ORGANIC VAPORS

(See Section 4)


Method no.: 07
Matrix: Air
OSHA PELs: Section 4
Procedure: Collection on charcoal, desorption with an organic solvent, and analysis by gas chromatography with flame ionization detector.
Recommended air volume
and sample rate:
Section 4
Status of method: This method has been used extensively in the OSHA Analytical Laboratory. With slight modification, this method is a generalized version of validated NIOSH methodology. This method has been revised to include validated OSHA methodology also.





Date: May 1979 by: Organic Methods Evaluation Branch
Last Update: November 1989 by: Michael Shulsky

Organic Methods Evaluation Branch
OSHA Analytical Laboratory
Salt Lake City, Utah

1. General Discussion

1.1. Background

Background information on the analytes may be obtained from a number of sources such as NIOSH Criteria Documents, chemical dictionaries and industrial hygiene manuals. Solvents are used for degreasing, for dry cleaning, and in the manufacture of many materials ranging from paints, varnishes, shellacs, and lacquers to rubber and synthetic resins. When not being used as solvents, they may function as fuels or act as chemical intermediates with or without regard to their ability to put materials into solution. Toxic effects of the analytes vary with many acting as irritants or causing narcosis, and some having more hazardous effects.

1.2. Statistical parameters

1.2.1. Each analyte included in this general procedure has a validated NIOSH method, (Ref. 5.1.) and/or a validated OSHA method. One of the NIOSH validation requirements is that the results obtained be within ±25% of the true values at the 95% confidence level at the air concentration equal to the OSHA standard. Although the OSHA evaluation procedure differs from that of NIOSH, the same validation requirements are used.

1.2.2. Refer to the validated NIOSH methods, (Ref. 5.1.) or OSHA methods for detailed information on individual analytes.

1.3. Advantages

1.3.1. The sampling device is small, portable, and involves no liquids.

1.3.2. The analysis is by a quick instrumental method.

1.3.3. Interferences can be eliminated by altering chromatographic conditions in most cases.

1.3.4. The method allows simultaneous analysis of two or more analytes.

1.4. Disadvantages

1.4.1. The air volume sampled is limited by the capacity of the charcoal tubes. Exceeding the capacity of the charcoal tube results in loss of sample. The adsorptive capacity is decreased by high humidity.

1.4.2. The method is limited by the reproducibility of the pressure drop across the tubes. The pressure drop affects the flow rate causing the air volume to be imprecise.

1.4.3. The analyst must work with toxic solvents.

1.4.4. When many components are present, elimination of interferences becomes difficult.

2. Sampling Procedure

2.1. Apparatus

2.1.1. A calibrated personal sampling pump whose flow can be determined within ±5% at the recommended flow rate with the sampling device attached.

2.1.2. Charcoal tubes: Glass tubes with both ends flame sealed, 7 cm long with a 6-mm o.d. and 4-mm i.d., containing 2 sections of 20/40 mesh activated charcoal separated by a 2-mm portion of urethane foam. The activated charcoal is prepared from coconut shells and is fired at 600°C prior to packing. The adsorbing section contains 100 mg of charcoal, the backup section 50 mg. A 3-mm portion of urethane foam is placed between the outlet end of the tube and the backup section. A plug of silylated glass wool is placed in front of the absorbing section. The pressure drop across the tube must be less than 1 in. of mercury at a flow rate of 1 L/min.

2.1.3. Certain analytes require petroleum base charcoal instead of coconut base charcoal. This requirement is specified in Section 4.

2.2. Reagents

None required in sampling procedure.

2.3. Technique

2.3.1. Immediately before sampling, break the ends of the tube to provide an opening at least one-half the internal-diameter of the tube (2 mm).

2.3.2. The smaller section of charcoal is used as a backup and should be positioned nearest the sampling pump.

2.3.3. The charcoal tube should be placed vertically during sampling to minimize channeling.

2.3.4. Air being sampled should not be passed through any hose or tubing before entering the charcoal tube.

2.3.5. Do not exceed the recommended air volume.

2.3.6. The charcoal tubes should be capped with the supplied plastic caps immediately after sampling. Under no circumstances should rubber caps be used.

