VANADIUM PENTOXIDE (ID-185) BACK-UP REPORT
This backup report was revised April, 1991
Introduction
The general procedure for the collection and X-ray confirmation of
vanadium pentoxide (V2O5) exposures is
given in OSHA Method ID-185
(9.1.). The general procedure for the collection and analysis of air
samples for V2O5 is given in OSHA
Method ID-125G
(9.2.). The sampling technique and analytical instrumentation of
these two methods differ in both detail and purpose. The inductively coupled plasma atomic
emission spectroscopy (ICP-AES) approach in ID-125G is an
elemental analysis and cannot identify the vanadium-containing compound. OSHA
Method ID-185 is used when there is doubt as to the specific V compound that
is the source of the V exposure.
This method was evaluated when the OSHA Permissible Exposure Limit (PEL) was a ceiling
value and was for exposures to total dust or fume. Currently, the V2O5
PEL is a time weighted average (TWA) for either a respirable dust or fume.
This back-up report consists of the following sections:
(1) Experimental procedure
(2) Analysis
(3) Determination of the accuracy
and precision
(4) Determination of detection limits
(5) Effect of particle-size
distribution on X-ray recoveries
(6) Method comparison
(7) Summary of results
(8) Conclusions
Notes: 1) This method is for confirmation use, where heavier sample loadings are
expected, and therefore larger amounts were used in the validation than might be expected
in a 8-h TWA sample. 2) The evaluation was additionally designed to assess
the effect of particle size on the analytical accuracy of X-ray analyses V2O5.
1. Experimental Procedure
1.1. Two analytical X-ray techniques were investigated and compared
against an atomic absorption spectrometry (AAS) procedure for analyzing V.
1.1.1. X-ray diffraction (XRD) was performed using custom OSHA Laboratory software and
two Model 3500 Automated Powder Diffractometers (APDs) (Phillips Electronics Instruments
Co., Mahwah, NJ) sharing the same generator. Results for these APDs are labeled APD-A
or APD-B in this report.
1.1.2. X-ray fluorescence (XRF) was performed using a Model 77-800
(upgraded to Model 77-900A) Energy Dispersive X-ray Fluorescence
(EDXRF) Spectrometer (Finnigan Corporation, Sunnyvale, CA) consisting of: a an X-ray
generator with a direct-beam Rh end-window X-ray
tube and an X-ray spectrometer console using a Computer Automation Alpha 16
computer.
More recently, field samples submitted to the Laboratory for V2O5
analyses have been analyzed using a Kevex 770/8000 Delta EDXRF system (Kevex Instruments
Inc., San Carlos, CA) consisting of: Kevex 770 X-ray generator, its
associated satellite box, vacuum system, helium flush system, firmware-based
8000 keyboard console, computer monitor, Digital Equipment Corporation (DEC) 11/73
computer, graphics memory, Kevex spectrum analyzer, and Toolbox II software. This latter
system uses an Fe secondary target for this analysis and offers improved sensitivity,
lower background and greater resolution of interferences (9.1.). The
system parameters for both XRF systems are given in the experimental section of the method
(9.1.).
1.1.3. For the purpose of comparison, the V2O5
samples on Ag membranes were re-analyzed for V by AAS procedure (9.3.) using a Model 603 Atomic Absorption Spectrometer (Perkin-Elmer
Corp., Norwald, CT).
1.2. Three particle-size distributions were investigated:
- "M" samples simulating respirable particle size range (0 to 10 µm with median
of approximately 3 µm). The "M" samples are referred to by the label
"Respirable".
- "W" samples simulating fine-respirable particle size range (0 to
3 µm). The "W" samples are referred to by the label "Fine
respirable".
- "Fine W" samples simulating fume-like particles.
1.2.1. Quantitation of V2O5 dust
(approximating respirable dust) was assessed using spikes at nominal levels of 233, 467
and 700 µg V2O5 ("M"
samples).
1.2.2. Quantitation of V2O5 dust
(approximating fine-respirable particle size) was assessed using spikes at
levels of 237, 474, and 710 µg V2O5
("W" samples). The spiked levels indicated in the NIOSH study were duplicated
and are not the usual validation levels of 0.5, 1, and 2 times the OSHA PEL. The
respirable characterization and spiking levels are further described in Reference 9.4.
1.2.3. To assess analyte sensitivities to very small fume-like particles,
aliquots of an unstirred acetonitrile suspension of the finest particles in the
"W" material were analyzed (See Section 2.4.2.).
1.2.4. All samples were analyzed by XRD, XRF, and AAS techniques.
2. Analysis
2.1. Filter Membranes
2.1.1. FWS-D (0.5 µm pore size) membranes were spiked with sonicated acetonitrile
suspensions of V2O5. Small pore size
membranes were used during the evaluation to prevent the loss of small particles during
the spiking with liquid suspensions. During sample collection of particles suspended in
air, the 5.0-µm pore size PVC membrane should sufficiently retain the
smaller particles due to static charge and collection characteristics. Collection
efficiencies exceeding 99% have been reported of 0.3-µm Dioctyl phthalate
aerosol collection on PVC filters (9.5.).
2.1.2. Silver metal membranes (25-mm diameter, 0.45-µm pore size) were used to support
the prepared thin films for presentation in X-ray analyses.
2.2. Preparation of Standard Materials
Procedure:
2.2.1. Reagent grade V2O5 (99.8%,
J.T. Baker, Phillipsburg, NJ) was used as the starting material for the X-ray
methods. For the AAS comparison method, a 1,000 µg/mL V standard in a dilute HCl matrix
(Lot #J141, RICCA Chemical Company, Arlington, TX) was used for preparing AAS standards.
2.2.2. Respirable stock material ("M") was prepared by hand grinding reagent
grade V2O5 in an alumina mortar and
pestle at room temperature. The ground material was added to 50 to 75 mL of
tetrahydrofuran (THF) in a glass beaker. The sonicated suspension was west-sieved
through 10-µm nylon mesh using a sieving bottle as shown and described in
Section 6.4. of the method (9.1.). The dust was isolated from the
suspension by filtering it onto a Ag membrane. This material was used in preparing only
spikes.
