.1. Subtract each baseline ABS from the corresponding standard ABS, and plot the net ABS versus the standard concentrations. Using a least squares method, determine the equation for the best curve fit. 7.2. Subtract each baseline ABS from the corresponding sample or blank ABS, and use the standard curve to calculate the concentration of each analyte in µg/mL. 7.3. Calculate the concentration for each air sample as:
Where:
7.4. For wipe or bulk samples, calculate the total amount (in µg) of analyte in each sample using the equation above. An air volume is not used. Convert bulk sample analytes to % composition using:
Where:
7.5. Reporting Results to the Industrial Hygienist For those samples only extracted with DI H2O, report the sample results as the soluble fraction of the sample. If more than one solution exists for a sample, and it is not necessary to report results separately, then combine these results. An example is a sample that was filtered due to insoluble particulate. The results from the filtrate plus results from the second particulate digestion are added together. 7.5.1. Report air sample results as mg/m3 analyte. 7.5.2. Report wipe sample concentrations as total micrograms or milligrams analyte. 7.5.3. Report bulk sample results as approximate percent by weight analyte (note: Sample results for bulk liquids may be reported as approximate percent by volume if volumetric aliquots were taken during sample preparation.) Due to differences in sample matrices between bulks and standards, bulk results are approximate. 8. References
8.1. Occupational Safety and Health Administration Analytical Laboratory: OSHA Manual of
Analytical Methods edited by R.G. Adler (Method No. 8.2. Clayton, G.D. and F.E. Clayton, ed.: Patty's Industrial Hygiene and Toxicology. 3rd ed. New York: John Wiley and Sons, 1978. 8.3. American Conference of Governmental Industrial Hygienists: Documentation of the Threshold Limit Values and Biological Exposure Indices. 5th Ed. Cincinnati, OH: American Conference of Governmental Industrial Hygienists, 1986. 8.4. National Institute for Occupational Safety and Health: The Industrial Environment--Its Evaluation and Control. Washington, DC: Government Printing Office, 1973.
8.5. Slavin, S., W.B. Barnett, and H.L. Kahn: The Determination of Atomic Absorption Detection
Limits by Direct Measurement. Atomic Absorption Newsletter 11:
8.6. Perkin-Elmer Corp.: Analytical Methods for Atomic Absorption Spectrophotometry. Norwalk,
CT: 8.7. National Institute for Occupational Safety and Health: NIOSH Manual of Analytical Methods. 2nd ed. (Method no. 173) Cincinnati, OH: National Institute for Occupational Safety and Health, 1977. 8.8. Occupational Safety and Health Administration Technical Center: Standard Operating Procedure for Atomic Absorption. Salt Lake City, UT. In progress (unpublished). 8.9. Fisher Scientific Company: Atomic Absorption Methods Manual. Waltham, MA: Fisher Scientific Co., 1977. 8.10. Instrumentation Laboratory Inc.: Atomic Absorption Methods Manual. Wilmington, MA: Instrumentation Laboratory Inc., 1975. 8.11. Occupational Safety and Health Administration Analytical Laboratory: Standard Operating Procedure for Microwave Digestions by D. Cook. Salt Lake City, UT. 1989 (unpublished).
8.12. "Air Contaminants; Final Rule": Federal Register 54:12 (19 Jan. 1989). pp. 8.13. Occupational Safety and Health Administration Analytical Laboratory: OSHA Laboratory Quality Control Division Data by B. Babcock, Salt Lake City, UT, 1989 (unpublished). 8.14. Slavin, Walter: Atomic Absorption Spectroscopy. New York: Interscience Publishers, 1968.
8.15. Ediger, R.D.: Atomic Absorption Analysis with the Graphite Furnace using Matrix
Modification. Atomic Absorption Newsletter. 14(5): Air Contaminants - OSHA Permissible Exposure Limits*
Note: Compounds having total and respirable dust PELs of 15 and 5 mg/m3, respectively, are normally analyzed
gravimetrically. Elements contained in these dust samples can be identified by this or other methods, if
necessary.
Detection Limits, Sensitivities, and Ranges
* DL = Detection Limit. See Appendix A for more information regarding definitions or calculations. Analytical DLs are approximate. ** Alternate line of 231.2 nm was used with one exception: The qualitative detection limit value is for the primary line (217.6 nm). *** Air/H2 flame used with the exception of the qualitative detection limit determination. This value is for Air/C2H2 flame. + N2O/C2H2 flame used.
++ Flame emission used to determine qualitative detection limit.
Precision and Accuracy*
CV Coefficient of Variation
Alternate Procedures
1) Digest samples with HNO3. Heat until the liquid is nearly gone. Allow the samples to cool to room temperature. 2) For 25 mL final sample solution volumes, add the following amount of conc. HCl (Adjust accordingly for alternate solution volumes):
3) Warm gently and swirl to dissolve the analyte. Allow samples to cool and dilute to a
1) Place the sample in a beaker and add an aliquot of 2) Place the beaker in an ultrasonic bath for approximately 10 min.
