CONFIRMATION OF VANADIUM PENTOXIDE IN WORKPLACE ATMOSPHERES
Method Number: |
ID-185 |
Matrix: |
Air |
OSHA Permissible Exposure Limits
Final Rule Limits (as V2O5):
Respirable dust:
Fume: |
0.05 mg/m3 Time Weighted Average (TWA)
0.05 mg/m3 (TWA) |
Transitional Limits (as V2O5):
Total dust:
Fume: |
0.5 mg/m3 (Ceiling)
0.1 mg/m3 (Ceiling) |
Sampler: |
Low-ash polyvinyl chloride (PVC) membrane, 37-mm diameter,
5-µm pore size. The membrane is placed in a cassette which is then attached to
a 10-mm nylon cyclone (optional for fume sampling) and a calibrated personal sampling pump. |
Recommended Sampling Rate: |
1.7 liter per minute (L/min) |
Analytical Procedure: |
The PVC membrane is dissolved in tetrahydrofuran and a
suspension is produced with the collected dust. The dust is transferred to a silver membrane for analysis by XRF and
verification by XRD. |
Recommended Air Volume: |
816 liters (8-h sample) For confirmation only |
Quantitative Detection Limits: (as V2O5) |
X-ray Diffraction |
|
X-ray Fluorescence |
DL
|
Total time
|
|
DL
|
Total time
|
25 µg 20 µg |
65 s 650 s |
|
14 µg 2 µg |
100 s 1,000 s |
|
Precision and Accuracy Validation Levels: |
0.29, 0.58, and 0.87 mg/m3 (as V2O5)
using 237, 474, and 710 µg V2O5 loadings
and assuming 816 liter air volumes |
Respirable Dust:
CV1 Mean Recovery Overall Analytical Error |
X-ray Fluorescence (See Section 3.2.)
|
Fine-Respirable 0.097 87.1% ±32% |
Respirable 0.064 96.5% ±16% |
|
Method Classification: |
Validated Confirmation Method |
Chemist: |
Mike C. Rose |
Date: |
October, 1987 (revised April, 1991) |
Commercial manufacturers and products mentioned in this method are for
descriptive use only and do not constitute endorsements by USDOL-OSHA.
Similar products from other sources can b substituted.
Inorganic Methods Evaluation Branch
OSHA Salt Lake Technical Center
Salt Lake City, Utah
1. Introduction
The goal of this method is to provide confirmation for occupational vanadium pentoxide (V2O5)
exposures. To achieve that end, the published X-ray diffraction (XRD) approach presented in a NIOSH
study (8.1.) for various V compounds was evaluated for applicability. Unlike the NIOSH study, this
method focuses on only V2O5. The method was further extended to evaluate the applicability of X-ray
fluorescence (XRF) to measure V2O5 exposures, since sample preparation was also compatible with XRF analysis.
The procedure used in the NIOSH XRD study was adapted from the published analytical procedure
(8.1.) and techniques (8.2.) that were in print prior to the publication of NIOSH Method 7504
(8.3.). The NIOSH sampling approach collects and analyzes only the respirable fraction because of its
toxic effects (8.1.). [Vanadium pentoxide is also toxic by other routes of exposure
(8.4.-8.6.)]
Particle size effects on the analysis were investigated during the OSHA validation when the
OSHA PEL was for total dust and fume (i.e. Transitional PELs). The respirable particle-size range was
used for validation of this method because of the size dependence associated with XRD. The validation
is applicable to the newer Final Rule V2O5 PELs because total dust is now excluded other than
consideration as nuisance dust. The OSHA V2O5 PELs are currently for respirable dust or fume and are
8-h time weighted average (TWA) values (8.7.).
1.1. History
The previous two OSHA methods for V were atomic absorption spectroscopy (AAS) (8.8.) and
inductively coupled plasma atomic emission spectroscopy (ICP-AES) (8.9.).
Both methods analyze for the element V and do not speciate the particular V compound present.
This method was validated in 1987 using a Finnigan direct-beam energy dispersive
X-ray fluorescence (EDXRF) spectrometer and a Philips automated powder diffractometer
(APD). More recently, field samples submitted to the Laboratory for V2O5 analyses have been
analyzed using a Kevex EDXRF spectrometer that features substantially improved software and
an Fe secondary target that reduces background and offers improved analytical performance.
Detection limit and sensitivity data for both the direct-beam and secondary target
instruments are included in this method. A more complete description of the X-ray
instrumentation used in this method is given in reference 8.10.
1.2. Analytical Principles
1.2.1. XRD:
Quantitative powder XRD relies upon the diffraction of monochromatic X-rayS by the
regular lattices of randomly oriented crystallites uniformly deposited in a thin flat film.
The relationship among the parameters of order (n), X-ray wavelength (l),
the lattice spacing (d), and the angle of diffraction (q) is given by the Bragg equation:
n(l) = 2d sin (q)
1.2.2. XRF:
Thin film quantitative X-ray fluorescence relies upon the excitation of atoms in a
uniformly deposited thin flat film by the application of X rays of sufficient energy to cause
the promotion of inner orbital electrons and subsequent decay accompanied by characteristic fluorescence.
1.2.3. XRD and XRF:
For uniform thin films, these X-ray techniques produce signal intensities that are
proportional to the amount of analyte present. Non-linear calibration curves can also be
used to correct for the fluorescing Ag support medium, shadowing of particles in Ag
membrane pores, preferred orientation, and particle-size effects (8.10.).
1.3. Advantages and Disadvantages
1.3.1. Advantages of both methods:
Both methods are non-destructive. Further analyses may be performed on the sample by dissolution
in acid and using atomic absorption (8.8.) or atomic emission (8.9.) methods.
1.3.2. Advantages of XRD:
The XRD method is specific for determining V2O5. The sampling and analytical
approach provides the opportunity for concomitant analysis for respirable quartz and/or
zinc oxide by XRD (8.11. and 8.12.).