2.2.3. For a finer particle-size distribution, additional stock material
("W") was prepared by a different grinding technique. Reagent grade V2O5
was ground in a freezer mill operated with liquid nitrogen for 10 minutes. This stock
material was also wet-sieved through 10-µm nylon mesh as
described in Section 2.2.2. In addition to spiked sample preparation, this material was
used to prepare calibration standards because the freezer mill produces more reproducible particle-size
distributions than those produced from hand grinding with a mortar and pestle.
2.2.4. Results: The reagent grade material,
"W", and "M" materials were analyzed by AAS and gave an assay of
101.3% V2O5 as shown below:
Reagent
|
"W" Material
|
"M" Material
|
mg Taken
|
Assay by AAS
|
mg Taken
|
Assay by AAS
|
mg Taken
|
Assay by AAS
|
1.307 |
104.59% |
2.263 |
101.59% |
2.069 |
102.66% |
1.633 |
101.84 |
1.581 |
101.39 |
1.535 |
103.45 |
2.348 |
99.96 |
2.295 |
99.48 |
5.663 |
97.30 |
2.457
|
98.82
|
1.297
|
100.93
|
2.138
|
100.09
|
Mean |
101.3% |
|
100.8% |
|
100.9% |
|
The manufacturer's assay of 99.8% was used in calculations for all materials derived
form the reagent grade V2O5.
2.3. Preparation and XRD Analysis of Calibration Standards
Procedure:
2.3.1. X-ray calibration standards preparation
Twenty-four calibration standards were prepared from THF suspensions of fine-respirable
stock material ("W"). To avoid reduction of V2O5
by warm organic agents, fixatives were not used to secure the dust on the membranes. This
also facilitated sample and standard re-analysis by AAS. Three standards were
prepared on Ag membranes at each of the following levels:
"W" Material - Standards
Standard
Delivered
µg V2O5
|
Volume (mL)
|
Reagent
Concentration
µg/mL
|
50
100 200 250 499 998 1996 2495 |
5
10
2
25
5
10
20
25 |
10.040
10.040
100.02
10.040
100.02
100.02
100.02
100.02 |
2.3.2. Calibration and analysis (XRD)
These calibration standards were analyzed on two different APDs. The sample analysis
order was scrambled to reduce instrumental drift effects. Two-theta
calibrations were performed using the primary 38.15° two-theta (2q) Ag calibration line to avoid potential interferences from other V2O5
lines. (In practice, the strong secondary 44.33° 2q Ag line is
within 0.03° 2q of a low intensity V2O5
peak; therefore, no significant error is introduced when using the secondary Ag line in 2q calibrations.) The X-ray generator settings were 40 kV
and 40 mA. Integration times of 1 s and 0.02° 2q steps were
used throughout the study. The detection limit was determined at both 1- and 10-s
integration periods. A custom OSHA computer program (9.6.) was used
to establish a two-piece calibration curve and calibration coefficients.
Standard data is shown in Table 1. A two piece curve fit or a
polynomial curve fit is generally performed to optimize recovery over the entire
analytical range. A second-order polynomial fit for the low end allows for
the correction of losses due to penetration of dust into the Ag membrane and to correct
for the conservative heuristic used when establishing integration limits. The second-order
fit of the upper end is smoothly spliced onto the upper end of the low range fit. The second-order
fit of the upper end allows for partial correction for sample self-absorption
effects. This is important because a Leroux correction (9.7.) is not
performed. In the past, Leroux corrections have been found to over-correct (9.8.).
2.4. Preparation and X-ray Analyses of Spiked Samples
Procedure:
2.4.1. Sample preparation
Analyses were performed on a total of 36 samples (six samples at each of three test
levels for the two different materials, "W" and "M"). Acetonitrile was
used as the vehicle for spiking FWS-D membrane filters. Neither V2O5
nor the PVC filter medium dissolves appreciably in acetonitrile. The "W" spiked
samples were prepared by filtration of acetonitrile suspensions of the freezer mill
material upon FWS-D (0.5-µm pore size) filters supported on a fritted-glass
filtering support. The "M" spiked samples were prepared by similar filtration of
acetonitrile suspensions of the mortar-and-pestle ground material. Upon
drying, the filter membranes were transferred to centrifuge tubes and the 10 mL of THF was
added to each to dissolve the membranes. The tubes were placed in an ultrasonic bath and
the tube contents were sonicated for approximately 10 minutes. The sonicated suspension
was filtered onto 25-mm Ag membranes (0.45-µm pore size) for
subsequent analysis. As in the preparation of the calibration standards (See Section 2.3.1.), fixatives were not used. The three test levels were
produced as follows:
|
Sample Delivered
µg V2O5
|
Aliquot
(mL)
|
Reagent Concentration
µg/mL
|
"W" |
237
474
710 |
2
4
6 |
118.62
118.62
118.62 |
"W" |
233
467
700 |
2
4
6 |
116.92
116.92
116.92 |
2.4.2. Fume-like sample preparation:
In order to assess the effect of very fine particles on recovery, three samples were
prepared (10-mL aliquots) from the center of the same 118.62 µg V2O5/mL
acetonitrile suspension of fine-respirable stock material "W" after
allowing the coarser material to settle out of the unstirred suspension. After 2.5 h, a
significant fraction of the larger particles had settled out leaving a hazy suspension.
Aliquots of the supernatant suspension were spiked directly onto 0.45-µm
pore size Ag membranes. These samples were referred to as "Fine-W"
samples.
2.4.3. The "W", "M", and "Fine-W" Samples
and blanks were analyzed by XRD, XRF, and AAS using the procedures which follow:
XRD analytical procedure
The Ag membrane samples and blanks were analyzed in the same manner as the calibration
standards described in Section 2.3.2.
2.4.4. XRF analytical procedure
The "W" calibration standards were analyzed by XRF using the program
described in Appendix 1. Count data were collected on all channels in
air at 20 kV, 0.5 mA, for 100 s with a narrow collimator and without an X-ray
filter. After analyzing the standards, a concentration-response curve was
prepared to calibrate on the integrated counts in the 17-channel (~0.64 kV)
region spanning the V Ka peak at
4.949 kV. Background counts were estimated using a linear background model between 3.4 to
5.9 kV. The equations obtained (Table 1) were used to calculate the amount of V2O5
present in spiked ("M" and "W") samples.