2.3.7. One tube should be handled in the same manner as the sample tube (break, seal and transport) except that no air is sampled through this tube. This tube should be labeled as a blank.

2.3.8. Capped charcoal tubes should be wrapped end to end with official OSHA seals. They should be packed tightly and padded before they are shipped to minimize tube breakage during shipping.

2.3.9. For certain analytes where migration on the charcoal is a significant problem, it may be requested that two charcoal tubes be used in series in order that breakthrough may be distinguished from migration. These tubes must be separated and individually capped and sealed before shipping.

2.4. Breakthrough

Breakthrough data is presented on each analyte in its respective validated NIOSH method (Ref. 5.1.) and/or OSHA method.

2.5. Desorption efficiency

2.5.1. The back end of a charcoal tube is opened and the backup portion of activated charcoal is removed, leaving the front 100-mg portion of activated charcoal intact in the tube. The activated charcoal must be of the same lot as that in the tubes used to collect the samples. A known amount of analyte is injected directly into the activated charcoal with a microliter syringe and the tube is capped.

2.5.2. Six tubes at each of three concentration levels (0.5, 1, and 2 times the standard) are prepared by adding an amount of analyte equivalent to that present in a recommended air sample at the selected level. The tubes are allowed to stand at least overnight to assure complete adsorption of the analyte onto the charcoal. These tubes are referred to as the samples. A parallel blank tube should be treated in the same manner except that no analyte is added to it. The sample and blank tubes are desorbed and analyzed in exactly the same manner as the sampling tube described in Section 3.

2.5.3. The desorption efficiency (DE) equals the average weight in mg recovered from the tube divided by the weight in mg added to the tube, or

DE(%) = average weight (mg) recovered
weight (mg) added
× 100

2.5.4. If there is a significant change in desorption efficiency over the range of loadings studied, a plotted curve of DE versus mg recovered must be used to correct for adsorption losses.

2.5.5. If there is no significant change in DE over the range studied, reconfirmations need only be carried out at one loading in the middle of the range.

2.6. Recommended air volume and sample rate

See Section 4. for recommended air volume and sampling rate.

2.7. Interferences

2.7.1. It is important to be aware of other components in the atmosphere which may interfere with the collection of the analyte.

2.7.2. High relative humidity may significantly affect the collection of some analytes.

2.8. Safety precautions

2.8.1. Care must be taken when opening the sealed ends of charcoal tubes to avoid cuts to the hands.

2.8.2. Safety glasses should be worn when opening the sealed ends of charcoal tubes to avoid injury to the eyes from glass splinters.

3. Analytical Procedure

3.1. Apparatus

3.1.1. Gas chromatograph equipped with flame ionization detector.

3.1.2. Columns. A variety of columns are suitable. Two good selections are a 10-ft × 1/8-in. i.d. stainless steel column containing 10% SP1000 on 80/100 mesh Supelcoport and a 10-ft × 1/8-in. i.d. stainless steel column-containing 20% SP2100 with 0.1% Carbowax 1500 on 100/120 mesh Supelcoport.

3.1.3. A suitable method of measuring peak areas, such as an electronic integrator or data system.

3.1.4. Two-milliliter vials with either screw-on or crimp-on caps which contain Teflon-lined septa.

3.1.5. Microliter syringes; one-microliter for GC injections and 10-µL for standard preparation, or other suitable sizes.

3.1.6. Pipets for dispensing desorbing solvent (DS). A Glenco 1-mL reagent dispenser is adequate and convenient.

3.1.7. Volumetric Flasks. Five-milliliter and other convenient sizes.

3.1.8. Glass tubing cutter.

3.2. Reagents

3.2.1. Chromatographic quality desorbing solvent (DS). Although carbon disulfide is commonly used as the DS, certain analytes can be more effectively desorbed with the use of alternate solvents or solvent solutions. These alternate DSs are listed in Table 4.1. and are normally used when the single analyte is requested or when the requested analytes are known to be effectively desorbed with that DS. When analysis for a number of analytes requiring different desorbing solutions is requested, the preferred DS will usually be carbon disulfide.