1.3.3. Advantages of XRF:
Energy dispersive XRF provides the opportunity for the serendipitous discovery of other toxic
elements that might be present in the air sample. For example, Pb and U are present in many V ores
(8.1., 8.6.).
Detection limits may be significantly decreased for the XRF analysis by increasing the data collection time.
1.3.4. Disadvantages of both:
The XRF and XRD instrumentation are expensive.
1.3.5. Disadvantages of XRD:
Quantitation by XRD is particle-size dependent and the sample distribution should
match that of the standard reference material. This necessitates using a 10-mm nylon
cyclone with a personal pump operated at 1.7 L/min to collect all samples except for
fume. The use of a cyclone will preclude certain particle-size ranges; therefore, if a
cyclone is used, the dust collected will only be in the respirable range. For total dust
samples where a cyclone is not used, the XRD method is used only as a confirmation technique.
The presence of a non-V2O5 interference in some blanks precludes significantly
decreasing detection limits in the XRD method by increasing the data collection time.
Preferred-orientation effects are potentially large for XRD analyses of V2O5,
due to cleavage along favorable planes (8.1.).
1.3.6. Disadvantages of XRF:
X-ray fluorescence (XRF) cannot speciate the vanadium-containing compound. Positive
interferences can occur from Ti or Cr (Also see Section 4.2.).
1.4. Vanadium Pentoxide (CAS 1314-62-1) Some Sources of Exposure (8.6.):
Application
|
|
Source of Exposure
|
catalyst | oxidation of nitrogen and sulfuroxides |
colorant | manufacture of yellow glass |
developer | photography industry |
coating | using welding rods |
alloys | manufacture of special steels |
contaminant | cleaning fuel oil burners |
1.5. Physical and Chemical Properties (8.5., 8.6.):
Molecular weight: | 181.90 |
Specific gravity: | 3.35 |
Melting point: | 690 °C |
Boiling point: | 1750 °C (decomposes) |
Vapor pressure: | non-volatile at room temp. |
Aqueous solublility: | 0.8 g/100 mL at 20 °C |
1.6. Toxicology
Information contained in this section is a synopsis of current knowledge of the physiological
effects of V2O5 and is not intended as a basis for OSHA policy.
1.6.1. When inhaled, the chief effects of V2O5 are on the respiratory passages.
Tracheitis, bronchitis, emphysema, pulmonary edema, or bronchial pneumonia may be observed,
but no specific chronic lung lesions have been described. Other symptoms reported include eye
irritation, conjunctivitis, dermatitis, green tongue, metallic taste, throat irritation,
increased mucus, and cough (8.6.).
1.6.2. The toxic effects of V2O5 are
primarily from exposures to dusts in the respirable particle-size range.
Exposure to non-respirable dust can also produce toxic effects.
1.6.3. Death has been observed when animals were exposed to 70 mg/m3
for a few hours (8.6.).
2. Range, Detection Limit, and Sensitivity (8.10.)
This method was evaluated when the OSHA PELs were Ceiling total dust and fume values. It should
perform well under the more recent 8-h TWA respirable dust and fume OSHA PEL
(8.7.).
2.1. Loadings of 237 to 710 µg V2O5 were
evaluated. The calibration curves are usable up to at least 2,500 µg V2O5,
the highest loading used in the method. Both X-ray methods require a thin layer deposition of sample;
therefore sample loadings greater than 2,500 µg V2O5
should be split up on separate Ag membranes.
2.2. The quantitative detection limits (DL) for V2O5 depend upon the
X-ray data collection time. The following detection limit estimates were obtained:
| | | X-ray Fluorescence
|
X-ray Diffraction
| |
Direct-Beam
| |
Secondary Target
|
DL
| Total time
| |
DL
| Total time
| |
DL
| Total time
|
25 µg 20 µg | 65 s 650 s | |
14 µg 2 µg | 100 s 1,000 s | |
0.25 µg 0.10 µg | 100 s 600 s |
2.3. The sensitivity (S) expressed in counts/µg (as V2O5)
is also dependent upon analytical time:
| | | X-ray Fluorescence
|
X-ray Diffraction
| |
Direct-Beam
| |
Secondary Target
|
S
| Total time
| |
S
| Total time
| |
S
| Total time
|
18 180 | 65 s 650 s | |
4 40 | 100 s 1,000 s | |
120 720 | 100 s 600 s |
3. Method Performance (8.10.)
3.1. X-ray Instrumentation
The XRF portion of the method was evaluated in 1987 using a direct-beam energy
dispersive X-ray fluorescence (EDXRF) spectrometer and an automated powder
diffractometer (APD). Advances in X-ray instruments should enable users to obtain
similar or better results than that obtained in the method evaluation. Other X-ray
targets and detectors may be used as appropriate. Some newer instrumentation
combines X-ray diffraction and fluorescence features.
3.2. Dust Characterization
In order to assess the method performance as a function of particle size, both coarse and
fine V2O5 respirable dust materials were investigated during the evaluation of this
method. Both materials passed through a 10-µm sieve. The approximate particle-size
distributions of the two materials were measured by optical microscopy. The approximate
size-distributions found are as follows:
Material Coarse Fine |
0.5 to 3 µm 50% 99% |
3 to 10 µm 50% 1% |
These results are approximate and are only used to show that the median visible particle size of
the coarse material approximated the 3.5 µm median diameter that is characteristic of respirable
dust (8.13.). This coarse material was referred to as "respirable dust"
in the evaluation and was used to assess the method performance for respirable dust.
Particles below 0.5 µm could not be counted directly using optical microscopy. Visual
observations of acetonitrile suspensions of the coarse and fine materials indicated that a
significant amount of the fine material had a very small particle size; after thorough mixing,
the suspension of the coarse material cleared over the course of a day whereas the suspension
of the fine material remained cloudy for several days. The fine material consisted of particles
of 3 µm or less and had a particle-size distribution strongly biased towards
smaller particle sizes than respirable dust; however, it includes a particle size range that is
also deposited in the lung. The size distribution of the fine material was comparable to that
expected for samples collected with a cyclone. This material was referred to as "fine respirable"
in the method and was used in preparing calibration curves, because the method of preparation
was more reproducible than that for the coarse material.
