b-Chloroprene
Method number: | 112 |
|
Matrix: | Air |
|
Target concentration: OSHA PEL: ACGIH TLV: |
25 ppm (90.5 mg/m3) TWA, (skin)
25 ppm (90.5 mg/m3) TWA, (skin)
10 ppm (36.2 mg/m3) (skin) |
|
Procedure: | Samples are collected by drawing known volumes of air through sampling
tubes containing 600 mg of Chromosorb 106 in the front section and 300 mg in the back section.
Keep the samples in a refrigerator to help minimize analyte migration. Samples are desorbed
with toluene and analyzed by GC using an electron capture detector. |
|
Recommended air volume and sampling rate: | 6 L at 50 mL/min (120 min) |
|
Reliable quantitation limit: | 22 ppb (80 µg/m3) |
|
Standard error of estimate at the target concentration: | 7.9% |
|
Status of method: | Evaluated method. This method has been subjected to the established
evaluation procedures of the Organic Methods Evaluation Branch. |
|
Date: September 1998 | Chemist: Yihlin Chan |
Organic Methods Evaluation Branch
OSHA Salt Lake Technical Center
Salt Lake City, UT 84115-1802
1. General Discussion
1.1 Background
1.1.1 History
For monitoring occupational exposure to chloroprene, NIOSH specifies sampling with coconut
shell charcoal tubes, desorption with carbon disulfide, and analysis by GC with a flame
ionization detector (FID) (Ref. 5.1). Work at OSHA SLTC, however, has
shown that the coconut shell charcoal tube does not retain chloroprene well
(Ref. 5.2). Also, the method is not very sensitive. The present work was
undertaken to develop a sampling and analytical method that is more sensitive.
Chloroprene auto-oxidizes easily, polymerizes spontaneously at room temperature, and forms
cyclic dimers on prolonged standing even in the presence of polymerization inhibitors
(Ref. 5.3). Attempts were made to convert chloroprene to a stable
derivative. The reagents investigated included bromine, hydrogen bromide, tetracyanoethylene,
and 7,7,8,8-tetracyanoquinonedimethane. None were successful because the reactions
were not quantitative. The chlorine atom on the 1,3-butadiene strongly enhances the
free-radical activity (polymerization) of the molecule and decreases its activity
in ionic and Diels-Alder reactions.
In 1987, a Chinese scientist reported a method of determining the concentration of chloroprene
in air down to 0.01 mg/m3 (3 ppb) (Ref. 5.4).
He sampled with tubes containing Chromosorb 101 and thermally desorbed the analyte directly into
a GC column. Because the sample is not diluted with solvents, sensitivity is greatly enhanced -
1000-fold, assuming 1-mL extraction and 1-µL
injection. Unfortunately SLTC does not have a thermal desorption unit equipped with an
autosampler at this time. We decided on a conventional solvent desorption method. The target
concentration was set at 25 ppm, the current OSHA PEL.
Chromosorb 106 was selected as the sampling media for its high surface area (700 to 800
m2/g versus less than 50 m2/g for
Chromosorb 101). The selection was also based on the consideration that the method may be
adapted to thermal desorption in the future. The sensitivity was increased 400-fold
by using an electron capture detector (ECD) instead of an FID.
There was concern for the purity of chloroprene used to prepare analytical standards because
chloroprene is unstable. NIOSH used freshly distilled chloroprene in their method.
(Ref. 5.1). We recommend the same. There are two commercially available
sources of chloroprene: Chem Service (45% in xylene) and Alfa/Aeser (50% in xylene). When
materials from these two suppliers were compared with freshly distilled chloroprene, their
nominal concentrations were within experimental error. They were found to be stable when stored
in a freezer (-13 °C) during the time of this work (two months).