3.2.2. Analyte standard, reagent grade.

3.2.3. Internal standard, (optional) reagent grade. p-Cymene is a suitable internal standard for many solvents.

3.2.4. Chromatographic quality helium, hydrogen, and air.

3.3. Standard preparation

3.3.1. Prepare analyte standard at a concentration of 1 µL of analyte per milliliter of DS by adding 5 µL of analyte to a 5-mL volumetric flask partially filled with DS. Fill the volumetric flask to the mark and invert 3 or 4 times for proper mixing. Other size volumetric flasks may also be used to prepare the 1 µL/mL analyte standards. At least two standards at 1 µL/mL are prepared. Standards must be used the day they are prepared. In some cases, analyte standards in concentrations other than 1 µL/mL may be more suitable, especially with analytes that have extremely high or low OSHA standards.

3.3.2. Injection of standards is accomplished with a 1-µL or other suitable syringe. The syringe is rinsed thoroughly in carbon disulfide between standards. Injector septa should be checked for wear daily.

3.3.3. Injection sizes other than 1-µL and injection by means of a gas chromatograph autosampler are acceptable in most cases.

3.4. Preparation of samples

3.4.1. The status of the seals on each charcoal tube is noted and recorded as intact, broken, or none.

3.4.2. The field identification number, the laboratory identification number and signature of the industrial hygienist on each sample seal are checked with those on the sample identification sheets.

3.4.3. The seal is removed and the charcoal tube is opened with a glass tubing cutter at the end containing the larger portion of charcoal. The front and back sections of charcoal are transferred to separate 2-mL capped vials. The glass wool plug and the small wad of urethane foam separating the two sections of charcoal are discarded.

3.4.4. The charcoal lot number is noted in order that the proper desorption efficiency is used in later calculations.

3.4.5. Gas chromatography parameters are set as recommended in the instruments manual. Oven temperature and column are varied until an optimum chromatogram is produced by the analyte standard.

3.4.6. Once the internal standard has been verified as not interfering with other peaks in the chromatogram, the samples are desorbed. One milliliter of DS is dispensed into each sample vial. The vial is immediately sealed. Each vial is swirled periodically to increase the rate of desorption. Twenty to thirty minutes is typical for the desorption process.

3.5. Analysis

3.5.1. The data processor can be calibrated to provide results directly in units of air concentration (normally ppm at 25°C and 760 mm Hg) by equating the peak areas from analytical standards with their equivalent air concentration values, assuming the recommended air volume was collected. Usually slight corrections must be made to the processor results to compensate for the volume of air actually collected for each individual sample. With a few of the analytes an additional similar correction may be necessary due to desorption efficiencies that change with concentration. The linear nature of the flame ionization detector allows the use of a point calibration, but the bracketing of samples with analytical standards is a good practice. The calculation of the equivalent air concentration for an analytical standard is detailed in Section 3.7.1. Specific instructions for data processor calibration procedures are found in the instruments' manuals.

3.5.2. Sample injection is accomplished with a 1-µL or other suitable syringe. The syringe is rinsed thoroughly in carbon disulfide between samples. Injector septa should be checked for wear periodically. Injection by means of a gas chromatograph autosampler is acceptable.

3.5.3. Bracket the samples with analytical standards if detected concentrations are above the PEL.

3.5.4. When the identity of a suspected analyte peak is in question, it should be confirmed by GC/MS, GC/IR, or by retention time on at least two GC columns containing different packing material. The identity of the analyte should be considered suspect when detected concentrations are above the PEL.

3.6. Interferences

Interferences to the analytical method will in most cases appear as poor resolution of the analyte peak from other components. This may be overcome by prudent selection of a more suitable chromatographic condition or column.