The XRD recovery for coarse material prepared and analyzed on Ag membranes was twice that found
for the fine material. The results for the fine-respirable and respirable dust
samples are tabulated below:
Dust
|
|
X-ray Diffraction
|
X-ray Fluorescence
|
Respirable | CV1 Range | 0.0555 - 0.0830 | 0.0392 - 0.0795 |
Respirable | CV1 Pooled | 0.068* | 0.064 |
Fine-respirable | CV1 Range | 0.0946 - 0.1663 | 0.0612 - 0.1456 |
Fine-respirable | CV1 Pooled | 0.121* | 0.097 |
|
Respirable | Recovery Range | 163.4 - 190.2% | 95.7 - 97.6% |
Respirable | Recovery Mean | 177.4%* | 96.5% |
Fine-respirable | Recovery Range | 85.9 - 91.1% | 84.8 - 89.4% |
Fine-respirable | Recovery Mean | 88.7%* | 87.1% |
|
Respirable | Overall Error | ± 91% | ± 16% |
Fine-respirable | Overall Error | ± 36% | ± 32% |
* Average results from two APDs sharing the same generator.
These results indicate there are unacceptably severe errors associated with particle-size
effects in the case of analysis of V2O5 by
XRD. The XRF method is more immune to errors due to particle-size effects and is
therefore potentially more accurate if the sample particle-size distributions cannot
be reliably matched to the standard material. It should be emphasized that the XRF method only
analyzes for V. Samples should normally be quantitated by XRF and qualitatively confirmed by XRD.
After the method validation, a secondary-target EDXRF spectrometer that offered
improved analytical performance was obtained. A preliminary study using field samples submitted
for V2O5 confirmation analysis has shown
that results from samples prepared on Ag membranes and analyzed by the secondary-target
XRF were in excellent agreement with results obtained using an ICP-AES method
(8.9.). The Ag membrane samples were acid-digested after the XRF
analysis and diluted solutions were analyzed for V using the ICP.
4. Interferences
4.1. XRD
For XRD, any crystalline material with a diffraction peak in the location of the XRD diffraction
line at 31.05° 2q is an interference. An earlier study
(8.1.) indicated no interferences in common mineral compositions that
contain V2O5. During the evaluation
(8.10.), a significant low-level positive interference was
observed in some of the blanks and was not attributable to V2O5.
4.2. XRF
Note: The potential interferences for XRF were assessed using a Link Systems EDX slide
rule (Nashua Corp., Los Angeles, CA) to gauge other analyte signals and their proximity to the
V Ka and V Kb peaks.
The V XRF lines have potential positive interferences. Known potential interferences are
tabulated in Table 1. Depending on the resolution of the instrument and
the available deconvolution software, the most significant likely interferences on an instrument
with a resolution of 0.15 to 0.17 kV peak width at half height are the
Ti Kb peak which overlap the
V Ka peak and the Cr Ka
peak that occur at the V Kb peak. Depending
on the instrument geometry, Ag diffraction peaks from the Ag membrane are also possible. The
Ti Ka peak may interfere with background correction.
Sum peaks occur when the counting rate is too high to accurately measure the energies of the
X-ray photons and the energies of two or more photons are counted as a single
photon. This error can be reduced by decreasing the X-ray tube current or through
corrections using computer software. Samples containing large amounts of sulfur or chlorine may
produce sum peaks at the V Ka and
V Kb peaks, respectively. Because the
sample is deposited on a Ag membrane, Ag sum peaks may interfere with the V Kb
peak. The Ag sum peaks occur in the range of 5.266 to 7.038 kV with a maximum intensity located
near 5.4 kV.
When present in large amounts, iron and cobalt could give Si escape peaks in the vicinity of
the V Ka and Kb
positions. An escape peak is generated by the low-probability quantum excitation
of the K-shell electrons in the Si atoms of the detector producing a small peak at
1.76 kV below its parent peak.
Absorption edges (ab), are step discontinuities in the Energy vs. X-ray absorption
function. The peak intensity reduction is greater if the absorption edge is on the low energy
side of the V Ka peak. Absorption-edge
effects are not expected to be significant for the thin films used in this method.
The remaining interferences are the L lines of other elements which, when present, tend to be of low intensity.
5. Sampling
5.1. Sampling Equipment
5.1.1. Sample assembly:
Filter holder consisting of a two- or three-piece cassette, 37-mm diameter.
Backup pad, 37-mm, cellulose.
Low ash PVC membrane filter, 37-mm, 5-µm pore size [part no. P-625413,
Mine Safety Appliances (MSA), Pittsburgh, PA or cat. no. P-503700, Omega Specialty Instrument Co., Chelmsford, MA].
Note: During preparation for analysis, the sample is dissolved in tetrahydrofuran (THF). Certain acrylic copolymers added to PVC filters
are insoluble in THF. If the membrane filter composition is unknown, a laboratory test should be conducted with THF to determine suitability
before use.
5.1.2. Cyclone: Nylon, 10-mm (BDX-99R, part no. 7010048-1 Sensidyne Inc.,
Largo, FL, or part no. 456243, MSA, Pittsburgh, PA).
Note: A cyclone is optional for fume sampling. If a mixture of fume and dust is anticipated, a cyclone should be used.
5.1.3. Pump calibration system: Stop watch and bubble tube or electronic meter.
5.1.4. Sampling pump: Calibrate the personal sampling pump to approximately 1.7 L/min. Use a
representative sampler (cyclone, filter, etc.) in-line with the pump to minimize
errors associated with uncertainties in the volume sampled. For fume sampling, the cyclone is
optional and the pump can be calibrated at 2 L/min.