1.1.2 Toxic effects (This section is for information only and should not be taken as the basis of OSHA policy.)
Chloroprene is much more toxic to rodents than butadiene or isoprene. Toxic effects in humans
from acute, high-level, chloroprene exposures have been reported for the liver,
circulatory system, hematopoietic system, central and peripheral nervous systems, immune system,
reproductive system, and the periodontia. Symptoms include headache, irritability, cardiac
palpitations, dizziness, insomnia, fatigue, respiratory irritations, chest pains, gastrointestinal
disorders, dermatitis, temporary hair loss, conjunctivitis, and corneal necrosis. There are
marked disparities between the studies reported by U.S. and Russian investigators of the
toxicity and hazards posed by exposure to chloroprene. Although two Russian epidemiological
studies suggested that exposure to chloroprene was related to increased incidences of skin and
lung cancer, both the International Agency for Research on Cancer (Ref. 5.5)
and the American Conference of Governmental Industrial Hygienist (Ref. 5.6)
have concluded that chloroprene is not classifiable as to its carcinogenicity to humans. The
NIOSH criteria document also stated "The presently available data appear to be insufficient to
formulate firm conclusions on the carcinogenicity of chloroprene." (Ref. 5.7).
1.1.3 Workplace exposure
Most chloroprene is polymerized to make polychloroprene (neoprene), a synthetic rubber used in
wire and cable covers, gaskets, automotive parts, adhesives, caulks, flame-resistant
cushioning, and other applications requiring chemical, oil, and weather resistance, or high gum
strength. (Ref. 5.3.) The major worker exposure to chloroprene occurs in the
manufacture of the monomer and during the polymerization to neoprene latex. It occurs chiefly by
inhalation of the vapor and skin contact with the liquid. (Ref. 5.7).
1.1.4 Physical properties and other descriptive information (Ref. 5.2)
CAS no.: | 126-99-8 |
|
synonyms: | 2-chloro-1,3-butadiene; chloroprene |
|
formula: | C4H5Cl |
|
formula weight: | 88.54 |
|
appearance: | clear liquid |
|
boiling point: | 59.4 °C |
|
melting point: | -128 to -132 °C |
|
density: | 0.9585 g/mL at 20 °C 0.9591 g/mL at -12 °C 0.9148 g/mL (50% in xylene, at -12 °C) 0.9106 g/mL (calculated for 45% in xylene, at -12 °C) |
|
flash point: | -20 °C (ASTM open cup) |
|
solubility: | slightly soluble in water (<1%) and miscible with most organic solvents |
|
vapor pressure (p): | log10p = 6.652 - 1545/T (T in K, p in kPa) |
|
reactivity: | readily forms dimers and oxidizes at room temperature |
|
structure: |  |
The analyte air concentrations throughout this method are based on the recommended sampling and
analytical parameters. Air concentrations listed in ppm are referenced to 25 °C and 101.3 kPa (760 mmHg).
1.2 Limit defining parameters
1.2.1 Detection limit of the analytical procedure
The detection limit of the analytical procedure is 2.4 pg. This is the amount of analyte that
will give a response that is significantly different from the background response of reagent
blank. (Sections 4.1 and 4.2)
1.2.2 Detection limit of the overall procedure
The detection limit of the overall procedure is 0.14 µg per sample (6.6 ppb,
24 µg/m3). This is the amount of analyte spiked on the sampler
that will give a response that is significantly different from the background response of a
sampler blank. (Sections 4.1 and 4.3)
1.2.3 Reliable quantitation limit
The reliable quantitation limit is 0.48 µg per sample (22 ppb,
80 µg/m3). This is the amount of analyte spiked on a sampler that
will give a signal that is considered the lower limit for precise quantitative measurements.