For the "Fine-W" spiked samples, a separate regression was
performed using single representative "W" standards at each of the loadings.
2.5. Calibration and Analysis (AAS) - Stock Materials and Spiked Samples
Procedure:
2.5.1. Preparation of atomic absorption standards:
Eight standards were prepared in a 4% HCl matrix by serial dilutions of the 1,000
µg/mL V standard. Concentration of the final standards ranged from 100 to 0.1 µg/mL.
2.5.2. Analytical procedure (AAS) - stock materials:
To check the purity of the stock materials used in the evaluation of the X-ray
methods, four samples of each of the reagent, "W", and "M" materials
were weighed out and transferred to 100 mL volumetric flasks using a total of 4-mL
HCl and approximately 5 mL deionized water (DI H2O). These were
brought just to a boil on a hot plate. After cooling, the samples were diluted to volume
with DI H2O to give a 4% HCl matrix. Analysis was performed
according to reference 9.3. The results are shown in Section 2.2.4.
2.5.3. Analytical procedure (AAS) - silver membrane samples:
The blanks and the spiked "W", "M", and "Fine-W"
samples were re-analyzed by AAS after the XRD and XRF analyses. The edges of
the membranes were bent before acid extraction to encourage the free flow of acid above
and below the membrane. The membranes were agitated and sonicated for 10 to 15 s in 250-mL
Phillips beakers containing 10-mL DI H2O and 1 mL
concentrated HCl. After the dust was visually released from the membranes, the beakers
were placed on a hot plate and brought just to a boil. They were then removed to cool to
ambient temperature (20 to 25 °C) while being agitated. The solutions were then
quantitatively transferred to 25-mL volumetric flasks using 4 to 5 small
rinses of DI H2O. The volumetric flasks were then diluted to
volume with DI H2O to give a 4% HCl matrix. Analysis was
performed according to Reference 9.3.
2.6. Results: The results for the three different analytical techniques
and sample materials are presented as follows:
Table
|
Results
|
2
3
4
5
6
7
8
9 |
XRD results for "W" material
XRD results for "M" material
EDXRF results for "W" material
EDXRF results for "M" material
AAS results for "W" material
AAS results for "M" material
Summary - analyses of "Fine-W" material
Summary - analyses of "W" and "M" material |
Tables 8 and 9 contain summary results of the three analytical techniques performed on
each sample.
3.
Determination of the Precision and Accuracy
3.1. Outlier and Bartlett's Tests (XRD and EDXRF)
The calibration data (Table 1) and all of the "W" and
"M" spiked-sample data passed the ASTM test for outliers at the 99%
confidence level (9.9.). All the spiked-sample data
passed Bartlett's test (9.10.), so the results were pooled as
appropriate. Statistical test results are shown below:
Bartlett's Test Results
"W" Reference Material (also used for calibration standards)
Bartlett's variance homogeneity tests: |
Critical Chi-squared value = 9.21 (99% Confidence Level
and N = 3) |
XRD (APD-A)
XRD (APD-B)
EDXRF |
Chi-squared = 1.11
Chi-squared = 2.11
Chi-squared = 9.21 |
N = 3
N = 3
N = 3 |
"M" Reference material (coarser than calibration standards)
Bartlett's variance homogeneity tests: |
Critical Chi-squared value = 9.21 (99% Confidence Level
and N = 3) |
XRD (APD-A)
XRD (APD-B)
EDXRF |
Chi-squared = 0.26
Chi-squared = 0.49
Chi-squared = 2.04 |
N = 3
N = 3
N = 3 |
3.2. The precision and accuracy (9.11.) for the XRD method:
Recoveries, precision, and overall errors are shown below. X-ray
diffraction results for "W" stock material (AAS results are shown in
parentheses) are:
Recovery: |
APD-A Ave. Recovery
APD-B Ave. Recovery
Combined XRD Recovery
|
= 0.894
= 0.880
= 0.887 (0.900)
|
Precision: |
APD-A CV1(Pooled)
APD-B CV1(Pooled)
Combined XRD CV1(Pooled)
|
= 0.117
= 0.125
= 0.121 (0.031)
|
Overall Error: |
= ± 36% |
X-ray diffraction results for "M" stock material (AAS results are shown in
parentheses) are:
Recovery: |
APD-A Ave. Recovery
APD-B Ave. Recovery
Combined XRD Recovery
|
= 1.680
= 1.867
= 1.774 (0.933)
|
Precision: |
APD-A CV1(Pooled)
APD-B CV1(Pooled)
Combined XRD CV1(Pooled)
|
= 0.062
= 0.073
= 0.068 (0.015)
|
Overall Error: |
= ± 91% |
3.3. Precision and Accuracy - XRF method
Recoveries, precision, and overall error are shown below. X-ray
fluorescence results for "W" material (AAS results shown in parentheses) are:
Ave. Recovery:
Precision: CV1(Pooled)
Overall Error: |
|
= 0.871 (0.900)
= 0.097 (0.031)
= ± 32% |
X-ray fluorescence results for "M" material (AAS results shown in
parentheses) are:
Ave. Recovery:
Precision: CV1(Pooled)
Overall Error: |
|
= 0.965 (0.933)
= 0.064 (0.015)
= ± 16% |
4. Determination of Detection Limits
4.1. Procedure: Blanks were analyzed in order to estimate the microgram
detection limits. Blanks were analyzed by XRF using the total analytical times indicated
in Appendix 1. The blanks were also analyzed by XRD as described in Section 2.3.2. using total analytical times of 65 and 650 s
(corresponding to integration times of 1 and 10 s respectively). The X-ray
detection limit estimates were based on the International Union of Pure and Applied
Chemistry (IUPAC) definition as three times the standard deviation of the measurements
performed on blanks divided by the slope (9.12., 9.13.).
The AAS detection limit was estimated using three times the minimum AAS reading.