5.1.5. Assorted flexible tubing.
5.1.6. High volume sampling pump with cyclone (optional - for bulk sample collection). Whenever
possible, take a high volume respirable V2O5 dust
sample for laboratory use as a respirable dust reference material. This is beneficial because
V2O5 has significant particle-size effects in XRD.
5.2. Sampling Procedure
5.2.1. Place the PVC filter and a cellulose backup pad in a two or three piece cassette.
5.2.2. Attach the cassette, which is preceded by a 10-mm nylon cyclone (optional for
fume sampling), to a calibrated personal sampling pump using flexible tubing.
Do not use tubing on the inlet port of the cassette.
5.2.3. Place the sampling assembly in the breathing zone of the worker or sampling
area and place the pump in an appropriate position. Take 816 L of air through
the cassette at approximately 1.7 L/min. Do not allow the cyclone to be
inverted during or after sampling. Take full shift samples if possible. Collect
fume samples without a cyclone at 2 L/min flow rates.
5.2.4. Check the pump and sampling assembly periodically to verify performance and
to monitor particulate loading on the sample filter. If the filter becomes
overloaded (>3 mg) during the sampling interval, replace it with another filter.
5.2.5. Terminate sampling at the predetermined time and record the pump flow rate and collection time.
5.2.6. Record on the OSHA 91 form all pertinent sample data. When other
compounds are known or suspected to be present in the air, such information,
including their suspected identities, should be transmitted with the samples.
Request "confirmation analysis of vanadium pentoxide".
5.2.7. Identify and submit an appropriate blank filter from each lot of filters used.
5.2.8. Seal each filter cassette and identify it with an OSHA Form 21. Mail samples
to the laboratory in a suitable container designed to prevent damage.
5.3. Wipe samples are not appropriate for this analysis, but may be submitted for analysis by ICP-AES.
5.4. Bulk samples are appropriate, especially if they represent settled dust.
5.5. Bulk samples should be shipped separately from air samples. They should be accompanied by
Material Safety Data Sheets if available. A description of the sample composition is useful in
resolving interferences and should accompany bulk samples. Check current shipping restrictions
and ship to the laboratory by the appropriate method.
6. Analysis
Samples are analyzed by both XRD and XRF. Only one sample preparation method is necessary; both
analytical techniques are compatible with this sample preparation.
6.1. Safety Precautions
6.1.1. Follow laboratory safety rules and regulations regarding solution preparation and instrument operation.
6.1.2. Refer to the appropriate manuals for proper instrument operation and maintenance.
Analysts that are unfamiliar with the instrumentation must be trained prior to using the
equipment. Refer all non-routine maintenance to an X-ray specialist
or authorized service personnel. Wear a radiation badge and a radiation ring when using the
X-ray equipment. Be sure any radiation monitors are operating properly.
6.1.3. Tetrahydrofuran (THF) is extremely flammable and tends to form explosive
peroxides. If static electric discharge is a potential problem, use a grounded
wrist strap when transferring THF. Fires involving THF are likely to rapidly
become fierce; in case of a large fire, sound an alarm and evacuate.
6.1.4. Use appropriate caution when handling chemicals. Use latex or other chemically-resistant
gloves to avoid skin contact with solutions and suspensions of THF and V2O5.
Both chemicals are toxic. Tetrahydrofuran defats skin. Read the Material Safety Data Sheets. Use
a hood for ventilation and use a fit-tested respirator if necessary when grinding
or preparing-samples. Dispose of any waste properly.
6.2. Equipment
Either a direct-beam or a secondary-target XRF instrument can be used. A
wavelength-dispersive XRF instrument can be used in place of an EDXRF instrument.
6.2.1. Direct-beam XRF:
The spectrometer should be equipped with appropriate monitors and collimators.
The spectrometer used at the OSHA Salt Lake Technical Center (OSHA-SLTC) included the following accessories:
- Indium-lined collimators.
- Polypropylene window film, 0.20 mil thick (part no. 3520, SPEX Industries, Edison, NJ).
- Rhodium X-ray tube.
- Laboratory press, 12 ton (Cat. no. A14-100, Kevex, San Carlos, CA.).
- Lithium-drifted silicon (SiLi) detector.
- Radiation safety monitor (S.E. International Instrumentation Model Radiation Alert
Monitor 4, S.E. International Instrumentation Division, Summertown, TN). This safety monitor
can be used with other X-ray equipment below.
6.2.2. Secondary-target XRF:
The spectrometer should be equipped with appropriate monitors, collimators, and secondary
targets. In addition to the accessories listed in 6.2.1. items c-f, the
secondary-target XRF spectrometer at the OSHA-SLTC included the following accessories:
- Lucite monitor.
- Tantalum collimator.
- Iron secondary target.
- Sample holders for 25-mm diameter Ag membranes (Cat. no. A00-213,
Kevex). Note: These holders may require light machining in order to center the Ag membrane
over the most sensitive Spot.
- Pellet die set for preparing multi-channel analyzer (MCA) energy calibration
sample, 31-mm diameter (Cat. no. A10-403, Kevex.) Alternately, a
13-mm diameter die set (Cat. no. A10-401, Kevex) may be used.
- Sample holder for 31-mm diameter MCA energy calibration sample (Cat. no.
A00-214, Kevex). Note: If a 13-mm diameter pellet is used,
substitute a 13-mm diameter sample holder (Cat. no. A00-212, Kevex).
6.2.3. XRD system consisting of:
- Automated Powder Diffractometer (APD).
- X-ray Generator.
- Long, fine-focus copper target X-ray tube.
- Scintillation, position sensitive, or xenon proportional counter detector.
- X-ray accessories include: pulse-height analyzer, graphite
monochromator, 2q compensating slit, 1° receiving slit, and sample spinner.
- Sample holders for 25-mm diameter Ag membranes (Model no. PU1813/26, Philips Electronics Instruments Co., Mahwah, NJ).