(Section 4.4)
1.2.4 Precision (analytical procedure)
The precision of the analytical procedure, measured as the pooled relative standard
deviation over a concentration range equivalent to 0.5 to 2 times the target concentration,
is 0.32%. (Section 4.5)
1.2.5 Precision (overall procedure)
The precision of the overall procedure at the 95% confidence level for the ambient
15-day storage test (at the target concentration) is ±16%. This includes an
additional 5% for sampling error. (Section 4.6)
1.2.6 Recovery
The recovery of chloroprene from samples used in a 15-day storage test remained
above 95%, when the samples were stored at ambient temperature. (Section 4.7)
1.2.7 Reproducibility
Six samples, collected from a controlled test atmosphere of chloroprene, with a draft copy of
this procedure, were submitted for analysis by an SLTC Service Branch. The samples were analyzed
after 2 days of storage at 5 °C. No individual sample result deviated from its theoretical
value by more than the precision reported in Section 1.2.5. (Section 4.8)
2. Sampling Procedure
2.1 Apparatus
2.1.1 A personal sampling pump calibrated to ±5% of the recommended flow rate with the sampling device attached.
2.1.2 Glass sampling tubes (150 mm × 10 mm o.d.) packed with two sections of Chromosorb 106.
The front section contains 600 mg and the back section contains 300 mg. The sections are held in
place with glass wool plugs. For this evaluation, commercially prepared sampling tubes were
purchased from SKC, Inc. (Catalog no. 226-111).
2.2 Reagents
None required.
2.3 Technique
2.3.1 Immediately before sampling, break off the ends of the sampling tube. All tubes should be from the same lot.
2.3.2 Attach the sampling tube to the sampling pump with flexible tubing.
2.3.3 Air should not pass through any hose or tubing before entering the sampling tube.
2.3.4 Cap both ends after sampling. Wrap each sample with a Form OSHA-21 seal.
2.3.5 Record the air volume for each sample.
2.3.6 Submit at least one blank with each set of samples. Blanks should be handled in the same
manner as samples, except no air is drawn through them.
2.3.7 List any compounds that could be considered potential interferences.
2.4 Sampler capacity
The capacity of the front section of the SKC 226-111 sampling tube was tested by
sampling from a test atmosphere of chloroprene (181 mg/m3, 80% RH,
22 °C) at 50 mL/min. The 5% breakthrough volume was determined to be 8.57 L. (Section 4.9)
2.5 Desorption efficiency
2.5.1 The average desorption efficiency for chloroprene from Chromosorb 106 over the range of
0.5 to 2.0 times the target concentration was 101.9%. (Section 4.10.1)
2.5.2 The desorption efficiencies at 0.05, 0.1, and 0.2 times the target concentration were found
to be 100.8%, 101.0%, and 101.1%, respectively. (Section 4.10.1)
2.5.3 Desorbed samples remain stable for at least 24 h. (Section 4.10.2)
2.6 Recommended air volume and sampling rate
2.6.1 The recommended air volume is 6 L at 50 mL/min.
2.6.2 For short-term sampling the recommended air volume is 1.5 L at 100 mL/min.
2.6.3 When short-term samples are collected, the air concentrations equivalent to
the reliable quantitation limits become larger. For example, the reliable quantitation limit is
84 ppb or 0.30 mg/m3 for chloroprene when 1.5 L is collected.
2.7 Interferences (sampling)
There is no known interference for sampling.
2.8 Safety precautions (sampling)
2.8.1 The sampling equipment should be attached to the worker in such a manner that it will not
interfere with work performance or safety.
2.8.2 All safety practices that apply to the work area being sampled should be followed.
3. Analytical Procedure
3.1 Apparatus
3.1.1 A GC equipped with an electron capture detector (ECD). An HP 5890 equipped with an
ECD and an autosampler were used in this evaluation.
3.1.2 A capillary column capable of separating chloroprene and the internal standard
(trichloroethylene) from any interferences. An Rtx-1 column (60 m × 0.32 mm i.d.,
df 1.0 µm) was used in this evaluation.
3.1.3 An electronic integrator or other suitable means of measuring detector response. The
Millennium Chromatography Manager System (Waters) was used in this evaluation.
3.1.4 Glass vials, 4-mL and 2-mL, with poly(tetrafluoroethylene)-lined caps for
desorbing samples and for use in the autosampler.
3.1.5 A dispenser capable of delivering 2.00 mL of desorbing solvent.
3.2 Reagents
3.2.1 Chloroprene. Chloroprene, 50% in xylene, was obtained from Alfa/Aesar. Chloroprene,
45% in xylene, was obtained from Chem Service. They were stored in a freezer at -13 °C.