4.2. Results: Detection limit results are summarized below and shown in Table 10. Detection limits determined for the analytical methods used
(µg V2O5):
X-ray Diffraction |
|
X-ray Fluorescence |
|
Atomic Absoprtion |
DL
|
Total time
|
|
DL
|
Total time
|
|
DL
|
Total time
|
25 µg
25 µg |
65 s
650 s |
|
14 µg
2 µg |
100 s
1,000 s |
|
9 µg
- |
4 s
- |
Some XRD blank results were abnormally high. This gave a large detection limit for the
XRD method. No V was identified using XRF or AAS analyses on the same blanks; therefore, V2O5
contamination was ruled out as a possible cause. A sample of ten Ag membranes from the
same lot also did not have the XRD interference. This indicates that the PVC membranes
and/or the THF solvent may be responsible. Salt (NaCl) has its primary peak near the V2O5
analytical peak. The non-stoichiometric compound, K0.2Na0.8Cl,
has its primary diffraction line at the V2O5
analytical peak. Finger prints which potentially contain salt did not produce significant
peaks in the range scanned.
5. Effect of
Particle-Size Distribution on X-ray Recoveries
Comparisons were performed on the results for the three different particle-size
distributions. Due to the sample preparation method used for the fume-like
samples, the amount of V2O5 taken was
not known beforehand. Using result ratios (Mean Relative Recoveries) allows making a
comparison.
5.1. The data used for this comparison study were taken from Tables 8 and 9.
5.2. The mean relative recoveries for the materials studied:
|
Mean Relative Recoveries
|
|
XRD/AAS
|
XRF/AAS
|
Fume-like particles "Fine-W"
Fine-respirable particles "W"
Respirable dust particles "M" |
0.651
0.986
1.901 |
0.998
0.968
1.034 |
6. Method Comparison
Related to the evaluation of accuracy and precision is method (system) comparison which
normally employs duplicate sampling (or spiking) to holistically compare the quality of a
known analytical system with one or more untested analytical systems. Duplicate spiking (a
separate set for each method) was not performed in these comparisons because the non-destructive
nature of the XRD and XRF analyses made that unnecessary and counter-productive.
Westgard and associates (9.14.) have proposed a detailed scheme for
method comparison. This evaluation scheme calls for the application of a least-squares
linear regression of the results from the candidate method and comparative analytical
method (assumed to be dependent and independent variables respectively). The regression is
then analyzed by statistical techniques such as the F-test, t-test,
least-squares analysis and correlation coefficients. This scheme is based on
the assumption that the comparative method gives the true value (9.15.).
The approach is possibly biased against discovering a better analytical system. In these
analyses, however, the AAS technique should give the most accurate value for the amount of
V2O5 captured on and in the Ag
membrane and is considered the reference method. The statistical evaluation is meaningful
in that limited context. Comparisons of the XRD and XRF candidate methods with the AAS
comparison method are presented below.
6.1. A summary of the AAS versus X-ray comparison data from
the computer calculations follows:
|
a
|
Slope
|
Sslope
|
Bias
|
r
|
r2
|
APD-A "W" dust
APD-A "M" dust |
2.11
-28.02 |
0.9848
1.8826 |
0.057
0.074 |
-4.33
340.92 |
0.9739
0.9886 |
0.9484
0.9773 |
APD-B "W" dust
APD-B "M" dust |
-23.69
-35.38 |
1.0440
2.0999 |
0.063
0.088 |
-4.99
424.44 |
0.9717
0.9870 |
0.9441
0.9743 |
EDXRF "W" dust
EDXRF "M" dust |
-3.69
-23.34 |
0.9779
1.0988 |
0.046
0.041 |
-13.06
17.95 |
0.9838
0.9900 |
0.9679
0.9800 |
Where: |
a (in µg) |
= |
intercept of regression line |
Slope |
= |
slope of regression line |
Sslope |
= |
standard deviation for the slope |
Bias (in µg) |
= |
mean µg V2O5 found by candidate
method less
mean µg V2O5 found by reference
method |
r |
= |
correlation coefficient |
r2 |
= |
Coefficient of determination (fraction of variation in candidate measurements due to
variation in reference measurements) |
The coefficients of determination are between 0.94 and 1.00. This indicates that, at
most, only 6% of the variance in the candidate measurement is not accounted for by
variance in the "independent" reference measurement.
The slopes approaching a value of 1 indicate small relative bias between the candidate
and comparison methods. This is the case in all but the XRD results for the "M"
dust. The slope of approximately 2 indicates an unacceptable relative bias.
6.2. Results (t-tests and F-tests)
|
t
|
t-crit
|
F
|
F-crit
|
df
|
APD-A "W" dust
APD-A "M" dust |
-.4557
9.0564 |
2.110
2.120 |
1.02257
3.62634 |
2.30
2.33 |
17
16 |
APD-B "W" dust
APD-B "M" dust |
-.4683
9.0849 |
2.110
2.120 |
1.15448
4.52635 |
2.30
2.33 |
17
16 |
EDXRF "W" dust
EDXRF "M" dust |
-1.659
2.3854 |
2.120
2.120 |
1.01211
1.23193 |
2.33
2.33 |
16
16 |
Where: |
t |
= |
calculated Student t-test value
(bias indicated by sign) |
t-crit |
= |
two-sided critical t value for 0.05 probability (from Reference 9.16.) |
F |
= |
calculated F-test value |
F-crit |
= |
critical F value for 0.05 probability (from Reference 9.16.) |
df |
= |
degrees of freedom
(no. paired observations - 1) |
For the fine respirable dust ("W") samples, no significant difference was
detected between the performance of the test and comparative methods.
The t-test data above indicate that for the respirable dust
("M") samples there is a significant difference in means between the X-ray
test methods and the AAS comparison method.
In the case of the "M" dust samples, the F-test data indicate
that there was a significant difference in precision between the XRD and AAS methods, but
there was not a significant difference in precision between the XRF and AAS methods.
7. Summary of Results
In order to get the best estimate of the Overall Error, the recoveries and CV1
Pooled results for the two different APDs used in the validation were averaged. Averaging
was not necessary for the XRF results, since only one instrument was used. The ranges
shown are for all experiments performed and are therefore somewhat larger than if only a
single APD instrument were used.
The results for the accuracy and precision calculations in Section 3 were based on the
assumption that the theoretical amount of V2O5
delivered to each PVC membrane represented the true amount. The X-ray
stock material was verified against the AAS standard giving approximately 100% V2O5.