3.7. Calculations

3.7.1. An equivalent air concentration for analytical standards is used to calibrate the data processor such that analytical results are obtained directly in the units of ppm. For a standard which is 1 µL analyte per milliliter of DS the calculation of the calibration amount code proceeds in the following manner:

mg analyte = 1 mL DS (Desorp. Vol.) × 1 µL analyte
1 mL DS
× D

  where D = analyte density in mg/µL

Assume the recommended air volume V had been sampled and determine the analyte air concentration in mg/m3.

The relation between concentrations in mg/m3, and ppm at 25°C and 760 mm is:

ppm = (mg/m3)(24.46)/(MW)

where ppm = the air concentration based on an air volume
of V and a DE of 100%
24.46 = molar Volume at 25°C and 760 mm Hg
MW = molecular weight of the analyte

If the DE is not 100%, the corrected air concentration (ppm') will be obtained by

ppm' = ppm/DE(decimal form)

ppm' is the calibration value assigned to the peak area generated by the 1 µL/mL analyte standard when it is analyzed.

3.7.2. After the results are obtained, one additional correction must be made. Each processor result (PR) must be corrected to the actual air volume (V') sampled. This final correction is made in the following manner:

ppm'' = PR/(V'/V)

PPM'' is the corrected result which is reported to the industrial hygienist.

3.7.3. The following example is the calibration code calculation for toluene:

1 mL DS (Desorp. Vol.) × 1 µL toluene
1 mL DS
× 0.8669 mg/µL = 0.8669 mg toluene

Assume 10-L air volume

0.8669 mg
0.01 m3
= 86.69 mg/m3

86.69 mg/m3 × 24.46
92.15
= 23.01 ppm

Desorption Efficiency for charcoal lot used, 97%

ppm' = 23.01/0.97 = 23.72

The calculations should be considered an example only, and various parameters confirmed before used in actual analysis.

3.7.4. The results obtained by the processor must be corrected as in the following examples:


  PR V'   V V'/V     PPM

52 ppm 8 10 0.8   65 ppm
52 ppm 5 10 0.5 104 ppm
52 ppm 8   5 1.6   32.5 ppm

3.8. Safety precautions

3.8.1. Care must be taken when opening charcoal tubes to avoid cuts to the hands.

3.8.2. Safety glasses must be worn throughout the analytical procedure.

3.8.3. Work involving solvents open to the atmosphere must be performed in a hood.

3.9. Reporting results

3.9.1. When results uncorrected for air volume are greater than 10 ppm, three significant digits will be reported. For results below 10 ppm, the chemist will use his judgment, but in no cases report more than three significant digits.

3.9.2. The estimated detection limit based on a 10-L air sample will be reported on the analyst worksheet when no analyte is detected.

3.9.3. All concentration levels down to the detection limit are reported.

3.9.4. If the concentration of analyte found on the back section of the charcoal tube is equal to or greater than 25% of the concentration found on the front section, the charcoal tube is considered to be saturated and reported as such on the analyst worksheet.

3.9.5. The presence of significant peaks caused by unrequested components in the sample is noted on the analyst worksheet and they are identified and quantitated if possible.

3.9.6. All data processor print-outs and chart recorder chromatograms are filed in a central file according to laboratory sample identification number.

3.9.7. Analytical data and results are checked by a fellow chemist before the completed analyst worksheets are given to the supervisor.

4. Analytes

The following table contains those analytes which can be analyzed by this procedure. Standard size charcoal tubes containing coconut base charcoal are used unless specified otherwise in the table. Listed PELs are 8-h time weighted averages unless denoted as a ceiling concentration with a "(C)", before the PEL value. Before taking samples, the OSHA Field Operations Manual should be consulted for additional and more detailed information.