3.2.2 Toluene. Toluene, b&j high purity solvent grade, was obtained from Baxter.
3.2.3 Trichloroethylene. Trichloroethylene, 99.5+%, was obtained from Aldrich Chemical.
3.2.4 Desorbing solvent. Dilute 5.0 µL of trichloroethylene with 1000 mL of toluene.
3.3 Standard preparation
3.3.1 Distill chloroprene from its xylene solution (Section 4.12).
Determine the concentration (w/w) of the 50% chloroprene in xylene using the freshly distilled
chloroprene as reference. The 50% chloroprene solution should be stored in a freezer and its
concentration checked within a month of analysis. Determine the density of the solution if it
is measured by volume instead of weight. Alternatively, use the reference standard of chloroprene
supplied by Chem Service whose concentration is guaranteed to be within ±0.5% prior to the expiration date.
3.3.2 Prepare analytical standards by diluting the reference standards with the desorbing solvent.
Prepare fresh analytical standards daily. A 272-µg/mL standard solution
corresponds to the target concentration.
3.4 Sample preparation
3.4.1 Transfer the sorbent of the front and the back section to separate 4-mL glass vials.
3.4.2 Add 2.00 mL of the desorbing solvent to each vial.
3.4.3 Cap the vials and shake them on a shaker for 30 min.
3.4.4. Pour the solution into a 2-mL autosampler vial.
3.5 Analysis
3.5.1 GC conditions
column: | Rtx-1 (60 m × 0.32-mm i.d., 1.0-µm df) |
oven temperature: | 70 °C for 8 min, 30/min to 250 °C, hold for 5 min at 250 °C. |
inlet temperature: | 250 °C |
detector temperature: | 250 °C |
injection size: | 1.0 µL |
retention time: | 4.8 min (chloroprene) 7.0 min (trichloroethylene) |
carrier gas (H2): | 1.8 mL/min |
make-up gas (N2): | 45.3 mL/min |
split vent: | 22.7 mL/min |
split ratio: | 13.7 : 1 |
Figure 3.5.1. Chromatogram at target concentration. 1 = chloroprene, 2 = internal standard.
3.5.2 An internal standard calibration method is used. A calibration curve can be constructed by
plotting concentration of the analyte versus ISTD-corrected response of standard
injections. Bracket the samples with freshly prepared standards.
| |
Figure 3.5.2. Calibration curve of chloroprene from the data of Table 4.5.
|
3.6 Interferences (analytical)
3.6.1 Any compound that produces an ECD response and has a similar retention time as the analyte
or the internal standard is a potential interference. If any potential interferences are
reported, they should be considered before samples are desorbed. Generally, chromatographic
conditions can be altered to separate an interference from the analyte.
3.6.2 Any compound that affects the ECD response is a potential interference. The oven
temperature program in Section 3.5.1 should be followed to remove any late
eluting peak after each injection.
3.6.3 When necessary, the identity or the purity of an analyte peak may be confirmed with
additional analytical data (Section 4.11).
3.7 Calculations
The amount (in micrograms) of chloroprene per milliliter is obtained from the calibration curve.
This amount is corrected by subtracting the amount (if any) found in the blank. The air
concentration is calculated using the following formula.
mg/m3 =
| |
(microgram per mL) × (desorption volume, mL)
(liters of air sampled) × (desorption efficiency)
|
where: | desorption volume = 2.00 mL desorption efficiency = 1.05 MW = 88.54 |
3.8 Safety precautions (analytical)
3.8.1 Follow the rules set down in your Chemical Hygiene Plan.
3.8.2 Wear appropriate gloves. Avoid skin contact and inhalation of all chemicals.
3.8.3 Wear safety glasses and a lab coat at all times while in the lab area.
4. Backup Data
4.1 Determination of detection limits
Detection limits (DL), in general, are defined as the amount (or concentration) of analyte that
gives a response (YDL) that is significantly different (three standard
deviations (SDBR)) from the background response (YBR).