- Recirculating cooling system for the X-ray tube.
6.2.4. Computer system consisting of:
- Hardware and software for data reduction and graphic presentations.
- Line profile library (JCPDS-International Center for Diffraction Data
Powder Diffraction File, JCPDS, Swarthmore, PA).
6.2.5. Standard and sample preparation:
- Low ash PVC membrane filters, 37-mm, 5-µm pore size [part no.
625413, Mine Safety Appliances (MSA), Pittsburgh, PA or cat. no. P-503700,
Omega Specialty Instrument Co., Chelmsford, MA].
- Analytical balance capable of 10 µg precision.
- Centrifuge tubes: Round bottom 40-mL (Pyrex 8260).
- Gloves, THF-resistant [such as latex gloves (Cat. no. 8852, American Pharmaseal Lab., Glendale, CA)].
- Volumetric pipettes, eyedropper, volumetric flasks and graduated cylinders.
- Magnetic stirring bar and stirrer.
- Forceps.
- Silver membrane filters: 25-mm diameter, 0.45-µm pore size
(Cat. no. FM25-0.45, Osmonics, Inc., Minnetonka, MN).
- Ultrasonic bath.
- Filtration apparatus: 25-mm (Filter Holder Hydrosol Manifold, cat. no.
XX25 047 00, filtering clamps, cat. no. XX10 025 03, fritted glass bases with stoppers, cat. no. XX10 025 02, and glass funnels,
cat. no. XX10 025 11, Millipore Corp., Bedford, HA).
- Liquid nitrogen cold-trap system for THF collection (dewar, polypropylene
2-liter suction flask, liquid nitrogen, etc.).
- Eyedropper.
- Hot plate, intrinsically safe (Model HP-11515B, Sybron/Thermolyne, Dubuque, IA).
- Teflon sheet, 0.3 to 1 mm thick (cut to fit top of hot plate).
- Plastic petri dishes (Product no. 7242, Gelman Sciences, Ann Arbor, MI).
- Vacuum system.
- Freezer mill (Model no. 6700, Spex Industries, Edison, NJ).
- Sieve, nylon, 10 µm (Spectra/Mesh N sieve, Cat. no. 08-670-205, Fisher
Scientific, Springfield, NJ) or (Cat. no. 146514, Spectrum Medical Industries, Inc., Los Angeles, CA).
- Polypropylene 250-mL wash bottle with tube cut out of top.
6.2.6. Bulk sample preparation for membrane deposition:
- Mortar and pestle.
- Sieve or sonic sifter: Sieve, 325-mesh, (or Model ATML3P Sonic Sifter with
325-mesh sieve, ATM Corporation, Milwaukee, WI).
6.3. Reagents (except for liquid nitrogen use reagent grade or better):
6.3.1. Sample and standards preparation
- Tetrahydrofuran
- Vanadium pentoxide
- Liquid nitrogen (for vapor trap and for maintaining EDXRF SiLi detector).
6.3.2. Reagents for preparing MCA calibration sample
- Titanium dioxide
- Zinc oxide
- Yttrium oxide
- Boric acid (for secondary-target instrument)
6.4. Preparation of Standard Materials
6.4.1. Use reagent grade V2O5 as the starting material.
6.4.2. Prepare respirable stock material by grinding reagent grade V2O5
in a freezer-mill operated with liquid nitrogen for 10 minutes.
6.4.3. Add this ground material to 50 to 75 mL of THF in a 100-mL glass beaker.
Stir and sonicate the suspension in an ultrasonic bath.
6.4.4. Remove the tube from the cap of a 250-mL polypropylene wash bottle. Ream
out a hole where the tube had been and smooth any rough edges. Prepare a sieving bottle by
mounting a 10-µm nylon mesh sieve in the cap. Secure the sieve in place by
screwing on the cap. Wet-sieve by first squeezing out some of the air in the bottle and inverting
the bottle in the sonicated THF. A graphic representation of the standard preparation assembly is
shown below:
WET-SIEVE METHOD FOR STANDARD PREPARATION
6.4.5. Isolate the dust from the suspension by removing the suspension from the bottle.
Filter the dust onto a Ag membrane. A filtering apparatus similar to the one described below can be used.
6.4.6. Allow the residual THF on the dust to evaporate at room temperature. Remove the dust
from the Ag membrane and store this respirable stock material in a 20-ml capped scintillation vial.
6.4.7. Assemble the filtering apparatus and liquid nitrogen cold trap, and connect to a
vacuum pump. Connect the cold trap in between the filtering apparatus and
vacuum pump to collect the waste THF. Any waste vapors should not enter the vacuum pump.
6.4.8. Prepare a 10 µg/mL THF-V2O5
suspension by weighing out 5 mg of respirable V2O5
stock material and quantitatively transferring it to a 500-mL volumetric flask.
Dilute to volume with THF. In the same manner, weigh out 50 mg of respirable V2O5
and prepare a 100 µg/mL THF-V2O5
suspension. Add a magnetic stirring bar to each and mark the new meniscus (due to the added
volume of the bar) on the neck of each volumetric flask. Sonicate 10 minutes and cool with
constant stirring on a magnetic stirrer until the meniscus reaches the new mark.
6.4.9. For each standard, center a Ag membrane on the fritted-glass base of the
filtering apparatus. Also center the glass chimney on top of the base and secure it with a clamp.
6.4.10. With the vacuum turned off, place 2 to 5 mL of THF in the chimney of the previously assembled vacuum filtering apparatus.
6.4.11. Turn off the stirrer and shake the suspension vigorously prior to removing
each aliquot from the center of the THF suspension. Consult the following
table for determining the appropriate reagent concentration and aliquot size to
take in preparing each of 24 calibration standards (three standards at each of the eight levels):
Standard µg V2O5
|
Volume (mL)
|
Reagent Concn (µg/mL)
|
50 100 200 250 500 1,000 2,000 2,500 | |
5 10 2 25 5 10 20 25 | |
10 10 100 10 100 100 100 100 | |
Note: Duplicate standards are prepared so that outliers may be identified and discarded.