Table 4.
Recommended Sampling Parameters
and Desorption Solvents for Analytes Covered by This Procedure.


analyte (1985)
PEL
(ppm)
rec.
air
vol.(L)
maximum
rate
(L/min)
DS (v/v)

Allyl chloride 1 10 0.2 Benzene
n-Amyl acetate 100 10 0.2 CS2
sec-Amyl acetate 125 10 0.2 CS2
Benzene (OSHA method #12) 10 10 0.2 CS2
Benzyl chloride 1 10 0.2 CS2
Bromoform 0.5 10 0.2 CS2
2-Butoxyethanol (butyl cellosolve) 50 10 0.2 MeCl2/MeOH (95/5)
Butyl acetate 150 10 0.2 CS2
sec-Butyl acetate 200 10 0.2 CS2
tert-Butyl acetate 200 10 0.2 CS2
Butyl alcohol 100 10 0.2 CS2/DMF (99/1)
sec-Butyl alcohol 150 10 0.2 CS2/DMF (99/1)
tert-Butyl alcohol 100 10 0.2 CS2/DMF (99/1)
n-Butyl glycidyl ether (BGE) 50 10 0.2 CS2
p-tert-Butyltoluene 10 10 0.2 CS2
Camphor 2 10 0.2 CS2/MeOH (99/1)
Carbon tetrachloride 10 15 0.1 CS2
Chlorobenzene (monochlorobenzene) 75 10 0.2 CS2
Chloroform (trichloromethane,
   OSHA method #05)
(C) 50 15 0.1 CS2
Chloroprene (2-chloro-1,3-butadiene) 25 10 0.2 CS2
Cumene 50 10 0.2 CS2
Cyclohexane 300 10 0.2 CS2
Cyclohexanol 50 10 0.2 CS2/DMF (99/1)
Cyclohexene 300   5 0.2 CS2
Diacetone alcohol
   (4-hydroxy-4-methyl-2-pentanone)
50 10 0.2 CS2/DMF (99/1)
o-Dichlorobenzene (C) 50   3 0.2 CS2
p-Dichlorobenzene 75   3 0.05 CS2
1,1-Dichloroethane 100 10 e="Arial" size="2">0.2 CS2
1,2-Dichloroethylene 200   3 0.2 CS2
Dichloroethyl ether (C) 15 15 1.0 CS2>
1,1-Dichloro-1-nitroethane** (C) 10 15 1.0 CS2
Difluorodibromomethane (F-12-B2)* 100 10 0.2 IPA
Diisobutyl ketone 50 10 0.2 CS2
N,N-Dimethylaniline 5 10 0.2 CS2
Dioxane (diethylene dioxide) 100 10 0.2 CS2
Dipropylene glycol methyl ether 100 10 0.2 CS2
Enflurane (OSHA method #29)* -- 10 0.1 CS2
Epichlorohydrin 5 20 0.2 CS2
2-Ethoxyethanol (Cellosolve,
   OSHA method #53)
200 10 0.1 MeCl2/MeOH (95/5)
2-Ethoxyethylacetate (Cellosolve
   acetate, OSHA method #53)
100 10 0.1 MeCl2/MeOH (95/5)
Ethyl acetate 400   5 0.2 CS2
Ethyl acrylate 25 10 0.2 CS2
Ethyl alcohol (ethanol) 1000   1 0.05 CS2/DMF (99/1)
Ethyl sec-amyl ketone
   (5-methyl-3-heptanone)
25 10 0.2 CS2/DMF (99/1)
Ethyl benzene 100 10 0.2 CS2
Ethyl bromide 200   5 0.2 IPA
Ethyl butyl ketone (3-heptanone) 50 10 0.2 CS2/DMF (99/1)
Ethylene chlorohydrin** 5 20 0.2 CS2/IPA (95/5)
Ethylene dichloride
   (1,2-Dichloroethane, OSHA
   method #03)
50 10 0.2 CS2
Ethyl ether 400   3 0.2 Ethyl acetate
Ethyl formate 100 10 0.2 CS2
Glycidol (2,3-epoxy-1-propanol) 50 50 1.0 THF
Halothane (OSHA method #29)* -- 10 0.1 CS2