The XRD, XRF, and AAS analyses of the "W" dust samples and the XRF and AAS
analyses of the "M" dust samples agreed well. A 10% negative bias was noted in
recoveries of the "W" dust samples. It was concluded that the negative bias in
these cases was probably due to losses incurred in spiking the PVC membrane with a V2O5
suspension in acetonitrile. Because the same samples were analyzed by XRD, XRF and AAS,
any physical losses incurred in spiking and transfer to the Ag membrane were the same for
each sample regardless of the analytical technique. The recoveries for "M" dust
samples by XRD disagreed considerably with the other methods investigated and followed the
trend expected for larger particles. The process of spiking by means of a suspension does
not duplicate aerosol sampling and may not be ideally representative of samples taken with
a cyclone.
The ability of each analytical technique to accurately determine each V2O5
material was assessed using overall error. The overall error should be within ± 25% and
is calculated using the following equation:
Overall Error = ± ( |mean bias| + 2[CVT
pooled])100%
CV1 pooled was used instead of CVT
pooled in this study. In Section 3 the low end of the range for the
overall error for "M" dust analyzed by XRD was reported as ± 91%. Regardless,
all the XRD work exceeds a 25% cutoff for overall error. Only the "M" dust XRF
results satisfy an overall error limit of ±25%.
8. Conclusions
8.1. X-ray diffraction (XRD) is the method of choice in identifying
V2O5. Due to the significant
dependence of XRD analytical sensitivity upon particle-size distribution, XRD
is only used as a confirmation technique for this analyte. Analytical lines are available
for qualitative verification in addition to the line available for quantitation. In order
to quantitate using XRD, the standard material must approximate the analyte particle-size
distribution of the samples. This may not be practical for the OSHA dust and fume
standards for V2O5. If fume is present
in an operation, the results in Section 5.2. indicate that recovery
may be low due to the reduced XRD analytical sensitivity for finer particles. Therefore,
XRD is used for confirmation only.
8.2. X-ray fluorescence (XRF) is the method of choice in quantitating
V2O5 because particle-size
effects are much less severe for XRF compared to XRD. In XRF, the V Kb peak is also available for qualitative verification
of V.
8.3. Due to the common sample preparation technique and superior performance of the XRF
methodology, this work suggests a hybrid method incorporating quantitation by XRF and
chemical species verification (and semi-quantitative support) by XRD. There
is a potential for multi-analyte analyses by such a hybrid approach. If the
industrial hygienist desired, both respirable V2O5
and respirable quartz [which coexist in certain industrial operations (9.4.)]
could be determined on the same prepared sample if the quartz sampling procedure is
employed (9.17.)
8.4. The XRD method was patterned after the NIOSH study (9.4.). A
discussion of the effects of deviation in V2O5
methodology between OSHA and NIOSH can be found in reference 9.18.
8.5 Concluding remarks:
As noted above, there was good agreement between the AAS and XRF results in this study
for all three particle-size distributions. The major biases observed for the
XRD analyses are most readily explained as due to the change in sensitivity with respect
to particle size.
The results in this report support the proposition that the quantitative analysis of V2O5
by XRD would require close matching of the particle-size distribution of the
standard material to that of air samples collected during industrial operations on PVC
filters. As seen in the samples subjected to removal of coarse particles by sedimentation
("Fine-W"), XRD evidenced decreased analytical sensitivity when
compared to both XRF and AAS. As shown in Section 5.2., there was a
large positive bias in the XRD analyses when the particle-size distribution
was biased towards larger particles. This was most clearly shown when the analyses of V2O5
finely ground in a freezer mill ("W" samples) are compared to the analyses of V2O5
more coarsely ground in a mortar and pestle ("M" samples). The coarse material
provided a doubling of recoveries when compared to the recoveries of the fine material.
The quantitative analysis of V2O5 by
XRF was more immune to the particle-size distribution, thus giving improved
recoveries and better precision than analysis by the XRD method. Aerosol generation and
particle sizing would be advantageous in more fully evaluating these particle-size
effects.
9. References
9.1. Occupational Safety and Health Administration Technical
Center: Confirmation of Vanadium Pentoxide in Workplace Atmospheres by M.C.
Rose (USDOL/OSHA-SLTC Method No. ID-185). Salt Lake City, UT.
Revised 1991.
9.2. Occupational Safety and Health Administration Technical
Center: Metal and Metalloid Particulated in Workplace Atmospheres (ICP Analysis)
by J.C. Septon (USDOL/OSHA-SLTC Method No. ID-125G). Salt Lake
City, UT. Revised 1991.
9.3. Occupational Safety and Health Administration Analytical
Laboratory: OSHA Manual of Analytical Methods edited by R.G. Adler (Method No. I-1).
Salt Lake City, UT. 1977.
9.4. Carsey, T.P.: Quantitation of Vanadium Oxides in
Airborne Dusts by X-ray Diffraction. Anal. Chem. 57:2125-2130 (1985)
9.5. Gelman Sciences: The Filter Book. Ann Arbor, MI:
Gelman Sciences, 1991
9.6. Occupational Safety and Health Administration Analytical
Laboratory: X-ray Diffraction Program Documentation. Salt Lake
City, UT. 1981 (unpublished)
9.7. Leroux, J. and C. Powers: Direct X-ray
Diffraction Quantitative Analysis of Quartz in Industrial Dust Film Deposited on Silver
Membrane Filters. Occup. Health Rev. 21:26-34 (1970).
9.8. National institute for Occupational Safety and Health: Collaborative
Tests of Two Methods for Determining Free Silica in Airborne Dust (DHHS Publication
No. 83-124). Cincinnati, OH: National Institute for Occupational Safety and
Health, and the Bureau of Mines, 1983.
9.9. Mandel, J.: Accuracy and Precision, Evaluation and
Interpretation of Analytical Results, The treatment of Outliers. In Treatise on
Analytical Chemistry, 2nd ed., edited by I. M. Kolthoff and P. J. Elving. New York:
John Wiley and Sons, 1978. pp. 282-285.
9.10. Youden, W.J.: Statistical Methods for Chemists.
New York: John Wiley and Sons, 1964. p 20.
9.11. Occupational Safety and Health Administration Analytical
Laboratory: Precision and Accuracy Data Protocol for Laboratory Validations. In OSHA
Analytical Methods Manual. Cincinnati, OH: American Conference of Governmental
Industrial Hygienists (Pub. No. ISBN: 0-936712-66-X), 1985.
