Physical Measurements and Analysis OSHA Salt lake Technical Center Salt Lake City, Utah 1. Introduction
1.2. History of the Method
1.2.2. In 1975, ignition was used at the OSHA Laboratory to eliminate interference from particulates that do not lose significant weight upon ignition, i.e. minerals. 1.2.3. In 1979, scanning electron microscopy was used to distinguish between carbon black and other organic substances. With this procedure, scanning electron microscopic analysis is no longer required. 1.3. Principle
1.4. Classification and Properties
1.5. Uses
2. Range and Detection Limit
2.2. The detection limit is 0.1 mg. This value corresponds to 0.21 mg/m3 carbon black exposure for the recommended air volume of 480 liters. 3. Recovery, Precision, and Accuracy
3.2. The average recovery of samples spiked with
3.3. The coefficient of variation for the total analytical and sample preparation procedure in the range of sample sizes validated in this study was 0.14. This value corresponds to a 0.49 mg/m3 standard deviation at the OSHA standard level of 3.5 mg/m3. 3.4. The standard deviation for weight measurements is 76 ug. 4. Interferences
4.1.2. Rubber-processing chemicals are usually grouped into accelerators, antioxidants and stabilizers, antiozonants, and miscellaneous materials (8.2). These chemicals are defined by the U.S. Tariff Commission as "organic compounds that are added to natural and synthetic rubbers to give them qualities necessary for their conversion into finished rubber goods" (8.3). Most of the commercial organic chemicals used in the rubber industry are listed in Tables of the Encyclopedia of Chemical Technology (8.4). 4.2. Controlling Interferences
THF is the preferred sovent extractant because of its unique
solvent properties and wide miscibility. It is a saturated cyclic
ether (CH2)4O
and is a good solvent for Other solvents may be used for the removal of organic chemicals
insoluble in THF. The efficacy of each should be
Note: Some Carbon Blacks may contain significant amounts of Polynuclear Aromatic Hydrocarbons (PAHs) such as:
Pyrene Chrysene Benzo - a - Pyrene (BaP), a = alpha If PAHs are suspected to be present, the following additional steps may
be performed:
B. Take separate personal air samples on glass fiber filters C. Take area samples on glass fiber filters The filter used for PAH sampling is a Gelman Glass Fiber Filter, Type A/E, Product Number 61652, 37 mm diameter, which has a Typical Aerosol Retention Rating of 99.98% for a 0.3 um aerosol particle size. The Glass Fiber Filters are then analyzed for PAH content using parameters established in OSHA Method 58. The PVC samples collected are analyzed for Carbon Black.
5.1.2. Personal Sampling Pump: A calibrated personal sampling pump whose flow can be determined within ±10% at the 95% confidence level at the recommended flow rate. The pump must be calibrated with a filter holder and filter in the line. 5.1.3. Polyvinyl chloride (PVC) membrane filter;
5.1.4. Filter back-up pads. 5.1.5. PetriSlide (Millipore) or plastic Petri dish - as filter holder for storage and weighing. 5.1.6. Desiccator containing anhydrous calcium sulfate or other suitable desiccant. 5.1.7. Microbalance with 0.01 mg scale divisions. 5.2. Procedure
5.2.2. The filter is placed on a back-up pad in a cassette filter holder. The three piece cassette filter holder is tightened and tape is placed around the cassette to prevent loss of sample. A piece of flexible tubing from the pump is connected to the cassette. The air is sampled at a flow rate of 2 liters per minute. After sampling, the openings of the cassettes are sealed with plugs. 5.2.3. The cassette filter holder is opened and the filter is carefully transferred to a PetriSlide. 5.2.4. The filter is dried in a desiccator and then weighed. 5.2.5. The sample weight is determined as the difference between the filter weight before and after sampling. 5.2.6. The filter is again placed in the cassette holder and sealed with tape and an official OSHA Form 21 seal. 5.2.7. Bulk samples of the materials contributing to the airborne dust in the working atmosphere should be placed in bottles for supporting test data and analysis. 6. Analytical Procedure
6.1.2. Drying oven capable of maintaining temperature at 150°C. 6.1.3. Balance with 0.01 mg scale divisions. 6.1.4. Filtering assembly and vacuum system. 6.1.5. Glass fiber filters without organic binder. 6.1.6. Centrifuge tubes (50 mL) 6.1.7. Porcelain plate. 6.1.8. Ultrasonic bath. 6.2. Reagent
6.3. Safety Precautions
When using THF, work in a well ventilated hood and wear protective gloves to avoid skin contact. 6.3.2. Ususal precautions in working near high temperature furnaces or ovens and fumes should be observed. Protective gloves and tongs should be used in the transfer of samples to and from the high temperature apparatus. 6.4. Laboratory Quality Control Samples
6.5. Sample Analysis
6.5.2. The samples are transferred (filtered) onto glass fiber filters (the THF is vaccumed througha the GFF and the filtered THF captured is a liquid nitrogen trap prior to the vaccum air and THF getting into the vaccum air system pump) and the carbon black is washed 3 times with 2 or 3 ml THF. Occassionally a carbon black sample has a very large amount of sample which will virtually not allow the THF to filter through the GFF. To facilitate filtration and allow efficient analytical work, the pre-filtrate sample material can be portioned onto one or more additional GFF (as many as needed) so as to allow for smaller amounts of sample material per each GFF and allow proper filtering and efficient sample processing. The result for each of the GFF filtrations for that sample can be added together to obtain the correct level of carbon black in the original heavily loaded sample. 6.5.3. After the final wash, immediately remove the glass fiber filter from the filter assembly in order to prevent adherence to the glass assembly. The glass fiber filter can be placed in an aluminum foil weighing pan (2 1/2" high perimeter foil wall and a 2" diameter). The sample number can be imprinted into the foil with a pen or pencil. Also, the sample on the glass fiber filter can be kept in the aluminum foil pan during sample analysis steps 6.5.4. and 6.5.5. (this aids in maintaining sample identification). 6.5.4. The filters are dried in an oven at 150 for 30 minutes, cooled to room temperature and weighed (W1). 6.5.5. The filters are placed on a porcelain plate and placed in a muffle furnace at 600° for 30 minutes. 6.5.6. The filters are cooled to room temperature and weighed (W2). 6.6. Calculations The amount of carbon black is determined by subtracting W2 from W1. This value in ug is divided by the sample air volume in liters and reported as mg/m3.
where W3 is the blank value in ug. 7. Reporting
7.2. Analytical data and results are checked by another analyst before the results are submitted to the supervisor for approval. 8. REFERENCES:
8.2. Kirk, Othmer, Encyclopedia of Chemical Technology, Vol. 17, 2nd Edition, 509 (1978). 8.3. U.S. Tariff Commission, U.S. Production and Sales of Rubber Processing Chemicals for 1966, Aug. 11, (1967). 8.4. Kirk, Othmer, Encyclopedia of Chemical Technology, 2nd
Edition, Vol. 10, p. 8.5. Kirk, Othmer, Encyclopedia of Chemical Technology, 2nd Edition, Vol. 10, p. 249, (1978).
|