n-Heptane 500   5 0.2 CS2
Hexachloroethane 1 30 0.2 CS2
n-Hexane 500   5 0.2 CS2
2-Hexanone (MBK) 100 10 0.2 CS2
Hexone (MIBK) 100 10 0.2 CS2
sec-Hexyl acetate 50 10 0.2 CS2
Isoamyl acetate 100 10 0.2 CS2
Isobutyl acetate 150 10 0.2 CS2
Isobutyl alcohol 100 10 0.2 CS2/DMF (99/1)
Isophorone** 25 10 0.2 CS2/DMF (99/1)
Isopropyl acetate 250 10 0.2 CS2
Isopropyl alcohol 400   3 0.1 CS2/DMF (99/1)
Isopropyl ether 500   3 0.05 CS2
Isopropyl glycidyl ether 50 10 0.2 CS2
Mesityl oxide 25 10 0.2 CS2/MeOH (99/1)
2-Methoxyethanol (methyl
   cellosolve, OSHA method
   #53)
25 10 0.1 MeCl2/MeOH (95/5)
2-Methoxyethylacetate (methyl
   cellosolve acetate, OSHA
   method #53)
25 10 0.1 MeCl2/MeOH (95/5)
Methyl acetate 200   5 0.2 CS2
Methyl acetylene-propadiene*
   mixture (MAPP)
1000   2 0.05 CS2
Methyl acrylate 10   5 0.2 CS2
Methylal (dimethoxymethane) 1000   2 0.05 Hexane
Methylcyclohexane 500   5 0.2 CS2
Methylene chloride* (superseded by
   OSHA methods #59 & #80)
500   2 0.2 CS2
Methyl isobutyl carbinol 25 10 0.2 CS2/DMF (99/1)
a-Methyl styrene (C) 100   3 0.2 CS2
Naphthalene (superseded by
   OSHA method #35)
10 20 0.2 CS2
Octane 500   5 0.1 CS2
Pentane 1000   2 0.05 CS2
2-Pentanone 200 10 0.2 CS2
Phenyl ether 1 10 0.2 CS2
Phenylglycidyl ether 10 50 0.1 CS2
n-Propyl acetate 200 10 0.2 CS2
Propyl alcohol 200 10 0.2 CS2/DMF (99/1)
Propylene dichloride 75 10 0.2 CS2
Propylene oxide 100   5 0.2 CS2
n-Propyl nitrate** 25 50 0.1 CS2
Pyridine 5 50 1.0 MeCl2
Styrene (OSHA method #09) 100 10 0.2 CS2
1,1,1,2-Tetrachloro-2,
   2-difluoroethane
500   2 0.05 CS2
1,1,2,2-Tetrachloro-1,2-
   difluoroethane
500   2 0.05 CS2
1,1,2,2-Tetrachloroethane** 5 10 0.2 CS2
Tetrachloroethylene
   (perchloroethylene)
100 10 0.2 CS2
Tetrahydrofuran 200 10 0.2 CS2
Tetramethyl succinonitrile 0.5 10 0.2 CS2
Toluene 200 10 0.2 CS2
1,1,1-Trichloroethane (methyl
   chloroform, OSHA method #14)
350   3 0.2 CS2
1,1,2-Trichloroethane (OSHA
   method #11)
10 10 0.2 CS2
Trichloroethylene 100 10 0.2 CS2
1,2,3-Trichloropropane 50 10 0.2 CS2
Vinyl chloride (OSHA method
   #04*, superseded by
   method #75)
1   1 0.05 CS2/DMF (99/1)
Vinyl toluene 100 10 0.2 CS2
Xylene 100 10 0.2 CS2

*    Use two charcoal tubes in series for sampling.
**  Use petroleum base charcoal for sampling.

5. References

5.1. "NIOSH Manual of Analytical Methods", ed. 2 Vol. 1-4 National Institute of Occupational Safety and Health, U.S. Government Printing Office, Washington, D.C. (1977-78)