YDL - YBR = 3(SDBR)
The direct measurement of YBR and SDBR in
chromatographic methods is typically inconvenient and difficult because YBR
is usually extremely low. Estimates of these parameters can be made with data obtained from the
analysis of a series of analytical standards or samples whose responses are in the vicinity of
the background response. The regression curve obtained for a plot of instrument response versus
concentration of analyte will usually be linear. Assuming SDBR and the
precision of data about the curve are similar, the standard error of estimate (SEE) for the
regression curve can be substituted for SDBR in the above equation.
The following calculations derive a formula for DL:
Yobs = |
observed response |
Yest = |
estimated response from regression curve |
n = |
total no. of data points |
k = |
2 for a linear regression curve |
At point YDL on the regression curve
YDL = A(DL) + YBR
A = analytical sensitivity (slope)
therefore
Substituting 3(SEE) + YBR for YDL gives
4.2 Detection limit of the analytical procedure (DLAP)
The DLAP is measured as the mass of analyte actually introduced into the chromatographic column.
Ten analytical standards of chloroprene whose concentrations were equally spaced from 0 to 0.532
µg/mL were prepared. The standard containing 0.532 µg/mL represented approximately 10 times
the baseline noise. These solutions were analyzed with the recommended analytical parameters
(1-µL injection, 13.7 : 1 split). The data obtained were used to determine the
required parameters (A and SEE) for the calculation of the DLAP. Values of
4.92 × 10-5 and 1.19 × 10-4 were obtained
for A and SEE respectively. DLAP was calculated to be 2.4 pg on column or 0.033 µg/mL.
Table 4.2 Detection Limit of the Analytical Procedure for Chloroprene
|
concentration (µg/mL) |
mass on column (pg) |
response × 10-6 |
|
0.000 0.053 0.106 0.159 0.213 0.266 0.319 0.372 0.425 0.478 0.532 |
0.0 3.9 7.7 11.6 15.5 19.4 23.3 27.2 31.0 34.9 38.8 |
0 370 544 612 875 1103 1550 1351 1623 1739 2024 |
|
|

Figure 4.2. Plot of data to determine the DLAP. |
* response = (peak area chloroprene)/(peak area ISTD) |
4.3 Detection limit of the overall procedure (DLOP)
The DLOP is measured as mass per sample and expressed as equivalent air concentration, based
on the recommended sampling parameters. Ten samplers were spiked with chloroprene ranging
from 0 to 1.06 µg. The latter amount, when spiked on a sampler, would produce a peak
approximately 10 times the baseline noise for a sample blank. These samples were analyzed with
the recommended analytical parameters, and the data obtained used to calculate the required
parameters (A and SEE) for the calculation of the DLOP. Values of
1.94 × 10-3 and 9.32 × 10-5 were obtained
for A and SEE respectively. DLOP was calculated to be 0.14 µg/sample
(24 µg/m3 or 6.6 ppb).
Table 4.3 Detection Limit of the Overall Procedure for Chloroprene
|
mass per sample (µg) |
response × 10-6 |
|
0.000 0.106 0.213 0.319 0.425 0.532 0.638 0.744 0.851 0.957 1.06 |
0 0 379 567 819 913 1108 1561 1532 1716 2017 |
|
| |

Figure 4.3. Plot of data to obtain DLOP and RQL. |
* response = (peak area chloroprene)/(peak area ISTD) |
4.4 Reliable quantitation limit
The RQL is considered the lower limit for precise quantitative measurements. It is determined
from the regression line data obtained for the calculation of the DLOP (Section 4.3),
providing at least 75% of the analyte is recovered. The RQL is defined as the amount of analyte
that gives a response (YRQL) such that
YRQL - YBR = 10(SDBR)
therefore
| |

Figure 4.4. Chromatogram near RQL. 1 = chloroprene, 2 = trichloroethylene, 3 = unknown impurity.
|
RQL = 0.48 µg per sample (80 µg/m3 or 22 ppb)
Recovery at this level is 88%.