Subsequent calibrations can be performed using fewer standards at levels selected to be comparable to the samples.
6.4.12. Apply vacuum to the filtering apparatus to draw the THF through the Ag membrane. This
should result in a thin, even layered deposition of the standard on the Ag membrane. Do not
rinse the chimney after the material has been deposited on the membrane. Rinsing can disturb the
thin layer deposition just created. Vacuum should be applied for sufficient time to dry the membrane.
6.4.13. Carefully disassemble the chimney and clamp. Remove the Ag membrane from the
fritted-glass base using forceps. Do not use a fixative to secure the dust on the
membranes. Place a Teflon sheet on top of an intrinsically-safe hot plate which is
set at the lowest setting. Place the membrane on top of this heated Teflon sheet.
6.4.14. When dry, place the standard in a labeled petri dish.
6.4.15. Place any THF waste in an explosion-resistant metal container and dispose appropriately.
6.5. Preparation of Bulk and Air Samples
6.5.1. Bulk sample preparation:
To prevent the possibility of contaminating air samples, prepare bulk samples in an area away from where air samples are prepared; a separate filtering
apparatus is recommended. Finely-divided bulk samples may be analyzed qualitatively without preparation as packed powders using XRD. If a
quantitative estimate is desired, prepare as follows:
- Bulk samples approximating respirable particle size:
Weigh an aliquot of 1 to 2 mg on a PVC filter, and place in a round bottom centrifuge tube. Prepare
as for air samples using the procedure in Section 6.5.2., steps 3-8.
- Non-respirable bulks:
Grind the sample to a fine powder using either a mortar and pestle or a freezer mill). Then size
the sample, using a 325-mesh sieve or sonic sifter. This results in a sample particle
size of less than 45 µm. Weigh out 1 to 2 mg of the sized sample on a PVC filter and place
it in a round bottom centrifuge tube. Prepare as for air samples using the procedure in
Section 6.5.2., steps 3-8.
6.5.2. Air sample preparation:
- When sample weights are greater than 2.5 mg, aliquots are taken to achieve depositions within the working range.
- Examine the filter and backup pad to determine if any breakthrough to the backup pad has occurred. If there is significant breakthrough, the sample
is either not analyzed or results are reported with a disclaimer. (See Section 7.3. for reporting results.)
- Carefully transfer the air sample (PVC filter) from the cassette to around-bottom
40-mL centrifuge tube. Add 10 mL THF to dissolve the filter and suspend the sample.
Sonicate the sample suspension for 5 to 10 min.
- For each sample, center a Ag membrane on the fritted-glass base of the filtering
apparatus. Also center the glass chimney on top of the base and secure it with a clamp.
- With the vacuum turned off, place 2 to 5 mL of THF in the chimney of the previously assembled vacuum filtering apparatus.
- Quantitatively transfer the suspension with rinses of THF to the glass chimney of the vacuum
filtering apparatus. The total volume in the chimney should not exceed 20 mL.
- Continue preparation of samples as described Sections 6.4.12. to 6.4.15.
6.6. XRD Calibration and Analysis
Note: The analytical procedures for XRD and XRF are written for in-house use;
other equipment and data reduction systems similar to those mentioned can be
used. The custom OSHA laboratory XRD software and XRD Standard Operating
Procedure (SOP) (8.14., 8.15.) for data collection and reduction used in this
analysis incorporate software which smoothes data, locates peaks, selects
integration methods, performs weighted one- or two-piece calibrations, calculates
Microgram quantities, and permits analyst review and direction.
Analyze the 24 standards prepared in Section 6.4. according to the XRD SOP
(8.15.). Instrument parameters (all angles are in terms of °2Q) for calibration are:
| | |
Ag reference calibration | 38.15° (primary Ag line) 44.33° (secondary Ag line) |
----------------------- V2O5 scan ranges ----------------------- |
Secondary line | 30.40 to 31.70° (quantitative) (31.05° analytical peak) |
Primary line | 19.62 to 20.92° (qualitative) |
Tertiary line | 25.56 to 26.86° (qualitative) |
|
Scan step size | 0.02° |
Integration time | 1 to 10 s |
X-ray tube power | 40 kV and 40 mA |
The primary or secondary Ag line is used as the reference calibration. The secondary
V2O5 line was chosen as the quantitative line
because interferences from quartz are present on the primary and tertiary lines. It is possible
to make calibration curves for these two lines only if the particle-size distribution
closely matches the dust used in the preparation of standards, because they are more severely
affected by particle-size and preferred-orientation effects than the
line at 31.05° 2Q.
Typical scans of the three recommended analytical lines for a 250 µg V2O5
standard are graphically displayed in Figure 1.
Use available peak-integration or profile fitting software to obtain normalized
count data. Construct a least-squares curve-fit and derive calibration
coefficients for use in subsequent V2O5 analyses.
A polynomial curve fit can be used. The presence of V2O5
should be qualitatively confirmed by the presence of the two additional confirmational
V2O5 lines from the 2q
scan ranges; a strong line at 19.62 to 20.92°, and a weak line at 25.56 to 26.86°.
Consult the XRD SOP (8.15.) for further instructions.
6.7. XRF Calibration and Analysis
6.7.1. Direct-beam XRF spectrometer:
- Consult the XRF SOP (8.16.) and power up the X-ray tube to 20 kV and
0.5 mA. Remove any filter and put the narrow collimator in place over the detector. Calibrate the element
markers using a Ti-Zn-Y calibration standard consisting of a 1- to 3-mm
deep equimolar mix of powdered TiO2, ZnO, and Y2O3
on a 0.25 mil mylar support. Calibrate on the corresponding Ka
lines at 4.508, 8.631, and 14.933 kV.