9.12. Long, G.L. and J.D. Winefordner: Limit of Detection
-- A Closer Look at the IUPAC Definition. Anal. Chem. 55: 712A-724A
(1983).
9.13. Analytical Methods Committee: Recommendations for the
Definition, Estimation and Use of the Detection Limit. Analyst 112:199-204
(1987).
9.14. Westgard, J.O. and M.R. Hunt.: Use and Interpretation
of Common Statistical Tests in Method Comparison Studies. Clinical Chemistry 19:49
(1973).
9.15. Ripley, B.D. and M. Thompson: Regression Techniques
for the Detection of Analytical Bias. Analyst 112:337-383 (1987).
9.16. Gore, W.L.: Statistical Methods. New York:
Interscience, 1952, pp. 189-191.
9.17. Occupational Safety and Health Administration Analytical
Laboratory: OSHA Analytical Methods Manual (USDOL/OSHA-SLCAL Method No. ID-142).
Cincinnati, OH: American Conference of Governmental Industrial Hygienists (Pub. No. ISBN: 0-936712-66-X),
1985.
9.18. Occupational Safety and Health Administration Analytical
Laboratory: The effects of Deviation in Methodology - OSHA vs. NIOSH Results by
M.C. Rose. Salt Lake City, UT. 1987 (unpublished).
Table 1
Calibration Data
Freezermill ("W") Material
APD-A X-Ray Diffractometer
V2O5
µg Taken
|
Counts
|
µg Calculated
|
Mean
|
Std Dev
|
CV1
|
50
50
50 |
818
827
860 |
46.9
47.4
49.3 |
|
(46.4 - 49.4)* |
47.9 |
1.27 |
0.0264 |
100
100
100 |
1522
1634
1718 |
87.8
94.4
99.3 |
|
(87.2 - 100.4) |
93.8 |
5.77 |
0.0615 |
200
200
200 |
3444
3573
3306 |
202.2
210.0
193.9 |
|
(192.7 - 211.3) |
202.0 |
8.05 |
0.0399 |
250
250
250 |
4040
4195
4509 |
238.5
248.1
267.4 |
|
(234.4 - 268.2) |
251.3 |
14.72 |
0.0586 |
499
499
499 |
8054
9131
9270 |
495.3
568.2
577.7 |
|
(495.3 - 598.9) |
547.1 |
45.08 |
0.0824 |
998
998
998 |
14909
13309
15791 |
969.3
855.8
1032.7 |
|
(849.5 - 1055.7) |
952.6 |
89.62 |
0.0941 |
1996
1996
1996 |
29359
27312
27943 |
2096.7
1923.8
1976.5 |
|
(1897.1 - 2100.9) |
1999.0 |
88.62 |
0.0443 |
2495
2495
2495 |
35312
34389
35014 |
2632.4
2545.7
2604.3 |
|
(2543.2 - 2645.0) |
2594.1 |
44.24 |
0.0171 |
* Acceptable ranges from the ASTM test are shown in parentheses.
Calibration fit spliced at 500 µg, 8125 counts: |
Low fit :
High fit: |
Counts =
Counts = |
0 + 17.562281 × µg -
0.002626 × µg2
399.9865 + 15.962334 × µg - 0.001026 ×
µg2 |
Table 1 (Continued)
Calibration Data
Freezer mill ("W") Material
APD-B X-ray Diffractometer
V2O5
µg Taken
|
Counts
|
µg Calculated
|
Mean
|
Std Dev
|
CV1
|
50
50
50 |
1327
812
974 |
*
43.5
52.3 |
|
(Test not applicable) |
47.9 |
6.22 |
0.1299 |
100
100
100 |
1736
1593
1888 |
94.0
86.1
102.4 |
|
(84.8 - 103.6)** |
94.2 |
8.15 |
0.0865 |
200
200
200 |
3691
3720
3922 |
204.5
206.1
217.9 |
|
(201.1 - 217.9) |
209.5 |
7.32 |
0.0349 |
250
250
250 |
4440
4358
4456 |
248.2
243.4
249.2 |
|
(243.3 - 250.5) |
246.9 |
3.10 |
0.0126 |
499
499
499 |
8097
9462
9216 |
475.9
567.9
551.0 |
|
(475.3 - 587.9) |
531.6 |
48.97 |
0.0921 |
998
998
998 |
15331
13357
15996 |
968.6
831.7
1015.2 |
|
(828.8 - 1048.2) |
938.5 |
95.38 |
0.1016 |
1996
1996
1996 |
31381
29308
29709 |
2163.3
1999.8
2031.2 |
|
(1965.0 - 2164.6) |
2064.8 |
86.76 |
0.0420 |
2495
2495
2495 |
36411
36811
34880 |
2573.9
2607.4
2446.8 |
|
(2445.3 - 2640.1) |
2542.7 |
84.72 |
0.0333 |
* |
One of the 50 µg standards appeared as an outlier. Although the result was acceptable
on APD-A, the standard was not used for calibrating APD-B. |
** |
Acceptable ranges from the ASTM test are shown in parentheses. |
Calibration fit spliced at 500 µg, 8461 counts: |
Low fit :
High fit: |
Counts =
Counts = |
0 + 18.836058 × µg -
0.003830 × µg2
773.2365 + 15.743113 × µg - 0.000737 ×
µg2 |
Table 1 (Continued)
Calibration Data
Freezer mill ("W") Material
EDXRF
V2O5
µg Taken
|
Counts
|
µg Calculated
|
Mean
|
Std Dev
|
CV1
|
50
50
50 |
246
206
220 |
60.7
50.6
54.2 |
|
(49.3 - 61.0)* |
55.2 |
5.12 |
0.0928 |
100
100
100 |
412
359
415 |
102.4
89.1
103.2 |
|
|
(89.1 - 107.3) |
98.2 |
7.92 |
0.0806 |
200
200
200 |
869
785
848 |
217.2
196.1
212.0 |
|
(195.8 - 221.1) |
208.4 |
10.99 |
0.0527 |
250
250
250 |
1005
976
1015 |
251.4
244.1
253.9 |
|
(244.0 - 255.7) |
249.8 |
5.09 |
0.0204 |
499
499