4.5 Precision (analytical method)
The precision of the analytical procedure is defined as the pooled relative standard deviation
(RSDP). Relative standard deviations were determined from six
replicate injections of analytical standards at 0.5, 0.75, 1, 1.5, and 2 times the target
concentration. After assuring that the RSDs satisfy the Cochran test for homogeneity at the
95% confidence level, RSDp was calculated.
Table 4.5 Instrument Response to Chloroprene
|
× target concn µg/mL |
0.5 × 137 |
0.75 × 205 |
1 × 274 |
1.5 × 411 |
2 × 548 |
|
Response* |
0.4174 |
0.6209 |
0.7990 |
1.1582 |
1.5048 |
|
0.4173 |
0.6210 |
0.7982 |
1.1585 |
1.5091 |
|
0.4177 |
0.6230 |
0.7997 |
1.1562 |
1.5225 |
|
0.4171 |
0.6217 |
0.8051 |
1.1555 |
1.5224 |
|
0.4177 |
0.6218 |
0.8065 |
1.1602 |
1.5121 |
|
0.4177 |
0.6220 |
0.8063 |
1.1599 |
1.5130 |
 |
0.4175 |
0.6217 |
0.8025 |
1.1581 |
1.5140 |
SD |
0.0003 |
0.0008 |
0.0039 |
0.0019 |
0.0072 |
RSD % |
0.06 |
0.12 |
0.49 |
0.16 |
0.47 |
* response = (peak area chloroprene)/(peak area ISTD) |
The Cochran test for homogeneity requires the calculation of the g statistics according to the
following formula:
g =
|
largest RSD2
RSD20.5 × +
RSD20.75 × +
RSD21 × +
RSD21.5 × +
RSD22 ×
|
= 0.4758
|
The critical value of the g statistic, at the 95% confidence level, for five variances,
each associated with six observations, is 0.5065. Because the g statistic obtained
(0.4758) does not exceed this value, the RSDs within each level can be considered equal and they
can be pooled (RSDP) to give an estimated RSD for the concentration range studied.
4.6 Precision (overall procedure)
The precision of the overall procedure is determined from the storage data in
Section 4.7. The determination of the standard error of estimate
(SEER) for a regression line plotted through the graphed storage data
allows the inclusion of storage time as one of the factors affecting overall precision. The
SEER is similar to the standard deviation, except it is a measure of
dispersion of data about a regression line instead of about a mean. It is determined with the
following equation:
|
n =
k =
k = |
total no. of data points
2 for linear regression
3 for quadratic regression |
Yobs = |
observed % recovery at a given time |
Yest = |
estimated % recovery from the regression line at the same given time |
An additional 5% for pump error (SP) is added to the SEER by the
addition of variances to obtain the total standard error of estimate.
The precision at the 95% confidence level is obtained by multiplying the standard error of
estimate (with pump error included) by 1.96 (the z-statistic from the standard
normal distribution at the 95% confidence level). The 95% confidence intervals are drawn about
their respective regression lines in the storage graphs, as shown in Figures
4.7.1 and 4.7.2. The precision of the overall
procedure of ±16% was obtained from Figure 4.7.1.
4.7 Storage test
Storage samples were prepared by drawing a controlled test atmosphere (80% relative humidity and
22°C) through samplers at 50 mL/min for 120 min. The concentration of chloroprene was
approximately at the target concentration. Thirty-six samples were prepared. Six
samples were analyzed on the day of preparation. One-half of the remaining samples
were stored in a refrigerator (5°C), and the other half were stored at ambient temperature
(about 22°C) in a closed drawer. At 2-4 day intervals, three samples were
selected from each of the two sets and analyzed. The results were compared to those of day 0.
Because a few of the samples were lost, the experiment was repeated with another batch of 18
samples. Some amount of analyte (up to ~15%) was found in the back section of the day-15
and day-13 samples stored at ambient temperature. No chloroprene was detected in
the back sections of the samples stored in a refrigerator.