- First analyze the standards. Next analyze the samples and repeat the standards. For each sample or standard perform the following steps:
- Close the shutter and open the cover.
- Wipe the yellow Kapton window before analyzing each sample.
- Place the sample face down on the yellow Kapton window over the detector, close the cover, open the shutter and initiate analysis.
- Label the data to identify standards and samples analyzed. Typical spectra for a 250 µg
V2O5 standard and a blank are shown in Figure 2.
- Save each spectrum to magnetic tape or disk (if possible). Note: On certain XRF systems without mass storage, it may be necessary to perform all
the calculations after each sample is analyzed. Others allow some calculations to be incorporated automatically in the analyses prior to saving the spectra.
6.7.2. Secondary-target XRF spectrometer:
- Open sample chamber lid.
- Install a Ta collimator at the detector if not already installed.
- Install the Fe secondary target if not already installed.
- Place Ag membranes in the autosampler. Position each sample for maximum analytical signal during analysis.
- Close sample chamber lid.
- Calibrate the MCA using the Ti-Zn-Y calibration standard consisting of a
pellet made of a mixture H3BO3 and 10-20% of an equimolar mix of
powdered TiO2, ZnO, and Y2O3.
Calibrate on the corresponding Zn and Y Ka
lines at 8.631 and 14.933 kV respectively using a Ag secondary target. One may substitute other
appropriate calibration materials or secondary targets for this purpose.
- Select the following conditions:
Fe secondary target.
Display range = 10 kV.
X-ray tube power = 12 kV and 3.3 mA.
(To reduce the affect of sum peaks, do not exceed 50% deadtime on standards or samples.)
Atmosphere = Vacuum.
Acquisition time = 600 s.
Analyze a Lucite monitor sample or other appropriate monitor sample to adjust for varying X-ray tube output.
- Analyze all standards and samples.
- Save spectra to magnetic tape or disk. Note: On certain XRF systems it may be convenient to perform escape and sum peak corrections operations prior to saving each spectrum.
7. Calculations
7.1. XRD Data Reduction
The XRD results from line to line may not agree well. Use the 2 line results for calculations
unless interferences are present. After calibration and analysis, calculate the microgram
quantities of the analyte from the calibration curve. The V2O5
exposure is estimated using the following equation:
Vanadium pentoxide exposure (As V2O5 in mg/m3) = A / B
A = µg V2O5 found (blank subtracted)
B = sample air volume in liters
7.2. XRF Data Reduction
Use software features available on instrument:
7.2.1. Recall spectrum.
7.2.2. Remove escape peaks.
7.2.3. Remove sum peaks.
7.2.4. Identify V lines and any interfering elements.
7.2.5. Perform background corrections.
7.2.6. Perform profile fit or Gaussian deconvolution of the identified lines.
7.2.7. Tabulate integrated peak counts and masses of standards.
7.2.8. Use the count (or count rate) data to construct a least-squares curve-fit and
derive calibration coefficients for use in subsequent V2O5
analyses. A polynomial curve fit can be used. Use the concentration-response curve to
calculate the g of analyte present. The V2O5
exposure is calculated as in Section 7.1.
7.3. Reporting Results
Compare the XRD and XRF results. As previously mentioned (Section 3.3.),
particle-size and preferred-orientation effects may cause the XRD
results to differ from the XRF results by as much as a factor of two. The XRD results are used
as confirmation. If V is found by XRF and V2O5
is not confirmed by XRD, the analyst may wish to identify the vanadium-containing
species using XRD wide-angle scans and a XRD line library similar to what is mentioned
in Section 6.2.4.
If a reasonable comparison is obtained between XRD and XRF, report results to the
industrial hygienist as mg/m3 V2O5.
Do not report XRD quantitative results; XRF results represent more reliable estimates of the
amounts of V2O5 potentially present. Results by XRD represent
estimates only.
Disclaimers:
Particulate present on the backup pad constitutes some sample loss. Occasionally this may be
seen and can be due to a poor cassette seal on the filter, improper positioning of the filter in the
cassette, or poor quality control of the filter and/or cassette. If this type of contamination
occurs, relay a note to the compliance officer indicating that some of the sampled material was
found on the backup pad and the reported value may be lower than the actual exposure.
8. Reference
8.1. Carsey, T.P.: Quantitation of Vanadium Oxides in Airborne Dusts
by X-ray Diffraction. Anal. Chem. 57:2125-2130 (1985).
8.2. National Institute for Occupational Safety and Health:
Collaborative Tests of Two Methods for Determining Free Silica in Airborne Dust
(DHHS Publication No. 83-124). Cincinnati, OH: National Institute for Occupational Safety and Health and Bureau of Mines 1983.
8.3. National Institute for Occupational Safety and Health:
NIOSH Manual of Analytical Methods, 3rd. ed., Cincinnati, OH: National Institute for
Occupational Safety, 1986. pp. 7504-1 to 7504-10.
8.4. Windholz, M.; ed.: The Merck Index. 10th ed., Rahway, NJ: Merck and Co., 1983. p. 1417-1418.
8.5. American Conference of Governmental Industrial Hygienists:
Documentation of the Threshold Limit Values and Biological Exposure Indices (Pub. No. ISBN:
0-036712-68-6). Cincinnati, OH: American Conference of Governmental Industrial Hygienists, 1986. p. 620.
8.6. National Institute for Occupational Safety and Health:
Criteria for a Recommended Standard ... Occupational Exposure to Vanadium (DHEW/NIOSH
Pub. No. 77-222). Washington, D.C., U.S. Government Printing Office, 1977. pp. 21-44.
8.7. "Air Contaminants; Final Rule" Federal Register 54:12 (19 Jan. 1989). pp. 2475-2476.
8.8. Occupational Safety and Health Administration Analytical Laboratory:
OSHA Manual of Analytical Methods edited by R.G. Adler (Method No. I-1. Salt Lake City, UT. 1977.