499 |
2076
2178
2093 |
520.5
546.2
524.8 |
|
(514.7 - 546.3) |
530.5 |
13.77 |
0.0259 |
998
998
998 |
3924
3579
3811 |
984.9
898.2
956.5 |
|
(895.7 - 997.3) |
946.5 |
44.20 |
0.0467 |
1996
1996
1996 |
8361
7419
7750 |
2099.7
1863.0
1946.2 |
|
(1831.6 - 2107.8) |
1969.7 |
120.07 |
0.0610 |
2495
2495
2495 |
9941
9826
10449 |
2496.8
2467.8
2624.4** |
|
(2423.9 - 2615.5) |
2529.7 |
83.31 |
0.0329 |
* |
Acceptable ranges from the ASTM test are shown in parentheses. |
** |
Although this standard appeared outside of the ASTM test range, it was used in the
calibration. |
EDXRF curve fit:
Counts = 4.427507051 + 3.979798615 × µg
+ 0 × µg2
Table 1 (Continued)
Calibration Data
Freezer mill ("W") Material*
EDXRF
V2O5
µg Taken
|
Counts
|
µg Calculated
|
50
100
200
250
499
998
1996
2495 |
210
404
756
945
2114
3873
7754
10146 |
41.1
93.7
188.6
239.3
548.9
1002.4
1956.8
2517.0 |
* |
Standards prepared from this material were used to analyze fume-like "Fine-W"
samples. |
EDXRF curve fit:
Counts = 59.00722287 + 3.670421406 × µg
+ 1.33910E-04 × µg2
Table 2
Analysis - Spiked Sample Data
Freezer mill ("W") Material
APD-A X-ray Diffractometer
V2O5
µg Taken
|
Recovery Range*
|
N
|
Mean
|
Std Dev
|
CV1
|
237 |
(0.738 - 1.085) |
6 |
0.911 |
0.0895 |
0.0982 |
474 |
(0.639 - 1.149) |
6 |
0.894 |
0.1314 |
0.1470 |
710 |
(0.706 - 1.046) |
6 |
0.876 |
0.0877 |
0.1001 |
Average Recovery
CV1 (Pooled) |
= 0.894
= 0.1173 |
APD-B X-ray Diffractometer
V2O5
µg Taken
|
Recovery Range*
|
N
|
Mean
|
Std Dev
|
CV1
|
237 |
(0.690 - 1.028) |
6 |
0.859 |
0.0871 |
0.1015 |
474 |
(0.601 - 1.174) |
6 |
0.887 |
0.1476 |
0.1663 |
710 |
(0.731 - 1.059) |
6 |
0.895 |
0.0846 |
0.0946 |
Average Recovery
CV1 (Pooled) |
= 0.880
= 0.1250 |
* Acceptable ranges from the ASTM test are shown in parentheses.
Table 3
Analysis - Spiked Sample Data
Alumina Mortar and Pestle ("M") Material
APD-A X-ray Diffractometer
V2O5
µg Taken
|
Recovery Range*
|
N
|
Mean
|
Std Dev
|
CV1
|
233 |
(1.483 - 1.841) |
6 |
1.662 |
0.0923 |
0.0555 |
467 |
(1.510 - 1.960) |
6 |
1.735 |
0.1158 |
0.0668 |
700 |
(1.450 - 1.818) |
5** |
1.634 |
0.1052 |
0.0644 |
Average Recovery
CV1 (Pooled) |
= 1.680
= 0.0624 |
APD-B X-ray Diffractometer
V2O5
µg Taken
|
Recovery Range*
|
N
|
Mean
|
Std Dev
|
CV1
|
233 |
(1.560 - 2.158) |
6 |
1.859 |
0.1543 |
0.0830 |
467 |
(1.629 - 2.176) |
6 |
1.902 |
0.1414 |
0.0743 |
700 |
(1.622 - 2.060) |
6 |
1.841 |
0.1129 |
0.0613 |
Average Recovery
CV1 (Pooled) |
= 1.867
= 0.0734 |
* |
Acceptable ranges from the ASTM test are shown in parentheses. |
** |
One of the 700 µg spiked samples was damaged in transfer from APD-B to APD-A;
it was not used in subsequent analyses. |
Table 4
Analysis - Spiked Sample Data
Freezer-mill ("W") Material
EDXRF
V2O5
µg Taken
|
Recovery Range*
|
N
|
Mean
|
Std Dev
|
CV1
|
237 |
(0.788 - 1.000) |
6 |
0.8938 |
0.0547 |
0.0612 |
474 |
(0.632 - 1.064) |
5** |
0.8481 |
0.1235 |
0.1456 |
710 |
(0.744 - 0.992) |
6 |
0.8681 |
0.0640 |
0.0737 |
Average Recovery
CV1 (Pooled) |
= 0.8713
= 0.0966 |
Table 5
Analysis - Spiked Sample Data
Alumina Mortand and Pestle ("M") Material
EDXRF
V2O5
µg Taken
|
Recovery Range*
|
N
|
Mean
|
Std Dev
|
CV1
|
233 |
(0.825 - 1.126) |
6 |
0.9759 |
0.0776 |
0.0795 |
467 |
(0.839 - 1.076) |
6 |
0.9574 |
0.0612 |
0.0639 |
700 |
(0.895 - 1.028) |
5** |
0.9615 |
0.0377 |
0.0392 |
Average Recovery
CV1 (Pooled) |
= 0.9651
= 0.0642 |
* |
Acceptable ranges from the ASTM test are shown in parentheses. |
** |
One sample was lost in analysis |
*** |
One of the 700-µg spiked samples was damaged in transfer from APD-B to APD-A;
it was not used in subsequent analyses. |
Table 6
Analysis - Spiked Sample Data
Freezer-mill ("W") Material
AAS
V2O5
µg Taken
|
Recovery Range*
|
N
|
Mean
|
Std Dev
|
CV1
|
237 |
(0.857 - 0.963) |
6 |
0.9096 |
0.0273 |
0.0300 |
474 |
(0.843 - 0.957) |
6 |
0.9004 |
0.0294 |
0.0327 |
710 |
(0.838 - 0.942) |
6 |
0.8900 |
0.0268 |
0.0301 |
Average Recovery
CV1 (Pooled) |
= 0.9000
= 0.0309 |
Table 7
Analysis - Spiked Sample Data
Alumina Mortand and Pestle ("M") Material
AAS
V2O5
µg Taken
|
Recovery Range*
|
N
|
Mean
|