Table 4.7 Storage Test for Chloroprene
|
time (days) |
percent recovery (ambient) |
|
percent recovery (refrigerated) |
|
0 |
lost |
100.7 |
99.3 |
|
lost |
100.7 |
99.3 |
0 |
100.1 |
99.9 |
lost |
|
100.1 |
99.9 |
lost |
0 |
100.4 |
99.3 |
100.3 |
|
100.4 |
99.3 |
100.3 |
3 |
101.3 |
103.8 |
100.1 |
|
102.2 |
lost |
102.9 |
4 |
96.8 |
|
|
|
99.5 |
99.4 |
|
6 |
97.5 |
97.0 |
74.6* |
|
100.3 |
100.9 |
98.6 |
7 |
95.1 |
96.8 |
|
|
101.0 |
|
|
9 |
97.5 |
89.9 |
95.5 |
|
100.6 |
53.5* |
101.9 |
11 |
83.4 |
|
|
|
96.8 |
89.2 |
|
13 |
87.4 |
83.7 |
81.1 |
|
92.0 |
93.1 |
92.2 |
13 |
100.1 |
100.8 |
|
|
95.7 |
|
|
15 |
105.1 |
102.2 |
104.0 |
|
99.7 |
104.1 |
99.1 |
15 |
99.0 |
|
|
|
96.3 |
98.0 |
|
Lost = tube broken or tubing came off during sampling *outlier, not used |
4.8 Reproducibility
Reproducibility samples were prepared by collecting them from a controlled test atmosphere
similar to that used in the storage test. The samples were submitted to an SLTC Service Branch
for analysis. The samples were analyzed after being stored for 2 days at 5°C. No sample result
had a deviation greater than the precision of the overall procedure determined in Section 4.7.
Table 4.8 Reproducibility Data for Chloroprene
|
ppm expected |
ppm found |
percent found |
percent deviation |
|
26.7 |
25.6 |
95.9 |
-4.1 |
26.7 |
26.0 |
97.4 |
-2.6 |
26.7 |
26.0 |
97.4 |
-2.6 |
26.7 |
25.4 |
95.1 |
-4.9 |
26.7 |
25.6 |
95.9 |
-4.1 |
26.7 |
26.4 |
98.9 |
-1.1 |
|
4.9 Sampler capacity
The capacity of the front section of a Chromosorb 106 tube (SKC 226-111) was tested
by sampling from a test atmosphere of chloroprene at two times the target concentration
(181 mg/m3, 80% relative humidity at 22°C). The flow rate was
0.05 L/min. The downstream air was monitored at 10-minute interval with a GC
equipped with a gas sampling valve. The 5% breakthrough point was determined to be 8.57 L.
Table 4.9 Breakthrough of Chloroprene with Chromosorb 106 Sampling Tubes
|
Tube 1 | Tube 2 |
|
air volume (L) |
breakthrough (%) |
air volume (L) |
breakthrough (%) |
|
5.93 6.44 6.95 7.47 7.98 8.49 9.00 |
0.00 0.00 0.00 0.00 4.49 6.46 12.20 |
6.15 6.66 7.18 7.69 8.20 8.71 9.22 |
0.00 0.00 0.00 0.00 1.55 4.90 11.41 |
|
| |
Figure 4.9. Breakthrough curve for chloroprene on Chromosorb 106 at two times the target concentration.
|
4.10 Desorption efficiency and stability of desorbed samples
4.10.1 Desorption efficiency
The desorption efficiencies (DE) test samples were prepared by liquid-spiking
the 600-mg portions of Chromosorb 106 with chloroprene at 0.05 to 2 times the
target concentration. These samples were stored overnight at ambient temperature, desorbed,
and analyzed. The average desorption efficiency over the working range of 0.5 to 2 times the
target concentration was 101.9%.