8.9. Occupational Safety and Health Administration Technical Center:
Metal and Metalloid Particulates in Workplace Atmospheres (ICP Analysis) by J.C. Septon
(USDOL-OSHA-SLTC Method No. ID-125G). Salt Lake City, UT. Revised 1991.
8.10. Occupational Safety and Health Administration Technical Center:
Vanadium Pentoxide Backup Data Report (ID-185) by M.C. Rose. Salt Lake City, UT. Revised 1991.
8.11. Occupational Safety and Health Administration Analytical Laboratory:
OSHA Analytical Methods Manual (USDOL/OSHA-SLCAL Method No. ID-142).
Cincinnati, OH: American Conference of Governmental Industrial Hygienists (Pub. No. ISBN: 0-936712-66-X), 1985.
8.12. Occupational Safety and Health Administration Analytical Laboratory:
OSHA Analytical Methods Manual (USDOL/OSHA-SLCAL Method No. ID-143).
Cincinnati, OH: American Conference of Governmental Industrial Hygienists (Pub. No. ISBN: 0-936712-66-X), 1985.
8.13. American Conference of Governmental Industrial Hygienists:
Threshold Limit Values and Biological Exposure Indices for 1989-1990
(Pub. No. ISBN: 0-936712-81-3). Cincinnati, OH: American Conference of Governmental Industrial Hygienists, 1989.
8.14. Occupational Safety and Health Administration Analytical Laboratory:
X-ray Diffraction Program Documentation. Salt Lake City, UT. 1981 (unpublished).
8.15. Occupational Safety and Health Administration Analytical Laboratory:
Standard Operating Procedure (XRD). Salt Lake City, UT. In progress (unpublished).
8.16. Occupational Safety and Health Administration Analytical Laboratory:
Standard Operating Procedure (XRF). Salt Lake City, UT. 1981 (unpublished).
Table 1 Potential Vanadium XRF Line Interferences (Not shown to scale.)
Interference
|
Line ID. = Energy (kV) =
|
V Ka 4.949
|
|
V Kb 5.427
|
| | | | | | |
Sn LI ab | 4.464 |
| | | : | | : |
Ba La1 | 4.467 |
| | | : | | : |
Ti Ka | 4.508 |
| | | : | | : |
I LIIIb2 | 4.507 |
| | | : | | : |
I LIII ab | 4.559 |
| | | : | | : |
Te LII l1 | 4.570 |
| | | : | | : |
Te LII ab | 4.613 |
| | | : | | : |
S 2 × Ka sum peak | 4.614 |
| | | : | | : |
Cs LIIb1 | 4.620 |
| | | : | | : |
Fe Ka Si escape peak | 4.64 |
| | : | | : |
La LIIIa1 | 4.651 |
| | : | | : |
Sb LI ab | 4.697 |
| | : | | : |
S Ka + Kb sum peak | | 4.775 |
| | | : | | : |
I LIIIl1 | | 4.800 |
| | | : | | : |
Ba LIIb1 | | 4.828 |
| | | : | | : |
Ce LIIIa1 | | 4.840 |
| | | : | | : |
I LII ab | | 4.856 |
| | | : | | : |
Ti Kb | | 4.931 |
| | | | | : |
Cs LIIIb2 | | 4.936 |
| | | | | : |
S 2 × Kb sum peak | | 4.936 |
| | | | | : |
Te LI ab | | 4.938 |
| | | | | : |
Cs LIII ab | | 5.011 |
| | : | | | : |
Pr LIIIa1 | | 5.034 |
| | : | | | : |
La LII b1 | | 5.043 |
| | : | | | : |
Ba LIIIb2 | | 5.156 |
| | : | | | : |
Co Ka Si escape peak | | 5.16 |
| | : | | | : |
I LI ab | | | 5.190 |
| | : | | | : |
Nd LIIIa1 | | | 5.230 |
| | : | | | : |
Cl 2 × Ka sum peak | | | 5.244 |
| | : | | | : |
Ba LIII ab | | | 5.247 |
| | : | | | : |
Ce LII b1 | | | 5.262 |
| | : | | | : |
Cs Ll1 | | | 5.280 |
| | : | | | : |
Fe Kb Si escape peak | | | 5.30 |
| | : | | | : |
Cs LII ab | | | | 5.358 |
| | : | | | : |
La LIIIb2 | | | | 5.384 |
| | : | | | : |
Ag sum peaks | | | | 5.266 - 7.038 |
| | : | | |< - : -> | |
Cr K | | | | 5.41 |
| | : | | | : |
Cl K + K sum peak | | | | 5.439 |
| | : | | : | |
Pr LIIb1 | | | | 5.489 |
| | : | | : | |
La LIII ab | | | | 5.489 |
| | : | | : | |
Ba LII1 | | | | 5.531 |
| | : | | : | |
Ba LII ab | | | | 5.623 |
| | : | | : | |
Cl 2 × Kb sum peak | | | | 5.634 |
| | : | | : | |
Sm La1 | | | | 5.636 |
| | : | | : |   |
Cs LI ab | | | | 5.720 |
| | : | | : | |
Ce LIII ab | | | | 5.729 |
| | : | | : | |
The L lines for the noble gas xenon (Xe) and the short-lived radioactive element promethium (Pm) were
not considered due to their extremely low probability of occurrence in compliance samples.
Analytical Peaks for X-ray Diffraction (250 µg V2O5 Standard)
Figure 1
Analytical Peaks for X-ray Diffraction (250 µg V2O5 Standard)
COMMENTS
|
50 µg V2O5 STD = V250C13BR
BLANK = VBLG |
|
ANALYST DATE COLLIMATOR SECONDS % DEAD TIME |
Mike Rose 3/13/87 Narrow 100 <10 |
SAMPLE # KV = MA = VACCUM = FILTER = VERT. SCALE = HOR. SCALE = SMOOTHED |
V250C13BR/VBLG 20 0.5 AIR TORR NONE 256 Full SMOOTHED 0 times |
Figure 2
|