Std Dev
|
CV1
|
233 |
(0.938 - 1.001) |
6 |
0.9692 |
0.0163 |
0.0168 |
467 |
(0.905 - 0.954) |
6 |
0.9295 |
0.0124 |
0.0134 |
700 |
(0.875 - 0.924) |
5** |
0.8994 |
0.0137 |
0.0153 |
Average Recovery
CV1 (Pooled) |
= 0.9327
= 0.0152 |
* |
Acceptable ranges from the ASTM test are shown in parentheses. |
** |
One of the 700-µg spiked samples was damaged in transfer from APD-B to APD-A;
it was not used in subsequent analyses. |
Table 8
µg V2O5 "Fine-W"
Recoveries
Spiked on Silver Membranes
Reported by Analysis
Sample
|
APD-A
|
APD-B
|
EDXRF
|
AAS
|
A
B
C |
190.2
199.3
259.9 |
188.0
227.2
256.0 |
309.5
329.0
373.9 |
297.9
344.9
371.5 |
Table 9
µg V2O5 "W" Recoveries
Spiked on PVC Membranes
Reported by Analysis
|
"True"
|
APD-A
|
APD-B
|
EDXRF
|
AAS
|
237
237
237
237
237
237
474
474
474
474
474
474
710
710
710
710
710
710 |
234.5
220.5
245.9
195.1
204.4
195.5
498.0
403.3
482.0
338.4
375.5
444.4
548.7
548.6
658.3
612.3
669.9
692.2 |
236.7
202.4
218.8
182.8
188.5
192.0
500.9
434.3
496.5
341.6
346.4
403.2
614.5
542.5
682.3
603.4
662.5
706.4 |
233.6
206.7
221.3
207.2
202.7
199.7
489.6
403.7
*
338.6
360.2
418.0
555.0
576.8
677.1
607.7
640.9
641.1 |
225.3
212.3
218.8
209.1
218.8
209.1
430.1
430.1
443.6
410.0
410.0
436.9
598.1
636.6
640.0
622.5
650.6
643.6 |
µg V2O5 "M"
Recoveries
|
"True"
|
APD-A
|
APD-B
|
EDXRF
|
AAS
|
233
233
233
233
233
233
467
467
467
467
467
467
700
700
700
700
700
700 |
355.2
412.2
410.8
378.6
382.5
384.5
761.9
886.6
787.7
867.2
797.7
760.2
1079.0
*
1111.3
1230.0
1215.6
1081.9 |
363.5
443.3
442.7
433.5
467.3
448.5
857.9
990.2
807.5
941.8
864.1
868.9
1173.3
1337.8
1277.4
1382.2
1337.0
1223.3 |
222.3
255.7
218.5
202.2
228.8
236.8
484.6
482.3
427.3
425.5
437.9
425.0
666.8
*
654.4
709.7
689.4
644.9 |
228.5
222.0
231.8
225.3
222.0
225.3
426.8
433.5
436.9
443.6
433.5
430.1
615.5
*
636.6
626.0
640.1
629.5 |
* Sample lost in analysis.
Table 10
Detection Limit Determination
Blanks Prepared as Spiked Samples
100-s Analysis Time & Normalized Counts
Sample
|
APD-A
Counts
|
APD-B
Counts
|
EDXRF
Counts
|
A
B
C
D
E
F
G |
330
135
105
187
196
108
408 |
434
113
106
165
231
191
545 |
59
11
22
41
42
8
39 |
|
Average |
209.86 |
255.00 |
31.714 |
|
SD |
116.46 |
168.96 |
18.599 |
Slope (Count/µg) |
17.562 |
18.836 |
3.9798 |
|
D.L. |
19.9 |
26.9 |
14.0 |
Estimated DL |
25 µg V2O5 (XRD) |
14 µg V2O5 (XRF) |
Detection Limit Determination
Blanks Prepared as Spiked Samples
1,000 s Analysis Time & Normalized to Compare with 100-s Results
Sample
|
APD-A
Counts
|
APD-B
Counts
|
EDXRF
Counts
|
A
B
C
D
E
F
G
H
I |
214
27
47
42
21
15
234
195
97 |
251
21
94
24
81
84
351
435
106 |
1.0
4.3
4.3
8.4
0
4.4
4.5
1.2
0 |
|
Average |
99.11 |
160.78 |
3.12 |
|
SD |
90.10 |
148.95 |
2.78 |
Slope (Count/µg) |
17.562 |
18.836 |
3.9798 |
|
D.L. |
15.4 |
23.7 |
2.10 |
Estimated DL |
20 µg V2O5 (XRD) |
2 µg V2O5 (XRF) |
Where DL = 3(SD) / sensitivity
Appendix 1
The multichannel analyzer was set to 512 channels and the instrument was calibrated
using the TiO2-ZnO-Y2O3
calibration standard. The spectrum range was approximately 1.2 to 20.6 kV.
Finnigan EDXRF Program
|
Changes for analysis on different days
|
Steps to be entered
on instrument console
|
Analyses at 100 s
|
Analyses at 1,000 s
|
(Begin programming.)
LEARN |
"Fine-W" and DL |
DL |
(Program system to set up analytical conditions.) |
100 SEC
1ST HALF
CLEAR+CLEAR
MARKER (Z=23)
OUTPUT |
100 s |
1,000 s |
(Program system to acquire spectrum.)
ACQUIRE |
(Program the instrument to obtain net counts using background
subtraction.)
NET
START/SPAN/SELECT
[Set background low energy region to ~3.4 kV
(between Ag La and V Ka
peaks).]
START = 71
SPAN = 5
INTENS
[Set background high energy region to ~5.9 kV (just beyond the V Kb peak).] |
START = 124
SPAN = 5
INTENS |
127 |
127 |
[Set region to span the V Ka
analytical peak at 4.949 kV (~0.64 kV span.)] |
START = 91
SPAN = 17
INTENS |
93 |
93
14 |
(End of programming.)
EXECUTE |
Note: This program was written for a specific instrument (Finnigan 77-900A).
Commands are capitalized. Background regions should be adjusted when interferences are
present.
|