Table 4.10.1 Desorption Efficiency of Chloroprene
|
× target concn (µg/sample) |
0.05 × 26.6 |
0.1 × 53.2 |
0.2 × 106 |
0.5 × 266 |
1.0 × 532 |
2.0 × 1063 |
|
DE (%) |
101.0 |
100.5 |
100.8 |
101.3 |
101.0 |
102.4 |
|
101.2 |
100.5 |
101.4 |
100.9 |
103.5 |
102.7 |
|
100.8 |
100.3 |
101.8 |
101.1 |
101.8 |
102.5 |
|
100.7 |
101.4 |
101.3 |
101.1 |
102.0 |
102.8 |
|
100.3 |
101.7 |
101.2 |
100.9 |
102.2 |
103.0 |
|
100.8 |
101.5 |
99.9 |
100.5 |
102.0 |
103.0 |
Average |
100.8 |
101.0 |
101.1 |
101.0 |
102.1 |
102.7 |
|
4.10.2 Stability of desorbed samples
The stability of the desorbed samples was investigated by re-analyzing the
one-times-the-target-concentration samples 24 hours later. After the desorption
efficiency analysis was performed, three vials were recapped with new septa while the remaining
three retained their punctured septa. The samples were re-analyzed with fresh
standards. The average percent change was -2.7 for the re-capped samples and -3.7
for those that were not re-capped.
Table 4.10.2 Stability of Desorbed Samples of Chloroprene
|
punctured septa replaced |
punctured septa retained |
|
initial DE (%) |
DE after one day (%) |
difference |
initial DE (%) |
DE after one day (%) |
difference |
|
101.0 |
99.0 |
-2.0 |
102.0 |
98.2 |
-3.8 |
103.5 |
99.2 |
-4.3 |
102.2 |
98.5 |
-3.7 |
101.8 |
99.9 |
-1.9 |
102.0 |
98.4 |
-3.5 |
|
average |
|
|
average |
|
102.1 |
99.4 |
-2.7 |
102.1 |
98.4 |
-3.7 |
|
4.11 Qualitative analysis
Chloroprene may be confirmed by GC/MS using GC conditions similar to those in Section 3.5.1.

Figure 4.11. Mass spectrum of 2-chloro-1,3-butadiene.
4.12 Distillation of chloroprene
Chloroprene was distilled at atmospheric pressure (650 mmHg). The cut boiling between 62°C
and 66°C was collected. Freshly distilled chloroprene was stored at -15°C
and used within two days. Chloroprene can also be distilled at reduced pressure (Ref. 1).
5. References
5.1. NIOSH Manual of Analytical Methods,
Fourth Edition, U.S. Department of Health and Human Services, Center for Disease Control,
National Institute for Occupational Safety and Health, Cincinnati, OH, DHHS (NIOSH)
Publication No. 94-113, 1994.
5.2. Eide, M. OSHA SLTC, Salt Lake City, UT, A Study of the Detection
Limits of b-Chloroprene Samples Collected on TBC Coated
Charcoal Tubes Using Flame Ionization and Electron Capture Detectors, unpublished, 1997.
5.3. Johnson, P. R. in Kirk-Othmer Encyclopedia of Chemical Technology,
3rd edition, 1982, Vol. 5, 773-785.
5.4. Wang, P. Chromatographical Determination of Traces of
2-Chloro-1,3-butadiene in the Atmosphere with Chromosorb 101 Trapping Tube,
China Synthetic Rubber Industry, 1987, 1(10), 17.
5.5. IARC Monographs on the Evaluation of the Carcinogenic Risk of Chemicals to Humans,
International Agency for Research on Cancer, Lyon, 1979, 19, 131-146.
5.6. Documentation of the Threshold Limit Values and Biological Exposure Indices,
6th Edition, 1991, American Conference of Governmental Industrial Hygienist, Cincinnati, OH.
5.7. NIOSH criteria for a recommended standard ... occupational exposure to Chloroprene,
U.S. Department of Health, Education, and Welfare, Center for Disease Control, National Institute for
Occupational Safety and Health, Cincinnati, OH DHEW (NIOSH) Publication No. 77-210, 1977.
|