3) Filter the sample through a 0.45 µm MCE filter and transfer the filtrate to a 4) Add reagents to achieve the final solution concentrations listed:
(1) Digest the sample using the procedure described in Section 6.6.1.a. (2) Transfer the sample to a volumetric flask. (3) Dilute the samples and add ionization suppressants to achieve the final solution concentrations listed:
(1) For Au, Sn, or SnO, add 9 mL HCl to each beaker, swirl, and then add 2 mL HNO3. CAUTION: Make sure the entire filter or sample is wetted with HCl and allow the filter/HCl solution to sit for a period of at least 2 to 3 min before adding the HNO3. (2) Digest the sample on a hot plate until nearly dry. (3) Allow the samples to cool and then quantitatively transfer the sample, using a small amount of DI H2O to rinse the beaker, to a clean volumetric flask. Dilute to volume, making the final solution 10% HCl. For example, add 2.5 mL conc. HCl to a sample if the total solution volume is 25 mL. (4) Results for either Sn or SnO are reported as total Sn.
For samples requiring analysis of total Cr, the following procedure should be used. This procedure avoids the loss of any Cr(VI) as chromyl chloride (CrO2Cl2). For chromate or chromic acid analysis, see OSHA method no. ID-103.
(1) Digest the samples collected on MCE filters with HNO3 and then allow to cool to room temperature. If PVC filters were used, digest with HNO3 plus 2 mL of HClO4 and then allow to cool. (2) Add 1 or 2 mL of 30% H2O2 to the cooled solution to reduce any Cr(VI) that may be present. Let the sample sit for several minutes. (3) Heat approximately 5 min to boil off the H2O2 and then allow to cool. At this stage HCl may be added if needed to dissolve other metals. (4) Dilute to volume with DI H2O and analyze.
(1) For compounds such as zirconium dioxide or hafnium dioxide, place the sample filter in a platinum crucible, char at 300 °C, then heat the residue at 800 °C in a muffle furnace. [As an alternative, the digestion can be performed using a microwave digestion system (8.11.).] (2) Add 1 to 2 mL of HF, swirl the solution, and then heat on a hot plate to dissolve the residue.
(3) Evaporate the solution to approximately 0.4 mL and then
transfer to a Another procedure can be used for elements which do not need to be converted to their fluoride salts: (1) Heat the HF solution on the hot plate until the liquid is nearly gone. (2) Add 2 to 3 mL HCl, and warm the solution until about 1 mL remains. (3) Quantitatively transfer the solution to a 10 mL volumetric flask and dilute to volume with the appropriate diluents mentioned in Table 5. (4) It is recommended to prepare quality control samples of the substance of concern. Digest the samples and analyze by the same procedure to check the recovery efficiency. For platinum: (1) Place the sample filter in a Teflon microwave digestion vessel and add 5 mL of the "acid mixture (HCl/HNO3) for platinum digestions" prepared in Section 6.3.4., part f.). (2) Digest the sample according to Microwave Digestion Standard Operating Procedure (8.11.) or manufacturer guidelines. (3) Allow the sample to cool and then transfer to a 25-mL volumetric flask. Dilute to volume with DI H2O. * Some Zr compounds, such as the oxide and sulfate, may be insoluble when using the HNO3 digestion (8.6., 8.7.). Hafnium dioxide may also be insoluble.
(1) Place the sample filter in a beaker and desorb with 15 mL of DI H2O for approximately 5 min.
(2) Decant the sample solution into a
For example, add 5 mL of 5,000 µg/mL potassium ion for Cs analysis and dilute to volume with DI H2O. Add 5 mL of 5,000 µg/mL sodium ion for KOH analysis. (3) Analyze by flame emission or atomic absorption.
(1) Digest the filter with 1 mL HNO3 and 2 mL H2SO4 in a Phillips beaker and heat until about 1 mL remains. (2) Quantitatively transfer the solution to a 25-mL volumetric flask, add 5 mL of 5,000 µg/mL potassium ion, then dilute to volume with DI H2O.
Digestion or Extraction Reagents
+ Standards should be prepared in this matrix.
* After completing the digestion with HNO3, add 1 or 2 drops of conc. HCl to facilitate particulate
dissolution.
Gravimetric Factors
Analytical Parameters
* Adapted from reference 8.6. or from laboratory determinations ** Due to the limited upper linear range, samples may have to be diluted, the burner head rotated, or an alternate wavelength used. The burner head is routinely rotated for Fe and Mg before analysis.
1 Air/Acetylene mixture, lean, blue flame 2 Air/Acetylene mixture, very lean, blue flame 3 Nitrous oxide/Acetylene mixture, rich, red flame 4 Nitrous oxide/Acetylene mixture, slightly rich, red flame 5 Air/Hydrogen mixture Terminology For the purposes of this method, the following definitions are used: Qualitative detection limit The concentration (µg/mL) of an element which would yield an absorbance (ABS) equal to twice the standard deviation of a series of measurements of an aqueous solution containing the element. The signal obtained from the aqueous solution must be distinctly greater than the baseline (8.10.). These detection limits were taken from reference 8.5. Analytical detection limit The lowest concentration (µg/mL) of an element that can be reliably quantitated. This limit is the largest value obtained from any of the three calculations:
b) Two times the average baseline variation, or
c) The lowest standard used to construct a
The average reading for this standard is within 20% of it's linear response. The linearity is
determined by the other standards used to construct the If the lowest standard ABS reading is more than 20% in error, then an algorithm is used and the concentration value is increased in 10% increments until a concentration is achieved that would display less than 20% error or until the lowest standard concentration is reached. Sensitivity The concentration (µg/mL) of an element in aqueous solution which will produce an ABS of 0.0044 (8.6.). Linear Range
The working range of a specific analyte. The range is considered linear if doubling the concentration of a
standard results in at least a 75% increase in ABS.
Potential Interferences
|